  1. Overview p. 1
    Product introduction and general description of the Unistat 912w unit
  2. Spec p. 2
    Technical data according to DIN 12876 for the unit
  3. p. 2 table, 23 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Förderleistung bei 0,65 bar90 l/min
Delivery at 1,0 bar74 l/min
Delivery at 1,2 bar59 l/min
Delivery at 1,4 bar31 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C516 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,9 l
Filling capacity expansion tank6,5 l
Overall dimensions WxDxH **630x704x1565 mm
Net weight328 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply requirement400V 3~ 50Hz
max. current20 A
Fuse3x20A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S393805

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 912w (Order-No.: 1055.0003.01)

CharacteristicValue
Operating temperature range-90...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C7 kW
Cooling power with Thermooil at 200°C7 kW
Cooling power with Thermooil at 100°C7 kW
Cooling power with Ethanol at 0°C7 kW
Cooling power with Ethanol at -20°C7 kW
Cooling power with Ethanol at -40°C6 kW
Cooling power with Ethanol at -60°C3,5 kW
Cooling power with Ethanol at -80°C0,9 kW
Cooling power with Ethanol at -90°C0,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2 kg
CO2 equivalent4,3 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage1,1 kg
CO2 equivalent 2nd stage16,3 t
Circulation pump: MK pump max. delivery110 l/min
Circulation pump: MK pump max. delivery pressure1,5 bar

Значения приведены дословно; строки «Cooling power with Thermooil» и «Cooling power with Ethanol» объединены с соответствующими температурными точками, как напечатано в исходной таблице.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a maximum current of 20 A and 3x20 A fusing, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water connection is G1/2 male and requires a minimum differential pressure of 1 bar and a maximum pressure of 6 bar, so the facility water loop must meet these limits.

Ensures the available cooling water supply can provide adequate flow and pressure without overpressuring the unit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake between 5 °C and 20 °C, and below the condensation point the cooling water pipes must be insulated.

Prevents condensation and ensures rated cooling performance is achievable at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase, so performance must be derated for warmer facility water.

Affects selection when the available cooling water is warmer than the 15 °C reference condition.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The pump connection is M30x1.5 male with a Y-shaped bore per DIN 3863, so mating hoses or pipework must use ball socket DIN 3863 and sleeve nut DIN 3870.

Ensures correct mating fittings are procured for leak-free connection of the external circuit.

Source
“Pump connection M30x1,5 male”
Safety

The refrigerant R-452A has a GWP of 2141 and 2 kg charge (4.3 t CO2e), and the second stage R-23 has a GWP of 14800 and 1.1 kg charge (16.3 t CO2e), which matters for refrigerant regulations and leak management.

Drives compliance obligations, leak detection requirements and environmental reporting for the installed equipment.

Source
“Global Warming Potential (GWP) 2141”
Operation

The unit has a minimum filling capacity of 3.9 l and an expansion tank filling capacity of 6.5 l, so the external circuit volume and fluid charge must be planned within these limits.

Prevents underfilling or overflow and ensures adequate expansion volume for the temperature range.

Source
“min. filling capacity 3,9 l”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, so the selected fluid must remain below this limit across the operating temperature range.

Guards pump performance and avoids excessive pressure drop or cavitation with viscous fluids at low temperature.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit has IP20 protection and must be installed at ambient temperatures between 5 °C and 40 °C, so it needs a dry, protected location within that range.

Defines the environmental envelope and enclosure protection required at the installation site.

Source
“Degree of Protection IP20”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.

Prevents installation on a supply whose combined voltage and frequency deviation exceeds the permitted envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance

The equipment is EMC classified as Class A, Group 1 to EN55011, so it is intended for industrial environments and may require additional measures in residential areas.

Informs EMC planning and whether the installation location is appropriate for Class A equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Three-phase devices with current consumption less than 63 A are delivered with cable but without plug, so a suitable plug and connection must be provided on site.

Ensures the electrical contractor supplies the correct plug and final connection hardware.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit has a sound pressure level of 65 dB(A) ±4 dB(A), which must be considered for workplace noise exposure and location planning.

Affects siting and any noise mitigation needed in the operating environment.

Source
“sound pressure level +/- 4 dB(A) 65 dB(A)”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL, so the external circuit must be compatible with these protective requirements.

Confirms the safety concept and the need to set and verify overtemperature protection for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”