Induztr
Engineering Digest Document Viewer Related Models

Agilent InfinityLab LC Series 1220 Infinity II LC System User Manual

Contents

  1. Overview pp. 9-16
    System overview, EMF counters, power cord information
  2. Spec pp. 21-28
    Pump and VWD performance specifications
  3. Procedure pp. 29-60
    Accessory kit, system installation, PC connection, priming, column and restriction capillary
  4. Procedure pp. 61-84
    Default settings, link configuration, MAC address, IP address via BootP and Telnet
  5. Overview pp. 85-100
    Pump description, compressibility compensation, solvent filter blockage prevention
  6. Overview pp. 101-116
    Injection sequence, transport assembly, ALS operation
  7. Overview pp. 117-136
    VWD and DAD detector description, spectral acquisition, flow cell selection
  8. Procedure pp. 137-192
    Flow cell tests, ALS alignment, oven calibration, intensity and filter tests, spectral flatness
  9. Procedure pp. 193-240
    General, pump, ALS and VWD error messages and troubleshooting
  10. Procedure pp. 241-324
    Cleaning, valve and pump head exchange, leak correction, flow cell capillary replacement
  11. Procedure pp. 325-340
    Firmware update, mainboard overview, purge valve, sample loops
  12. Spec pp. 369-384
    Module to integrator connections, CAN/LAN cables
  13. Quality pp. 385-396
    Radio interference and declaration of conformity
  14. p. 25 table, 14 rows
  15. p. 26 table, 14 rows
  16. p. 20 table, 4 rows
  17. p. 65 table, 3 rows
  18. p. 82 table, 6 rows
  19. p. 21 table, 12 rows
  20. p. 22 table, 12 rows
  21. p. 23 table, 9 rows
  22. p. 24 table, 5 rows
  23. p. 69 table, 5 rows

Table 7 Performance Specifications Agilent 1220 Infinity II LC VWD

TypeSpecificationComments
Detection typeDouble-beam photometer
Light sourceDeuterium lamp
Maximum data rate80 Hz
Short term noise (ASTM)<± 0.25·10-5 AU at 230 nmSee Specification Conditions below
Drift< 1·10-4 AU/h at 230 nmSee Specification Conditions below
Linearity> 2 AU (5 %) upper limitSee Specification Conditions below
Wavelength range190 – 600 nm
Wavelength accuracy± 1 nmSelf-calibration with deuterium lines, verification with holmium oxide filter
Slit width6.5 nm typical over whole wavelength range
Flow cellsStandard: 14 µL volume, 10 mm cell path length and 40 bar (580 psi) pressure maximum
High pressure: 14 µL volume, 10 mm cell path length and 400 bar (5800 psi) pressure maximum
Micro: 1 µL volume, 5 mm cell path length and 40 bar (580 psi) pressure maximum
Semi-micro: 5 µL volume, 6 mm cell path length and 40 bar (580 psi) pressure maximum
Can be repaired on component level

Table 8 Performance Specifications Agilent 1220 Infinity II LC DAD

TypeSpecificationComments
Detection type1024-element diode array
Light sourceDeuterium and tungsten lampsThe UV-lamp is equipped with RFID tag that holds lamp typical information.
Number of signals8
Maximum sampling rate80 Hz
Short term noise (ASTM) Single and Multi-Wavelength< ± 0.7·10-5 AU at 254 nm and 750 nmsee "Specification Conditions" below
Drift< 0.9·10-3 AU/h at 254 nmsee "Specification Conditions" below
Linearity> 2 AU (5 %see "Specification Conditions" below
Wavelength range190 – 950 nm
Wavelength accuracy± 1 nmSelf-calibration with deuterium lines, verification with holmium oxide filter
Slit width1, 2, 4 , 8, 16 nmProgrammable slit
Diode width~ 1 nm
Flow cellStandard: 13 µL volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximumThe flow cell is equipped with RFID tags that hold cell typical information.
pH range1.0 – 9.5
Time programmableWavelength, polarity, peak width, lamp bandwidth, autobalance, wavelength range, threshold, spectra storage mode

Table 2 Performance Specifications Agilent 1220 Infinity II LC

TypeSpecification
Safety featuresExtensive diagnostics, error detection and display, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
Control and data evaluationAgilent EZChrom Compact, Agilent ChemStation, Agilent Instrument Utilities, Agilent Lab Advisor
CommunicationsController-area network (CAN), RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN
GLP featuresEarly maintenance feedback (EMF), electronic records of maintenance and errors

Table 12 Using Default Parameters

ParameterDefault
IP address:192.168.254.11
Subnet Mask:255.255.255.0
Default Gatewaynot specified

Table 16 Telnet Commands

ValueDescription
?displays syntax and descriptions of commands
/displays current LAN settings
ip <x.x.x.x>sets new ip address
sm <x.x.x.x>sets new subnet mask
gw <x.x.x.x>sets new default gateway
exitexits shell and saves all changes

Table 3 Performance Specifications Agilent 1220 Infinity II LC Pump

TypeSpecificationComments
Hydraulic systemDual plunger in series pump with proprietary servo-controlled variable stroke drive, floating plungers and passive inlet valve
Flow rangeSettable: 0.001 – 10 mL/min Recommended: 0.2 – 10.0 mL/minSet points in 0.001 mL/min increments
Flow precision≤0.07 % RSD, or ≤0.02 min SD whichever is greaterBased on retention time at constant temperature
Flow accuracy± 1 % or 10 µL/min whatever is greaterPumping degassed H2O at 10 MPa (100 bar)
Pressure operating rangeUp to 60 MPa (600 bar, 8702 psi) up to 5 mL/min Up to 20 MPa (200 bar, 2901 psi) up to 10 mL/min
Pressure pulsation< 2 % amplitude (typically < 1.3 %), or < 0.3 MPa (3 bar, 44 psi), whichever is greater
Compressibility compensationUser-selectable, based on mobile phase compressibility
Recommended pH1.0 – 12.5Solvents with pH < 2.3 should not contain acids which attack stainless steel
Gradient formation (optional)Low pressure binary mixing/gradient capability using proprietary high-speed proportioning valve
Delay volume600 – 900 µL, dependent on back pressureMeasured with water at 1 mL/min (water/water with tracer)
Composition rangeSettable 0 – 100 % Recommended 5 – 95 %In 0.1 % increments
Composition precision< 0.2 % RSD or < 0.04 min SD, whichever is greaterAt 1 mL/min; based on retention time at constant temperature

Table 4 Performance Specifications Agilent 1220 Infinity II LC Pump VL

TypeSpecificationComments
Hydraulic systemDual plunger in series pump with proprietary servo-controlled variable stroke drive, floating plungers and passive inlet valve
Flow rangeSettable 0.001 – 10 mL/minRecommended: 0.2 – 10.0 mL/min Set points in 0.001 mL/min increments
Flow precision≤0.07 % RSD, or ≤0.02 min SD whichever is greaterBased on retention time at constant temperature
Flow accuracy± 1 % or 10 µL/min whatever is greaterPumping degassed H2O at 10 MPa (100 bar)
Pressure operating rangeUp to 40 MPa (400 bar, 5801 psi) up to 5 mL/min Up to 20 MPa (200 bar, 2901 psi) up to 10 mL/min
Pressure pulsation< 2 % amplitude (typically < 1.3 %), or < 0.3 MPa (3 bar, 44 psi), whichever is greater
Compressibility compensationUser-selectable, based on mobile phase compressibility
Recommended pH range1.0 – 12.5Solvents with pH < 2.3 should not contain acids which attack stainless steel
Gradient formationLow pressure binary mixing/gradient capability using proprietary high-speed proportioning valve
Delay volume600 – 900 µL, dependent on back pressureMeasured with water at 1 mL/min (water/water with tracer)
Composition rangeSettable: 0 – 100 %Recommended: 5 – 95 %, In 0.1 % increments
Composition precision< 0.2 % RSD or < 0.04 min SD, whichever is greaterAt 1 mL/min; based on retention time at constant temperature

Table 5 Performance Specifications Agilent 1220 Infinity II LC Autosampler

TypeSpecificationComments
Injection range0.1 – 100 µL in 0.1 µL incrementsUp to 1500 µL with multiple draw (hardware modification required)
Precision<1 % RSD of peak areas from 1 µL to 5 µL <0.25 % RSD of peak areas from 5 µL to 100 µLMeasured with caffeine
Replicate injections1 – 99 from one vial
Pressure range0 – 60 MPa (0 – 600 bar, 0 – 8702 psi)
Sample viscosity range0.2 – 50 cp
Sample capacity100 × 2 mL vials in 1 tray 40 × 2 mL vials in ½ tray 15 × 6 mL vials in ½ tray (Agilent vials only)
Carry overTypically < 0.1 %, < 0.05 % with external needle cleaning
Injection cycle timeTypically 50 s depending on draw speed and injection volume
Minimum sample volume1 µL from 5 µL sample in 100 µL microvial, or 1 µL from 10 µL sample in 300 µL microvial

Table 6 Performance Specifications Agilent 1220 Infinity II LC Column Oven

TypeSpecificationComments
Temperature range5 °C above ambient to 80 °C10 °C above ambient to 80 °C (G4294B with DAD)
Column capacity1 column up to 25 cm
Temperature stability± 0.15 °CConstant composition and flow rate
Temperature accuracy± 0.8 °C
Internal volume6 µL

Table 15 Link Configuration Switches

SW 3SW 4SW 5Link Configuration
OFFspeed and duplex mode determined by auto-negotiation
ONOFFOFFmanually set to 10 Mbps, half-duplex
ONOFFONmanually set to 10 Mbps, full-duplex
ONONOFFmanually set to 100 Mbps, half-duplex
ONONONmanually set to 100 Mbps, full-duplex

Engineering Facts

14 facts · every claim linked to a page
Installation

The instrument requires additional clearance on the sides and rear for air circulation and electrical connections.

Bench space must be planned with side and rear clearance to avoid overheating and allow cable access.

Source
“It needs an additional 2.5 cm (1.0 in) of space on either side and approximately 8 cm (3.1 in) at the rear for air circulation and electric connections.”
Safety

The instrument remains partially energized even with the front power switch off, so the power cord must be unplugged for repair work.

Service personnel must unplug the power cord before opening the detector cover to avoid shock hazard.

Source
“The power supply still uses some power even when the power switch on the front panel is turned OFF.”
Safety

The instrument must be connected only to a grounded outlet and only with the Agilent power cord specified for the region.

Prevents electric shock and short circuit during installation.

Source
“Never operate your instrument from a power outlet that has no ground connection.”
Installation

Condensation can damage electronics, so a cold-shipped module must be left in its box to warm slowly to room temperature.

Prevents condensation-induced damage to system electronics after cold transport.

Source
“If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation.”
Performance

Drift performance depends on temperature stability, and the highest performance requires limiting ambient changes to below 1 °C/hour.

Lab HVAC stability directly affects detector drift and must be considered at site selection.

Source
“To realize the highest performance, minimize the frequency and the amplitude of the temperature changes to below 1 °C/hour (1.8 °F/hour).”
Operation

The pump pressure range is limited to 60 MPa up to 5 mL/min and 20 MPa up to 10 mL/min.

Method flow rate and column backpressure must stay within these combined limits.

Source
“Up to 60 MPa (600 bar, 8702 psi) up to 5 mL/min”
Operation

The pump should not be run without a solvent filter installed.

Prevents particles and algae from blocking the pump and degrading performance.

Source
“Never use the system without a solvent filter installed.”
Maintenance

Aqueous solvents and phosphate buffers (pH 4–7) promote algae growth that shortens solvent filter life, so sterile amber bottles and periodic solvent exchange are recommended.

Mobile phase choice and bottle handling determine filter maintenance frequency.

Source
“This is especially true for aqueous solvents or phosphate buffers (pH 4 – 7).”
Maintenance

The pump leak test requires blocking the pump with a blank nut and running with isopropanol while monitoring the pressure profile.

Technicians need the correct setup and solvent to obtain a valid leak test result.

Source
“The test requires blocking the pump with a blank nut, then running the test with isopropanol (IPA), while monitoring the pressure profile.”
Maintenance

Any trace of other solvents or an air bubble in the flow path will cause the pump leak test to fail.

Thorough IPA flushing is mandatory before pressurizing, otherwise results are invalid.

Source
“Any trace of other solvents or the smallest air bubble inside the flow path will definitely cause the test to fail!”
Maintenance

The pump must never be started with the pump head removed because it can damage the pump drive.

Critical precaution during seal or plunger exchange to avoid drive damage.

Source
“Never start the pump when the pump head is removed.”
Operation

For DAD, a 4 nm slit generally gives the best results, while narrow slits suit fine spectral structures or high-concentration quantification.

Slit width selection affects sensitivity, spectral resolution and linearity in method development.

Source
“In most situations, a slit width of 4 nm will give the best results.”
Compliance

Oven temperature calibration is normally not required, but a 2-point calibration and verification may be performed to comply with local regulatory requirements.

Regulated labs must know when oven calibration is required for compliance.

Source
“in order to comply with local regulatory requirements, the 2-point calibration and verification procedure may be performed.”
Operation

If the gradient valve is disconnected and the pump rebooted, the pump functions as an isocratic pump and the error reappears at each switch-on.

Explains unexpected loss of gradient capability and recurring error after service.

Source
“However, the pump will function as an isocratic pump in this configuration.”