Unichiller 020T-H Technical Data Sheet

Contenido

  1. Overview p. 1
    Product name and general description of the chiller unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 33 rows

Technical data according to DIN 12876 — Unichiller 020T-H (Order-No.: 3024.0056.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power1,6 kW
Cooling power at 15°C2 kW
Cooling power at 0°C2 kW
Cooling power at -10°C1,5 kW
Cooling power at -20°C0,8 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:B
max. delivery30 l/min
max. delivery pressure3,8 bar
Delivery at 0,12 bar29 l/min
Delivery at 0,5 bar25 l/min
Delivery at 1,0 bar21 l/min
Delivery at 1,5 bar17 l/min
Delivery at 2,0 bar13,5 l/min
Delivery at 2,5 bar10 l/min
Delivery at 3,0 bar6,5 l/min
Pump connectionG3/4 male
min. filling capacity2,5 l
Volume of expansion2 l
Overall dimensions WxDxH **450x510x1230 mm
Net weight141 kg
Power supply requirement208V 2~ 60Hz
max. current14 A
Fuse2x16 A

** Overall dimensions WxDxH — как напечатано в исходном тексте.

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is intended only for externally closed applications, so the user circuit must be closed.

Prevents selecting this unit for open-loop or tank-venting duty where it would be unsuitable.

Source
“For externally closed applications.”
Installation

The unit requires a 208 V two-phase 60 Hz supply with 14 A maximum current and 2x16 A fusing.

Electrical supply and protection must be provisioned before installation.

Source
“Power supply requirement 208V 2~ 60Hz max. current 14 A Fuse 2x16 A”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Avoids assuming a ready-to-plug power connection and plans for local plug installation.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are explicitly not allowed.

Supply quality must be checked because some voltage/frequency combinations invalidate the tolerance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises.

Cooling performance must be derated for warmer installation rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit is rated IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.

Defines environmental limits and protection against ingress for the installation site.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant is R-449A with GHS classification H280 and GWP 1397, so refrigerant handling and leak response must follow safety rules.

Affects safety planning, refrigerant labeling, and environmental compliance.

Source
“R-449A (A1, H280)”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Overtemperature protection is a required safety function for the external circuit.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation

The pump connection is G3/4 male, so mating hoses and couplings must match this thread.

Prevents mismatch of process hoses and fittings during installation.

Source
“Pump connection G3/4 male”
Operation

The minimum filling capacity is 2.5 l and expansion volume is 2 l, which sets the minimum heat-transfer fluid charge.

Ensures enough fluid volume and expansion space for safe operation.

Source
“min. filling capacity 2,5 l Volume of expansion 2 l”
Performance

Cooling power falls from 2 kW at 0 °C to 0.8 kW at -20 °C, so low-temperature duty has reduced capacity.

Selection must account for reduced cooling capacity at the lowest operating temperatures.

Source
“at -20°C 0,8 kW”
Performance

Pump delivery drops with increasing pressure, from 29 l/min at 0.12 bar to 6.5 l/min at 3.0 bar.

External circuit pressure drop must be checked against the pump curve.

Source
“Delivery at 3,0 bar 6,5 l/min”
Cumplimiento

The equipment is EMC Class A Group 1, which is industrial disturbance classification and may require EMC precautions.

Class A equipment may need additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Space requirements around the unit must be respected according to the operating conditions.

Air-cooled units need clearance for airflow and service access.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”