Agilent Technologies 5977A GC MSD
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 5977A GC MSD

Agilent Technologies 5977A GC MSD — Gas chromatography.

The 5977A GC MSD is a single-quadrupole mass-selective detector with a diffusion high-vacuum pump, rated for maximum recommended helium flow of 1.5 mL/min and maximum column id of 0.25 mm. It is not CI-capable; CI requires a turbo-pump sibling. It serves as the mass-selective detector in a GC/MS system for EI-mode analysis.

Full specifications

Core Performance & Functional Specifications
Gas Flow 2 mL/min
High Vacuum Pressure Only with Micro-Ion vacuum gauge installed.
Maximum Recommended Gas Flow 1.5 mL/min
Optimal He Column Flow 1 mL/min
Max column id mfr 0.25 mm
CI capability mfr No No
Inert ion sources available mfr Yes Yes
GC compatibility mfr 9000
Foreline pumps available mfr Pfeiffer Duo Duo
DIP capability (3rd party) mfr Yes Yes
High vacuum pump mfr Diffusion Diffusion
Turbo Pump Speed mfr Displays the turbo pump speed. speed.
Foreline Pressure mfr Displays the foreline pressure. pressure.
CI Reagent mfr Displays CI reagent gas and flow rate .
CI capability‡ mfr No No
DIP** capability (3rd party) mfr Yes Yes
Carrier gas purity minimum mfr 99.9995
JetClean default hydrogen flow mfr 0.13
PCI recommended reagent gas flow mfr 20
NCI recommended reagent gas flow mfr 40
Reagent gas line evacuation time mfr 6 minutes
Post-pumpdown thermal equilibration wait mfr 2 hours
CI bakeout time before tuning mfr 2 hours
Hydrogen dissipation wait after power failure mfr 10 minutes
MSD cool-down wait after power failure mfr 1 hour
Dimensions, Weight & Mechanical Design
Depth 54 cm
Foreline Pump Weight 11 kg
Height 41 cm
Weight 39 kg
Width 30 cm
MSD housing height mfr 41 cm
MSD housing width mfr 30 cm
MSD housing depth mfr 54 cm
Diffusion pump mainframe weight mfr 39 kg
EI/CI turbo pump mainframe weight mfr 46 kg
Electrical, Power & Drive Specifications
Primary fuse rating mfr 12.5 V
Operating Conditions, Environment & Protection
Sound pressure 70 dB
Standards, Safety & General Data
Standard EI turbo pump mainframe weight mfr 44 kg

Description

Construction and series positioning

Behavior under load and media limits

Installation, control and operation

Integration & Operating Notes

Process Limits
Keep total helium flow at or below 1.5 mL/min

The diffusion pump cannot handle higher gas loads: sum column flow and any reagent gas against the 1.5 mL/min recommended maximum (2 mL/min absolute). Exceeding these flows will likely cause spectral degradation and reduced sensitivity, so column dimensions must also respect the 0.25 mm id limit.

User Manual, Table 3, page 17
Process Limits
Cap hydrogen/methane flow at 50 mL/min total

When using hydrogen or methane as carrier or reagent gas, the combined flow to the MS must never exceed 50 mL/min. This is an explosion-safety limit, not a performance limit; configure and monitor flows accordingly, and after venting with hydrogen open the analyzer side plate at least 45 degrees for 10 minutes before powering on.

User Manual, page 26
Maintenance
Replace foreline pump oil every 6 months

Foreline pump oil must be replaced every 6 months (every 3 months if using ammonia reagent gas, though this model is not CI-capable). Check foreline pump fluid weekly and check for leaks monthly. Diffusion pump fluid and dry foreline pump seals are annual items; IDP-3 tip seal annually.

User Manual, page 143
Maintenance
Do not disturb the mass filter or HED insulator

The quadrupole mass filter requires no periodic maintenance and must never be cleaned or adjusted outside Agilent service; the HED ceramic insulator must never be touched. Wear a grounded antistatic wrist strap before opening the analyzer chamber to prevent ESD damage to the side board.

User Manual, page 145
Electrical
Use exact T12.5A, 250 V replacement fuses

Two primary fuses of type T12.5A, 250 V (part 2110-1398) are required. A blown primary fuse most likely indicates a problem with the foreline pump, so investigate the pump before replacing fuses and restarting.

User Manual, page 33

Selection & Integration Q&A

When should I choose this diffusion-pump 5977A over a turbo-pump sibling?

Choose the diffusion-pump model for EI-only work with helium flows ≤1.5 mL/min and columns ≤0.25 mm id. If you need CI (PCI/NCI), wider 0.53 mm columns, or flows up to 4 mL/min, select a turbo-pump sibling such as G7077B, G7078B, G7079B, or G7081B; CI is not available on diffusion-pump models.

What adjacent system parts are required or recommended for this MSD?

A Pfeiffer Duo foreline pump (11 kg) is used with this model. The high-vacuum gauge controller (G3397B) is recommended for EI to monitor vacuum, though mandatory only for CI. The MSD is compatible with 9000 GCs; ensure column id ≤0.25 mm and total helium flow ≤1.5 mL/min to match the diffusion pump capacity.

What mounting and duty-cycle limits apply from the manual?

Bench must support the 39 kg mainframe (41×30×54 cm) plus the 11 kg foreline pump, with access for weekly fluid checks. With hydrogen, total flow to the MS must never exceed 50 mL/min, and after venting open the side plate ≥45° for 10 minutes before power-on. Fasten the top thumbscrew finger-tight when using hazardous gas.

What maintenance schedule and consumables are required beyond base delivery?

Plan for foreline pump oil changes every 6 months, weekly fluid checks, monthly leak checks, annual diffusion pump fluid and dry foreline pump seal replacement, and annual IDP-3 tip seal. Consumables include Apiezon L high-vacuum grease for the side-plate O-ring and T12.5A, 250 V fuses (part 2110-1398, 2 required); these are not included in the base instrument.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
5977C Series MSD Operating Manual
Manual ·232 pp.