CC-415wl Immersion Circulator Datasheet

Contenido

  1. Overview p. 1
    Introduces the CC-415wl immersion circulator and its main features
  2. Spec p. 2
    Technical data per DIN 12876 including temperature and performance ratings
  3. p. 2 table, 15 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. cooling water pressure6 bar
Bath volume5 l
min. filling capacity4 l
Width bath opening WxD / bath depth120 x 110 / 150 mm
Height of bath opening565 mm
Overall dimensions WxDxH **410x480x764 mm
Net weight61 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current9 A
min. Fuse10A
max. Fuse16A
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S433624

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-40...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V1,6 kW
Heating power at 230V1,5 kW
Heating power at 220V1,3 kW
Cooling power at 100°C1,2 kW
Cooling power at 20°C1,2 kW
Cooling power at 0°C1 kW
Cooling power at -20°C0,6 kW
Cooling power at -30°C0,2 kW
Cooling power at -40°C0,05 kW
Refrigeration machineair- and water-cooled, CFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,38 kg
CO2 equivalent0,8 t
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C72 l/h
Consumption at water 15°C, flow 0°C66 l/h
Consumption at water 15°C, flow -40°C42 l/h
min. cooling water differential pressure3 bar

Order-No.: CC-415wl; 2018.0002.01

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires cooling water with a minimum differential pressure of 3 bar between inlet and outlet, so the facility water supply must be verified before installation.

Insufficient differential pressure will prevent proper cooling and may cause the unit to trip or underperform.

Source
“min. cooling water differential pressure”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.

Aggressive or incompatible water will corrode the circuit and void reliable operation.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

If the cooling water temperature is below the dew point, the cooling water pipes inside the unit must be insulated to prevent condensation.

Condensation can cause electrical faults and water damage; insulation is a mandatory installation measure.

Source
“If the cooling water temperature is below the dew point, it is essential to insulate the cooling water pipes in the unit.”
Operation

The unit is designed for cooling water temperatures between 5 °C and 20 °C; higher water temperature reduces cooling power and may require increased flow.

Operating outside this range will degrade performance and may exceed the unit's cooling capacity.

Source
“The temperature control unit is designed for cooling water temperatures between 5 °C and 20 °C.”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

This asymmetric tolerance rule must be respected when assessing site power quality to avoid damage or malfunction.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Missing plug specification at order time will delay commissioning or require field wiring.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant R-452A has a GWP of 2141 and a charge of 0.38 kg, corresponding to 0.8 t CO2 equivalent, which affects leak-check and F-gas obligations.

Operators must know the CO2 equivalent to comply with refrigerant leak-check and reporting regulations.

Source
“Global Warming Potential (GWP)”
Safety

The refrigerant is classified A1 with H280 hazard statement, indicating a non-flammable but pressurized gas hazard.

Personnel must be aware of the pressurized-gas hazard during handling and maintenance of the refrigeration circuit.

Source
“R-452A (A1, H280)”
Installation

The unit has IP20 degree of protection, so it must be installed in a dry indoor location protected from water ingress.

IP20 offers no water protection; installing in a wet or washdown area risks electrical failure.

Source
“Degree of Protection”
Installation

The ambient temperature must be between 5 °C and 40 °C; higher ambient temperature may reduce cooling capacity.

Site ambient conditions directly affect achievable cooling power and must be considered in selection.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The maximum cooling water pressure is 6 bar, so the supply must be regulated or limited to avoid overpressure damage.

Exceeding 6 bar can damage the cooling water circuit and internal components.

Source
“max. cooling water pressure”
Cumplimiento

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which has implications for residential or sensitive environments.

Class A equipment may cause interference in residential areas and may require additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The pump connection is M16x1 male and the cooling water connection is G1/2 male, so mating fittings must be procured accordingly.

Knowing the thread types prevents delays and ensures correct hose/pipe adaptation during installation.

Source
“Pump connection”
Maintenance

The maximum permissible kinematic viscosity of the pumped medium is 50 mm²/s, which limits the fluids that can be circulated.

Using a fluid with higher viscosity will overload the pump and reduce circulation performance.

Source
“max. permissible kin. viscosity”