Introduces the CC-205B circulator and its main features
Specpp. 1-2
Performance, temperature, electrical and physical data per DIN 12876
p. 2table, 3 rows
p. 1table, 36 rows
Technical data according to DIN 12876
Parameter
Value
max. ambient temperature
40 °C
from Serial-No.:
394677
1.2/20
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
45...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure pump
yes
max. delivery
27 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
22 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
Bath volume
4,8 l
Filling capacity
3 l
Width bath opening WxD
105x90 mm
Bath depth
150 mm
Height of bath opening
190 mm
Overall dimensions WxDxH **
178x337x355 mm
Net weight
9 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method chosen: 45–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.
Determines whether the unit can reach the required process temperatures with the available cooling utility.
Source
“Operating temperature range 45...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Performance
Heating power scales with supply voltage from 1.5 kW at 200 V up to 2.1 kW at 240 V, so the actual heat-up performance depends on site voltage.
Needed to verify the available electrical supply delivers the expected heating capacity.
Source
“Heating power at 240V 2,1 kW Heating power at 230V 2 kW Heating power at 220V 1,8 kW Heating power at 208V 1,6 kW Heating power at 200V 1,5 kW”
Installation
The unit requires a 200–240 V single/two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.
Defines the branch circuit and protection device sizing for safe installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
External branch circuit protection according to UL 489 is mandatory.
Dictates the type of upstream protective device that must be provided in the installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances of ±10% are permitted only if they do not act in opposite directions simultaneously.
Prevents installation on unstable grids where combined deviations could damage the unit.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
Ensures the correct plug is supplied for the destination country and avoids rework on site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
The unit needs a cooling water connection via a 3/8" cooling coil for tap water.
Requires planning a water supply and drain line at the installation location.
Source
“Cooling coil for (tap) water (3/8").”
Installation
The pump connection is M16x1 male, which must match the external circuit plumbing.
Determines the fittings and adapters needed to connect the external loop.
Source
“Pump connection M16x1 male”
Operation
The unit must be operated within an ambient temperature range of 5 °C to 40 °C.
Constrains the room conditions where the circulator can be installed and operated.
Source
“min. ambient temperature 5 °C”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.
Confirms the built-in safety cut-out that protects the bath and process from overheating.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance
The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar.
Used to check whether the pump can overcome the pressure drop of the external circuit.