CC-K15 Immersion Circulator

Contenido

  1. Overview p. 1
    Product identification and general description of the CC-K15 unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits per DIN 12876
  3. Spec p. 2
    Tabulated DIN 12876 technical parameters and values
  4. p. 2 table, 9 rows
  5. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
max. current refrigerated bath1,5 A
max. current immersion thermostat10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S379761

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 20°C0,25 kW
Cooling power at 0°C0,2 kW
Cooling power at -10°C0,12 kW
Cooling power at -20°C0,05 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,041 kg
CO2 equivalent0,0 t
Pressure / Suction pump max. delivery27 l/min
max. delivery pressure0,7 bar
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume15 l
Width bath opening WxD290x152 mm
Bath depth200 mm
Height of bath opening265 mm
Overall dimensions WxDxH **350x560x430 mm
Net weight28 kg
Power supply requirement220-240V 1~/2~ 50/60Hz

** см. сноску в оригинале (в тексте страницы пояснение не приведено)

Engineering Facts

14 facts · every claim linked to a page
Selection

The bath covers a working temperature range of -20 to 200 °C, so it suits both sub-ambient and high-temperature duties.

Defines whether the unit can meet the process temperature window before any other sizing is done.

Source
“Operating temperature range -20...200 °C”
Performance

Cooling capacity falls from 0.25 kW at 20 °C to only 0.05 kW at -20 °C, so low-temperature loads must be sized against the coldest point.

Prevents oversizing assumptions based on the 20 °C figure when the real duty is near the lower limit.

Source
“at 20°C 0,25 kW at 0°C 0,2 kW at -10°C 0,12 kW at -20°C 0,05 kW”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which drives room ventilation and ignition-source rules.

A3 flammability classification affects installation location, ventilation and site risk assessment.

Source
“R-290 (A3, H220)”
Installation

The unit is air-cooled, so it needs adequate air flow and the stated space clearances around the condenser.

Air-cooled condensers reject heat to the room and lose capacity if clearances are not respected.

Source
“Refrigeration machine air-cooled, natural refrigerant”
Installation

Ambient temperature must stay between 5 and 40 °C, and cooling capacity drops as ambient rises.

Sets the allowable plant-room conditions and warns that hot rooms erode the published cooling capacity.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The supply must be 220-240 V single or two-phase 50/60 Hz, with a minimum 10 A and maximum 16 A fuse.

Determines the electrical feed, breaker sizing and whether a neutral or second phase is available.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the unfavourable direction.

Guides acceptance of the site power quality before commissioning the compressor and electronics.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order.

Avoids a missing or wrong plug on delivery and rework at the installation site.

Source
“with power cable and country-specific plug (please specify when ordering)”
Operation

The pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction.

Needed to check whether the external circuit and its pressure drop can be served by the pump.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Operation

Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is fitted.

Prevents selection of an incompatible fluid that would damage the pump adapter and wetted parts.

Source
“Polyglycol is not permissible to be used as a heat transfer fluid”
Safety

The unit has adjustable overtemperature protection per DIN 12876 and safety classification III / FL.

Confirms the built-in safety cut-out required for unattended operation and flammable-liquid service.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Cumplimiento

The equipment is classified Class A, Group 1 for EMC, i.e. industrial disturbance class, not residential.

Class A equipment may not be suitable for residential environments and affects site EMC planning.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Cumplimiento

The unit is built to pressure equipment category Art. 4.3 PED and is IP20, so it needs a dry indoor location.

Defines the pressure-vessel compliance route and the ingress protection limits for the installation room.

Source
“Pressure equipment category Art. 4.3 PED”
Maintenance

A drain valve with cap is available as an accessory, which simplifies bath draining during service.

Worth ordering up front to make fluid changes and maintenance practical.

Source
“Drain valve with cap #6839”