Nabertherm RHTH 120/600/..
Nabertherm RHTH 120/600/.. — Tube furnaces.
The RHTH 120/600/.. is a horizontal high-temperature tube furnace accepting tubes up to 120 mm outer diameter, with a 600 mm heated length and 1800 °C maximum temperature. Its 14.8 kW three-phase connection and 830 mm tube length distinguish it from the smaller RHTH 80/300/.. sibling. It serves as a high-temperature process stage for tube-contained samples requiring controlled atmosphere or vacuum.
Nabertherm RHTH Series — Model Comparison & Selection
| Model | Max temperature, °C | Weight, kg | Electrical connection |
|---|---|---|---|
| RHTH 120/600/.. Current | 1800 | 110 | 3-phase4 |
| RHTH 80/300/.. | 1700 | 90 | 3-phase4 |
Full specifications
| Electrical connection | 3-phase4 |
| Nominal power | 14.8 kW |
| Max. outer tube Ø mfr | 120 |
| Outer dimensions (W × D × H) mfr | 550 × 640 mm |
| Weight | 110 kg |
| Heated length mfr | 600 |
| Tube length mfr | 830 |
| Max temperature | 1800 °C |
Description
Construction and series positioning
- The RHTH 120/600/.. is the larger of the two listed RHTH models: it accepts tubes up to 120 mm outer diameter versus the RHTH 80/300/..'s smaller bore, and provides a 600 mm heated length.
- The furnace is configured for horizontal tube operation; the series title also permits vertical operation, but no vertical-specific mounting facts are provided for this model.
- Physical envelope: 550 × 640 mm (W × D), 110 kg. Tube length for this model is 830 mm, giving 115 mm of cold tube beyond each end of the 600 mm heated zone.
Thermal behavior
- Maximum temperature is 1800 °C. The constant-temperature zone is 200 mm at one stated condition and 300 mm at another; both values are retained as separate conditions rather than reconciled.
- No digitized curve points are available, so heat-up rate, cooling behavior and load-dependent uniformity cannot be stated from the verified facts.
Electrical and control
- Electrical connection is 3-phase, 4-wire; nominal power is 14.8 kW. No voltage value is given, so supply voltage must be confirmed against the nameplate before installation.
- Controller and process-data functions are covered only by a general Nabertherm controller document (series 500 and MyNabertherm app); no model-specific control configuration is verified for this furnace.
Documentation scope
- The attached PDF digests cover controllers, fiber-optics glass production, foundry furnaces and dental furnaces. None contains model-specific installation, connection or operating facts for the RHTH 120/600/.., so integration details below are limited to what the verified spec set supports.
Integration & Operating Notes
The verified facts state a 3-phase, 4-wire connection and 14.8 kW nominal power but give no voltage. The supply must be matched to the nameplate rating; connecting to an assumed voltage risks overcurrent or underpowered operation, and the 14.8 kW load requires a dedicated three-phase circuit sized for continuous furnace duty.
The 120 mm figure is the maximum outer tube diameter the furnace accepts. An oversized tube will not seat correctly in the furnace bore and can cause mechanical damage or unacceptable temperature distribution; tube selection must also account for the 830 mm tube length and the 600 mm heated zone.
The constant-temperature zone is listed as 200 mm under one condition and 300 mm under another. Workpieces requiring a defined uniform zone must be placed according to the condition actually used; treating the larger figure as universal can place samples outside the uniform region.
The available controller document covers Nabertherm series 500 and the MyNabertherm app, including process data storage and AMS2750F instrumentation types, but does not confirm which controller is fitted to this model. Control configuration and any recorder option must be specified at order.
Selection & Integration Q&A
Choose the 120/600/.. when the process needs tubes up to 120 mm outer diameter and a 600 mm heated length at up to 1800 °C. The RHTH 80/300/.. is the smaller sibling; select it only when its smaller bore and shorter heated length fit the workload, since the 120/600/.. provides the larger tube capacity.
A 3-phase, 4-wire circuit rated for 14.8 kW is required, with voltage confirmed from the nameplate since no voltage is stated. The process tube must not exceed 120 mm outer diameter and should be about 830 mm long to span the 600 mm heated zone with cold ends for connections.
The verified facts list only the furnace itself: no tube, controller configuration, recorder, thermocouple or process-gas fittings are confirmed as included. The controller document describes series 500 and MyNabertherm functions as options, so controller, recorder and tube accessories should be specified explicitly at order.