Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model KT 115

Technical Reference Summary

Covered Models & Document Scope

BINDER KT 115 cooling incubator with Peltier technology, interior volume 102 L. Two standard variants: KT115-230V (Art.-No. 9020-0313, 200...240 V 1~ 50/60 Hz, CEE 7/7 plug) and KT115UL-120V (Art.-No. 9020-0314, 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug).

Technical Data

Parameter Value Condition
Temperature range 4...100 °C p.2 ↗ KT115-230V / KT115UL-120V
Temperature uniformity at 37°C 0.3 ± K p.2 ↗ KT115-230V / KT115UL-120V
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ KT115-230V / KT115UL-120V
Max. heat compensation at 40°C 100 W p.2 ↗ KT115-230V / KT115UL-120V
Recovery time after door was opened for 30 s at 37°C 3 min p.2 ↗ KT115-230V / KT115UL-120V
Rated voltage 200...240 V p.2 ↗ KT115-230V
Rated voltage 100...120 V p.2 ↗ KT115UL-120V
Power frequency 50/60 Hz p.2 ↗ KT115-230V / KT115UL-120V
Nominal power 0,7 kW p.2 ↗ KT115-230V / KT115UL-120V
Unit fuse 10,0 A p.2 ↗ KT115-230V / KT115UL-120V
Phase (nominal voltage) 1~ p.2 ↗ KT115-230V / KT115UL-120V
Interior volume 102 L p.2 ↗ KT115-230V / KT115UL-120V
Net weight of the unit (empty) 83 kg p.2 ↗ KT115-230V / KT115UL-120V
Load per rack 30 kg p.2 ↗ KT115-230V / KT115UL-120V
Permitted load 100 kg p.2 ↗ KT115-230V / KT115UL-120V
Wall clearance back 100 mm p.2 ↗ KT115-230V / KT115UL-120V
Wall clearance sidewise 240 mm p.2 ↗ KT115-230V / KT115UL-120V
Width net 860 mm p.2 ↗ KT115-230V / KT115UL-120V
Height net 715 mm p.2 ↗ KT115-230V / KT115UL-120V
Depth net 655 mm p.2 ↗ KT115-230V / KT115UL-120V
Interior width 600 mm p.2 ↗ KT115-230V / KT115UL-120V
Interior height 455 mm p.2 ↗ KT115-230V / KT115UL-120V
Interior depth 355 mm p.3 ↗ KT115-230V / KT115UL-120V
Sound-pressure level 48 dB(A) p.3 ↗ KT115-230V / KT115UL-120V
Energy consumption at 25°C 75 Wh/h p.3 ↗ KT115-230V / KT115UL-120V

Operating Limits

  • Temperature range: 4...100 °C p.2 ↗
  • Max. heat compensation at 40°C: 100 W p.2 ↗
  • Permitted load: 100 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 240 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ KT Product series: cooling incubator with Peltier technology
3-5 p.1 ↗ 115 Model / size (interior volume 102 L)
suffix p.1 ↗ -230V Mains voltage variant 200...240 V 1~ 50/60 Hz, CEE 7/7 plug
suffix p.1 ↗ UL-120V Mains voltage variant 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
KT115-230V p.1 ↗ 9020-0313 200...240 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard model version
KT115UL-120V p.1 ↗ 9020-0314 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug, Standard Standard model version

Related models