Induztr
Resumen técnico Visor de documentos Modelos relacionados

KISS 205B Immersion Circulator

Contenido

  1. Overview p. 1
    Product introduction and main features of KISS 205B
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 37 rows

Technical data according to DIN 12876 — KISS 205B (Order-No.: 2004.0009.98)

CharacteristicValue
Operating temperature range45...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
Interface
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure / Suction pump
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connectionM16x1 male
Bath volume4,8 l
Filling capacity2,5 l
Width bath opening WxD105x90 mm
Bath depth150 mm
Height of bath opening190 mm
Overall dimensions WxDxH **178x337x355 mm
Net weight9 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S393791

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 45–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.

Determines whether the unit can reach the required process temperatures and which cooling option must be specified.

Source
“Operating temperature range 45...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz with a maximum current of 10 A and an external fuse between 10 A and 16 A.

Needed to size the electrical supply and branch circuit protection before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

An external branch circuit protection according to UL 489 is required.

Mandatory for compliant electrical installation in UL jurisdictions.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage tolerance is ±10% only if the frequency tolerance does not run in the opposite direction; for example, -10% voltage with +3% frequency is not allowed.

Prevents specifying the unit on a supply where combined voltage and frequency deviations exceed the allowed envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

The pump connection is M16x1 male, and the included accessories are hose connector NW12, blank plug, sleeve nuts, cover for bath bridge, cooling coil, and pump adaptor.

Determines the mating hose/adaptor needed for external circuit plumbing.

Source
“Pump connection M16x1 male”
Installation

The cooling coil is for (tap) water with a 3/8" connection.

Needed to plan the water cooling supply line and fittings.

Source
“Cooling coil for (tap) water (3/8").”
Operation

The plastic pump is not resistant to acetone and polyglycols, so water mixed with glysantine or ethylene glycol is recommended for freeze protection.

Prevents selecting an incompatible heat transfer fluid that would damage the pump.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

The optional Pt100 sensor connection can only display (not control) process temperature and is available only factory fitted at additional cost.

Must be ordered with the unit if external temperature monitoring is needed; it cannot be retrofitted in the field.

Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Confirms the safety concept required for unattended operation and flammable liquids classification.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The autostart function restarts the unit after a power failure.

Affects risk assessment for unattended processes and restart behavior after power loss.

Source
“Autostart function for power failure”
Installation

The degree of protection is IP20, and the ambient temperature must be between 5 °C and 40 °C.

Defines the permissible installation environment and protection against ingress.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Space requirements must be respected per the operating conditions.

Prevents installing the unit in a location without adequate clearance for operation and ventilation.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”