Unichiller P015w-H Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data, performance and operating parameters
  3. Spec p. 2
    Detailed values according to DIN 12876 standard
  4. p. 2 table, 17 rows
  5. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -10°C48 l/h
Consumption at water 15°C, flow -20°C36 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight61 kg
sound pressure level +/- 4 dB(A)54 dB(A)
Power supply requirement208-230V 1~/2~ 60Hz
max. current12 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S618096

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P015w-H — Technical data according to DIN 12876

ParameterValue
Order-No.3051.0015.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at ambient temperature +20°C at 15°C1,5 kW
Cooling power at ambient temperature +20°C at 0°C1 kW
Cooling power at ambient temperature +20°C at -10°C0,7 kW
Cooling power at ambient temperature +20°C at -20°C0,3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,25 kg
CO2 equivalent0,3 t
Circulation pumpB
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C68 l/h
Consumption at water 15°C, flow 0°C60 l/h

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.

The site cooling water loop must be able to deliver at least 1 bar differential pressure and must not exceed 6 bar, otherwise the unit will not perform or may be damaged.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake temperatures between 5 °C and 20 °C, and pipes must be insulated if the water temperature can fall below the condensation point.

Cooling water outside this range invalidates the published cooling capacity and can cause condensation, so the facility water supply must be checked before installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Published cooling capacities are only valid at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, and higher cooling water temperatures reduce cooling power.

The engineer must derate the cooling capacity if site conditions differ from these reference values, otherwise the chiller may be undersized.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Water quality and treatment must be compatible with these materials to avoid corrosion and fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so water mixed with glysantine or ethylene glycol is recommended for freeze protection.

The heat transfer fluid must be selected to avoid attacking the pump; acetone and polyglycol-based fluids are unsuitable with the standard pump.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Cumplimiento

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the combination of -5 % voltage with +2 % frequency is not allowed.

The site power supply must be checked against this asymmetric tolerance rule, otherwise the warranty and safe operation may be compromised.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug, while three-phase devices below 63 A come with cable but no plug.

The correct plug type must be specified at ordering, and for three-phase units the customer must provide the plug and final connection.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The unit has degree of protection IP20 and must be installed in an environment between 5 °C and 40 °C ambient temperature.

IP20 offers no water protection, so the installation location must be dry and within the specified ambient temperature range.

Source
“Degree of Protection IP20”
Selection

The refrigerant is R-449A with GWP 1397 and a charge of 0.25 kg, giving a CO2 equivalent of 0.3 t.

This information is needed for F-gas reporting, leak-check obligations and refrigerant handling requirements.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Selection

The circulation pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with delivery dropping to 6 l/min at 2.0 bar.

The pump curve must be checked against the pressure drop of the external circuit to ensure the required flow is actually achieved.

Source
“max. delivery pressure 2,5 bar”
Installation

The minimum filling capacity is 3.8 l and the expansion volume is 1.7 l.

The external loop and bath must be filled with at least this volume, and the expansion volume limits how much thermal expansion the system can absorb.

Source
“min. filling capacity 3,8 l”
Operation

The unit is classified as pressure equipment category Art. 4.3 PED.

This determines the applicable pressure equipment compliance regime and documentation required for the installation.

Source
“Pressure equipment category Art. 4.3 PED”