Table 2 Performance Specifications Agilent 1220 Infinity II LC
Type
Specification
Safety features
Extensive diagnostics, error detection and display, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
speed and duplex mode determined by auto-negotiation
ON
OFF
OFF
manually set to 10 Mbps, half-duplex
ON
OFF
ON
manually set to 10 Mbps, full-duplex
ON
ON
OFF
manually set to 100 Mbps, half-duplex
ON
ON
ON
manually set to 100 Mbps, full-duplex
Engineering Facts
14 facts · every claim linked to a page
Installation
The instrument requires additional clearance on the sides and rear for air circulation and electrical connections.
Bench space must be planned with side and rear clearance to avoid overheating and allow cable access.
Source
“It needs an additional 2.5 cm (1.0 in) of space on either side and approximately 8 cm (3.1 in) at the rear for air circulation and electric connections.”
Safety
The instrument remains partially energized even with the front power switch off, so the power cord must be unplugged for repair work.
Service personnel must unplug the power cord before opening the detector cover to avoid shock hazard.
Source
“The power supply still uses some power even when the power switch on the front panel is turned OFF.”
Safety
The instrument must be connected only to a grounded outlet and only with the Agilent power cord specified for the region.
Prevents electric shock and short circuit during installation.
Source
“Never operate your instrument from a power outlet that has no ground connection.”
Installation
Condensation can damage electronics, so a cold-shipped module must be left in its box to warm slowly to room temperature.
Prevents condensation-induced damage to system electronics after cold transport.
Source
“If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation.”
Performance
Drift performance depends on temperature stability, and the highest performance requires limiting ambient changes to below 1 °C/hour.
Lab HVAC stability directly affects detector drift and must be considered at site selection.
Source
“To realize the highest performance, minimize the frequency and the amplitude of the temperature changes to below 1 °C/hour (1.8 °F/hour).”
Operation
The pump pressure range is limited to 60 MPa up to 5 mL/min and 20 MPa up to 10 mL/min.
Method flow rate and column backpressure must stay within these combined limits.
Source
“Up to 60 MPa (600 bar, 8702 psi) up to 5 mL/min”
Operation
The pump should not be run without a solvent filter installed.
Prevents particles and algae from blocking the pump and degrading performance.
Source
“Never use the system without a solvent filter installed.”
Maintenance
Aqueous solvents and phosphate buffers (pH 4–7) promote algae growth that shortens solvent filter life, so sterile amber bottles and periodic solvent exchange are recommended.
Mobile phase choice and bottle handling determine filter maintenance frequency.
Source
“This is especially true for aqueous solvents or phosphate buffers (pH 4 – 7).”
Maintenance
The pump leak test requires blocking the pump with a blank nut and running with isopropanol while monitoring the pressure profile.
Technicians need the correct setup and solvent to obtain a valid leak test result.
Source
“The test requires blocking the pump with a blank nut, then running the test with isopropanol (IPA), while monitoring the pressure profile.”
Maintenance
Any trace of other solvents or an air bubble in the flow path will cause the pump leak test to fail.
Thorough IPA flushing is mandatory before pressurizing, otherwise results are invalid.
Source
“Any trace of other solvents or the smallest air bubble inside the flow path will definitely cause the test to fail!”
Maintenance
The pump must never be started with the pump head removed because it can damage the pump drive.
Critical precaution during seal or plunger exchange to avoid drive damage.
Source
“Never start the pump when the pump head is removed.”
Operation
For DAD, a 4 nm slit generally gives the best results, while narrow slits suit fine spectral structures or high-concentration quantification.
Slit width selection affects sensitivity, spectral resolution and linearity in method development.
Source
“In most situations, a slit width of 4 nm will give the best results.”
Compliance
Oven temperature calibration is normally not required, but a 2-point calibration and verification may be performed to comply with local regulatory requirements.
Regulated labs must know when oven calibration is required for compliance.
Source
“in order to comply with local regulatory requirements, the 2-point calibration and verification procedure may be performed.”
Operation
If the gradient valve is disconnected and the pump rebooted, the pump functions as an isocratic pump and the error reappears at each switch-on.
Explains unexpected loss of gradient capability and recurring error after service.
Source
“However, the pump will function as an isocratic pump in this configuration.”