LC-40P SF Instruction Manual - Solvent Delivery Module

Technical Reference Summary

Covered Models & Document Scope

This manual (228-92873) covers the LC-40P SF solvent delivery module for Shimadzu SFC systems. It is used with the CO2 Cooling Unit and SFC-40P back pressure regulator. Optional parts include the LC-20AP tool kit (228-53062-41).

Section-by-Section Technical Breakdown

pp. 9
Overview and Features

The LC-40P SF is a solvent delivery module for Shimadzu SFC systems, delivering CO2 at high flow rates using a parallel double-plunger reciprocating pump. It supports pressures up to 44 MPa and flow rates from 10 to 150 mL/min. An automatic rinsing kit is standard to extend plunger seal life.

  • Intended for Shimadzu SFC systems with CO2 Cooling Unit and SFC-40P back pressure regulator.
  • Parallel double-plunger reciprocating pump.
  • Maximum pressure 44 MPa; flow rate range 10–150 mL/min.
  • Automatic rinsing kit standard.
pp. 10
Component Parts

Lists the parts included with the instrument, including the LC-40P SF unit, safety guideline, DVD-ROM, coolant bottle assembly, tubing panel, pump head cover, drain components, wrenches, and screws. Optional parts include the LC-20AP tool kit.

  • LC-40P SF main unit.
  • Safety Guideline (Booklet) 228-92890.
  • Instruction Manual/System Guide DVD-ROM 228-97175-41.
  • Coolant bottle ASSY 228-68438-41, coolant bottle holder 228-67401-41.
  • Inlet tubing panel 228-75673-41 with ghost trap.
  • Pump head cover 228-75699-41.
  • DRAIN, CTO 228-42206; drain tube 228-25162-03.
  • Wrenches: 11×13 (086-03047-08), 24×27 (086-03078, qty 2), hex 2.5 (086-03819).
  • Sub rear panel 228-75920 with screws.
  • Optional LC-20AP tool kit 228-53062-41.
pp. 11-15
Parts Identification and Function

Describes the front, behind front cover, right side, and back panels, including the display area, operation area, power switch, CO2 Cooling Unit, rinse solution bottle, thermo-sensor, coolant bottle, rinsing pump, pressure sensor block, check valves, ghost trap, and connectors.

  • Front: display area, operation area, power switch, front cover, handholds, CO2 Cooling Unit.
  • Behind front cover: rinse solution bottle (2-propanol 10%), thermo-sensor, coolant bottle (50% ethylene glycol), rinsing pump, leakage drain tubing, pump head cover, coolant pump outlet, pump head cooling block, pressure sensor block, check valves PREP OUT/IN, pump head, ghost trap, drain outlet.
  • Back: [REMOTE] connector, external input/output terminals, [PUMP PRESS] connector (0-5 MPa to 0-50 MPa in 10 steps), [FLOW METER] connector, fuse holders, power supply connector.
pp. 16-18
Display and Operation Area

The display area shows flow rate, pressure, and indicators. The operation area has 24 keys for controlling the pump, time program, valve, VP functions, and parameter entry.

  • Display area: flow rate (mL/min), pump indicator, prog.run indicator, remote control mode indicator.
  • Operation area keys: Display, Pump, Run, Valve, VP, Sleep, Delete, Edit, Clear, Back, Concentration, Function, Enter, numeric keys.
  • Pump key starts/stops pump; liquid CO2 shut-off valve opens/closes in concert.
  • Run key starts/stops time program.
  • Valve key opens/closes liquid CO2 shut-off valve; pump stops when valve closed.
pp. 19-22
Basic Operation Settings

Covers setting flow rate, maximum pressure limit, and minimum pressure limit. Ranges and steps are specified. Maximum pressure limit protects flow line components; minimum pressure limit prevents pumping during pressure drop.

  • Flow rate range: 0 to 150.00 mL/min, step 0.01 mL/min, initial 0 mL/min.
  • P.max range: 1.0 to 44.0 MPa (0-100 mL/min) or 1.0 to 33.0 MPa (100.01-150 mL/min), step 0.1 MPa, initial 10 MPa.
  • P.min range: 0.0 to 40.0 MPa (0-100 mL/min) or 0.0 to 30.0 MPa (100.01-150 mL/min), step 0.1 MPa, initial 0 MPa.
  • If pressure exceeds P.max, pumping stops automatically and alarm sounds.
  • If pressure drops below P.min for more than 1 minute, pumping stops and alarm sounds.
  • Setting P.min to 0 disables automatic stop on pressure drop.
pp. 24
Operation

Describes starting and stopping the pump, and the importance of starting the SFC-40P before operating the pump to avoid air delivery and plunger seal degradation.

  • Start SFC-40P operation before operating the pump.
  • Operating pump without SFC-40P running delivers air and degrades plunger seal.
  • When connected to system controller, unstable cooling in LC-40P SF causes system controller to stop pump.
pp. 26-27
Display Screens

Describes the types of screens: basic setting screen, auxiliary function setting screens, VP function screen, and time program edit screen, and navigation between them.

  • Initial screen displayed on power ON.
  • Basic setting screen sets flow rate, P.max, P.min.
  • Auxiliary function screens: parameter settings, file operation, control settings, system settings, monitor settings.
  • VP function screens: mobile phase monitor, product information, maintenance information, validation support, calibration support.
  • Time program edit screen for editing time programs.
pp. 28-40
Auxiliary Functions

Lists and describes auxiliary function groups: Parameter Settings (EVENT, COMP, PMIN T), File Operation (FILE NUM, FILE CPY, FILE DEL), Control Settings (P-SET, ZERO ADJ, MODE CHANG), System Settings (LOCAL, ADRS, KEY CLOSE, BRIGHT, EXT-S, SYS, S-PROT, LINK CHK, PRS-UNIT, RANGE, BEEP MODE), and Monitor Settings (MON TIME, MON ID, MON REV, CONDITION).

  • EVENT: controls output relays (0=OFF/OFF, 1=ON/OFF, 2=OFF/ON, 12=ON/ON).
  • COMP: compressibility compensation, default 3.00 (GPa)^-1 for liquid CO2; do not change unless necessary.
  • PMIN T: time ignoring minimum pressure limit, default 50.
  • FILE NUM: select file 0-9; up to 10 files stored.
  • P-SET: sets plunger stop position (1=left retracted, 2=right retracted, 3=intermediate).
  • ZERO ADJ: zero pressure sensor at atmospheric pressure.
  • MODE CHANG: switches constant flow/constant pressure; use constant flow for CO2; switch only when pump stopped.
  • LOCAL: 0=remote (initial), 1=local.
  • ADRS: address 1-12 for system controller; 85 for high-pressure gradient with two units.
  • KEY CLOSE: disables key input; release by pressing Clear while pressing Enter.
  • BRIGHT: display brightness 1-4.
  • EXT-S: 0=EVENT control, 1=relay1 start signal, 2=relay2 error signal, 3=both.
  • SYS: 1=independent/other system, 2=master pump in high-pressure gradient.
  • S-PROT: 0=off, 1=halves flow rate repeatedly to keep pressure below P.max without stopping.
  • LINK CHK: 0=off, 1=stops delivery if system controller disconnected during delivery.
  • PRS-UNIT: 0=MPa, 1=kgf/cm2, 2=bar, 3=psi.
  • RANGE: full scale (1 mV) = set value × 5.0 MPa; e.g., 1=5.0 MPa, 10=50.0 MPa.
  • BEEP MODE: 0=key+alarm, 1=alarm only, 2=all off.
  • MON TIME: displays elapsed time of time program.
  • MON ID: displays remote control ID (A-D).
  • MON REV: displays pump revolution counter; approx. 3,750,000 revolutions (approx. 600 hours at 5 mL/min, 10 MPa with 2-propanol) estimated seal replacement timing.
  • CONDITION: diagnoses delivery conditions; displays pressure increase/decrease and right/left pump head pressure decrease.
pp. 41-55
VP Functions

VP functions support validation and information display: Mobile Phase Monitor, Product Information, Maintenance Information, Validation Support, and Calibration Support.

  • Mobile Phase Monitor: MOBILE PHASE (remaining/set volume), ALARM LEVEL (warning when remaining volume below set %; 0% deactivates).
  • Product Information: SERIAL NUMBER, S/W ID.
  • Maintenance Information: TOTAL OP TIME, L SEAL DELIVERED, R SEAL DELIVERED, PART REPLACEMENT, MAINTENANCE LOG (20 records), ERROR LOG (10 records), OPERATION LOG (10 records).
  • Validation Support: DATE, TIME, MEMORY CHECK, PULSE CHECK, GE TEST PROGRAM.
  • Calibration Support: Input PASSWORD (5 digits, default 00000), FLOW COMP FACT, PRESS COMP FACT, SEAL REPLACEMENT (recommended 3840 L), INITIALIZE PARAM, CHANGE PASSWORD, CBM PARAMETER.
pp. 51-58
Time Program

Describes creating and executing time programs, including commands, screen description, creation flow, high-pressure gradient programs, deleting steps, and starting/stopping.

  • Time program commands and screen described.
  • High-pressure gradient programs supported.
  • Steps can be deleted.
  • Time program can be started and stopped.
pp. 61-63
Connection with System Controller or Workstation

Covers setting the instrument for connection to a system controller or workstation, basic parameters, and external input/output terminals.

  • Setting the instrument for remote control.
  • Basic parameters for system controller connection.
  • External input/output terminals and remote cable connection.
pp. 64-84
Maintenance

Covers periodic inspection and maintenance, replacing plunger seal, cleaning/inspecting plunger, replacing rinse seal, inspecting/cleaning check valves, inspecting line filter, replacing ghost trap, replacing fuse, and cleaning exterior.

  • Periodic inspection and maintenance list.
  • Plunger seal replacement procedure.
  • Plunger cleaning/inspection/replacement.
  • Rinse seal replacement.
  • Check valve inspection/ultrasonic cleaning.
  • Line filter inspection/replacement.
  • Ghost trap replacement/inspection.
  • Fuse replacement.
  • Exterior cleaning.
pp. 85
Specifications

Provides technical specifications for the LC-40P SF.

  • Maximum pressure 44 MPa.
  • Flow rate range 10–150 mL/min.
  • Parallel double-plunger reciprocating pump.
pp. 87
Maintenance Parts

Lists consumable and replacement parts.

  • Consumable parts list.
  • Replacement parts list.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Maximum pressure: 44 MPa. p.9 ↗ Ensure flow line components and columns are rated for at least 44 MPa. LC-40P SF
Flow rate range: 10 to 150 mL/min. p.9 ↗ Select flow rates within this range for stable delivery. LC-40P SF
Flow rate setting range: 0 to 150.00 mL/min, step 0.01 mL/min, initial 0 mL/min. p.19 ↗ Flow rate can be set precisely; initial value is zero. LC-40P SF
P.max range: 1.0 to 44.0 MPa (0-100 mL/min) or 1.0 to 33.0 MPa (100.01-150 mL/min), step 0.1 MPa, initial 10 MPa. p.19 ↗ Maximum pressure limit must be set according to flow rate range; protects column and flow lines. LC-40P SF
P.min range: 0.0 to 40.0 MPa (0-100 mL/min) or 0.0 to 30.0 MPa (100.01-150 mL/min), step 0.1 MPa, initial 0 MPa. p.19 ↗ Minimum pressure limit prevents pumping during pressure drop; setting 0 disables automatic stop. LC-40P SF
If pressure exceeds P.max, pumping stops automatically and alarm sounds. p.21 ↗ Set P.max to protect column and flow line components. LC-40P SF
If pressure drops below P.min for more than 1 minute, pumping stops automatically and alarm sounds. p.22 ↗ Set P.min to prevent air delivery when CO2 runs out or leakage occurs. LC-40P SF
Compressibility compensation default value: 3.00 (GPa)^-1 for liquid CO2. p.31 ↗ Do not change unless specific reason; affects pulsation. LC-40P SF
Pump revolution counter: approx. 3,750,000 revolutions (approx. 600 hours at 5 mL/min, 10 MPa with 2-propanol) estimated seal replacement timing. p.39 ↗ Use MON REV to monitor seal life. LC-40P SF
Recommended plunger seal replacement after pumping 3840 L. p.53 ↗ Set SEAL REPLACEMENT to 3840 L for maintenance scheduling. LC-40P SF
Recorder output full scale (1 mV) = set value × 5.0 MPa; e.g., set value 1 = 5.0 MPa, 10 = 50.0 MPa. p.37 ↗ Configure RANGE to match external recorder. LC-40P SF
[PUMP PRESS] connector output voltage can be adjusted from 0-5 MPa up to 0-50 MPa in 10 steps. p.15 ↗ Set pressure monitoring range for external device. LC-40P SF
Rinse solution: 2-propanol 10% in liquid form. p.14 ↗ Use correct rinse solution for plunger seal cleaning. LC-40P SF
Coolant: 50% ethylene glycol. p.14 ↗ Use correct coolant for pump head cooling. LC-40P SF
Start SFC-40P operation before operating the pump. p.24 ↗ Prevents air delivery and plunger seal degradation. LC-40P SF
Use only constant flow delivery mode for CO2 delivery. p.33 ↗ Do not use constant pressure mode for CO2. LC-40P SF
Switch delivery mode only when pumping stopped. p.33 ↗ Changing mode during pumping may cause incorrect operation. LC-40P SF
Default password for calibration support: 00000 (five digits). p.52 ↗ Required to access calibration functions. LC-40P SF
Address 85 for high-pressure gradient mode with two units. p.34 ↗ Set ADRS to 85 when using two LC-40P SF in high-pressure gradient. LC-40P SF
SYS setting: 1 for independent/other system, 2 for master pump in high-pressure gradient mode of two units. p.35 ↗ Only settings 1 and 2 work correctly. LC-40P SF

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Press Dec Right p.40 ↗ Air remains inside the right pump head; or there is a leak in the right pump head's seal; or the right pump head's inlet check valve or the left pump head's outlet check valve are not operating properly. Check for air, seal leak, or check valve malfunction as indicated. Section 3.5.7 CONDITION display
Press Dec Left p.40 ↗ Air remains inside the left pump head; or there is a leak in the left pump head's seal; or the left pump head's inlet check valve or the right pump head's outlet check valve are not operating properly. Check for air, seal leak, or check valve malfunction as indicated. Section 3.5.7 CONDITION display
Press Increase p.40 ↗ Pressure has been at least 10% higher than the level shown 5 minutes ago. May be normal during pump start/stop; investigate if persistent. Section 3.5.7 CONDITION display
Press Decrease p.40 ↗ Pressure has been at least 10% lower than the level shown 5 minutes ago. May be normal during pump start/stop; investigate if persistent. Section 3.5.7 CONDITION display