Tab. 2
Parameter
Specification
Input: Voltage
100 ÷ 240 Vac (±10 %), 1-phase
Frequency
50 to 60 Hz
Power
800 VA maximum
Output for pump: Power during pump ramp-up
430 W maximum
Power normal (pump water cooled)
350 W maximum
Power normal (pump air cooled)
220 W maximum
Operating temperature
5 °C to +45 °C
Storage temperature
-20 °C to +70 °C
Fuse: Mains
2 x T10 AH (slow blow) 250 V 5x20 fuses
Serial communication (T-Plus kit)
RS232 cable with a 9-pin D type male connector and a 9-pin D type female connector, and T-Plus software (optional)
Power cable
With European or NEMA plug 3 meters long (optional)
Pump cable
5 meters long fixed cable
In compliance with norms
EN 61010-1 (2001) EN 61326-1 (2006) Emission level : Class B for Residential Application Immunity level : for Industrial Application
Max altitude
2000 m
Protection category
IP 20
Internal use only
Weight (both models)
6.3 kg (13.9 lbs)
Installation category
II
Pollution degree
2
Window Meanings
N.
Read/Write
Data Type
Description
Admitted Values
310 to 399
Reserved to Agilent service
400
R
A
CRC EPROM (QE)
QE8XXXX (where “XXXX” are variable)
402
R
A
CRC Param. (PA)
PA8XXXX (where “XXXX” are variables)
404
R
A
CRC parameter structure
XXXX
406
R
A
Program Listing code & revision
XXXX
407
R
A
Parameter Listing code & revision
XXXX
500 to 502
Reserved to Agilent service
503
R/W
N
RS 485 address
0 to 31 (default = 0)
504
R/W
L
Serial type selection
0 = RS 232 1 = RS 485 (default = 0)
Window Meanings
N.
Read/Write
Data Type
Description
Admitted Values
126
R/W
N
Vent valve opening delay expressed in 0.2 sec
0 to 65535 Corresponding to 0 to 13107 sec (Default = 15)
143
R/W
N
External Fan Configuration
0 = Always ON 1 = Automatic 2 = Serial (default = 2)
144
R/W
L
External Fan activation
0 = OFF 1 = ON (default = 0)
145
R/W
L
Purge Valve open/close (N.C.)
0 = close 1 = open (Default =1)
147
R/W
N
Vent Valve opening (Time expressed in 0.2 s)
0 to 65535 bit 0 = infinite min 0.2 s, Max 13107 s (default = 0)
155
R
N
Power Limit Applied expressed in W
157
R/W
L
Gas Load Type
0 = Ar 1 = N2 (default = 0)
Counters / Controller Data
Display
Description
CYCLE XXXX
Number of start/stop cycles
TIME XXXXXm
Time of last cycle (minutes)
P.LIFE XXXXXX
total operating time (hours) of the pump
MN
Model Number of the controller
SN
Serial Number
FW
Firmware version
PA
Parameter Listing version
Pressing once the COUNTERS pushbutton displays CYCLE, TIME, P.LIFE; pressing twice displays MN, SN; pressing a third time displays FW, PA.
Tab. 8
Description
Part number
Turbo-V 551 Rack Controller
X3501-64002
T-Plus communication software and serial cable kit
9699883
Turbo-V HT Rack Air Cooling kit
X3501-68001
Mains cable NEMA plug 3m long
9699958
Mains cable European plug 3m long
9699957
Turbo-V vent
X3501-68002
RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION
EUROPE
NORTH AMERICA
PACIFIC RIM
Fax
00 39 011 9979 330
1 781 860 9252
Fax Free
00 800 345 345 00
Toll Free
00 800 234 234 00
800 882 7426, Option 3
vpt-customercare@agilent.com
vpl-ra@agilent.com
Please visit our website for individual office information
http://www.agilent.com
Tab. 1
MESSAGE
DESCRIPTION
REPAIR ACTION
CHECK CONNECTION TO PUMP
Wrong connection between the pump and the controller.
Check connections between controller and pump. Press the START push-button twice to start the pump.
PUMP WAITING INTERLOCK
The interlock signal of J1 connector is activated by an interruption of the link between pin 3 and 9 or between pin 4 and 15 of P1 connector, or because the external interlock signal is open.
Reset the short circuit between pin 3 - 9 and between pin 4 - 15 of J1 connector, or close the external interlock signal.
FAULT: PUMP OVERTEMP.
The pump upper bearing temperature exceeded 60 °C.
Wait until the temperature decrease below threshold value. Press the START push-button twice to start the pump.
FAULT: CONTROLLER OVERTEMPERATURE
The controller internal temperature exceeded 65 °C.
Wait until the temperature decrease below threshold value. Check that the pump cable length has been set correctly. Press the START push-button twice to start the pump.
Error Messages
MESSAGE
DESCRIPTION
REPAIR ACTION
FAULT: TOO HIGH LOAD
In normal operation, the current drawn by the pump is higher than programmed.
Check that the pump rotor is free to rotate. Press the START push-button twice to start the pump.
FAULT: SHORT CIRCUIT
During normal functioning (after the start-up phase), a short-circuit condition has been detected in the power circuit.
Check connections and shortages between pump and controller. Press the START push-button twice to start the pump.
OVERVOLTAGE
A failure has occurred in the controller power supply section, or the controller has received a disturbance from the mains.
Press the START button twice to start the pump again. If the message is redisplayed, contact Agilent for maintenance.
Engineering Facts
14 facts · every claim linked to a page
Installation
The controller must be powered from a properly grounded 3-wire socket to avoid electric shock and meet CE requirements.
Grounding and approved cord/plug are mandatory for safe installation and CE compliance.
Source
“Use this power cord and plug in conjunction with a properly grounded power socket to avoid electrical shock and to satisfy CE requirements.”
Safety
The input power cable must be disconnected before any servicing because high voltage inside can be lethal.
Defines the lockout step required before maintenance work.
Source
“High voltage developed in the controller can cause severe injury or death. Before servicing the unit, disconnect the input power cable.”
Installation
The controller must be placed so that free air flows through its ventilation holes, whether used on a bench or in a rack.
Obstructed airflow causes overheating and overtemperature faults.
Source
“The Turbo-V controller can be used as a bench unit or a rack module, but it must be positioned so that free air can flow through the holes.”
Installation
The controller must not be installed in explosive, dusty, wet, or high-fire-risk environments.
Site selection must exclude hazardous and contaminated atmospheres.
Source
“Do not install or use the controller in an environment exposed to atmospheric agents (rain, snow, ice), dust, aggressive gases, or in explosive environments or those with a high fire risk.”
Operation
Operating ambient must stay between 0 °C and +45 °C with 0–95 % non-condensing humidity.
Exceeding the ambient range risks overtemperature trips and reduced lifetime.
Source
“temperature: from 0 °C to +45 °C;”
Operation
The pump must never be run unless its inlet is connected to the system or blanked off.
Prevents injury and equipment damage during commissioning.
Source
“Never operate the Turbo-V pump if the pump inlet is not connected to the system or blanked off.”
Maintenance
The controller itself requires no maintenance, but any system work must be done by authorized personnel only.
Defines who may service the unit and that no routine controller maintenance is needed.
Source
“The Turbo-V 551 Rack Controller does not require any maintenance. Any work performed on the system must be carried out by authorized personnel.”
Operation
After an automatic fault stop, START must be pressed once to reset and a second time to restart the pump.
Operators must know the two-press reset/restart sequence after faults.
Source
“If the pump has been stopped automatically by a fault, this push-button must be pressed once to reset the controller and a second time to restart the pump.”
Operation
The interlock is satisfied only when pins 3–9 and 4–15 of J1 are shorted or the external interlock is closed.
Wiring of the interlock loop is required for the pump to start.
Source
“Reset the short circuit between pin 3 - 9 and between pin 4 - 15 of J1 connector, or close the external interlock signal.”
Operation
A pump upper-bearing temperature above 60 °C triggers a pump overtemperature fault.
Cooling and thermal design must keep bearing temperature below this threshold.
Source
“The pump upper bearing temperature exceeded 60 °C.”
Operation
A controller internal temperature above 65 °C triggers a controller overtemperature fault, and the pump cable length setting must be verified.
Rack airflow and correct cable-length configuration are needed to avoid this trip.
Source
“The controller internal temperature exceeded 65 °C.”
Installation
Remote I/O and serial cables longer than 3 m must be shielded.
Shielding is required for EMC when routing long signal cables.
Source
“All connecting cables for remote I/O and serial connections, if longer than 3 m, must be of the shielded type.”
Installation
Serial and I/O cables must be shielded, no longer than 30 m, with the shield bonded to the connector case at least on the controller side.
Sets the maximum cable run and shield termination for reliable, EMC-compliant wiring.
Source
“A shielded cable of 30 m maximum length has to be utilized for both serial and I/O port connections.”
Operation
Purge and vent valves are normally closed, so they stay closed on power failure.
Fail-safe valve behavior must be considered in vent/purge system design.
Source
“The two valves are Normally Closed (N.C.) so if a power fail occurs, the valves will remain closed.”