CC-410wl Immersion Circulator Datasheet

Contenido

  1. Overview p. 1
    Introduces the CC-410wl immersion circulator and its main features
  2. Spec pp. 1-2
    Lists operating data, dimensions, and performance parameters per DIN 12876
  3. p. 2 table, 14 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
max. cooling water pressure6 bar
Bath volume22 l
min. filling capacity17 l
Bath capacity with displacement rack8,5 l
Width bath opening WxD / bath depth280 x 280 / 200 mm
Height of bath opening520 mm
Net weight71 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current15,5 A
Fuse16 A
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S574279

Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 3 bar differential pressure between cooling water inlet and outlet.

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V3 kW
Heating power at 230V3 kW
Heating power at 220V2,7 kW
Cooling power at 100°C0,8 kW
Cooling power at 20°C0,8 kW
Cooling power at 0°C0,8 kW
Cooling power at -20°C0,5 kW
Cooling power at -30°C0,15 kW
Cooling power at -40°C0,1 kW
Refrigeration machineair- water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,25 kg
CO2 equivalent0,0 t
UN-numberUN 3358
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
min. cooling water differential pressure3 bar

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers a very wide working range from -45 to 200 °C, so it can serve both deep-freeze and high-temperature applications with one device.

Defines whether the circulator suits your process temperature window before purchase.

Source
“Operating temperature range -45...200 °C”
Performance

Cooling capacity collapses from 0.8 kW at 0 °C to only 0.1 kW at -40 °C, so deep-cold heat loads must be sized against the low-temperature figures.

Prevents oversizing expectations of cooling power at the cold end of the range.

Source
“at -40°C 0,1 kW”
Installation

The cooling water circuit needs at least 3 bar differential pressure and tolerates up to 6 bar, which dictates the site water supply and any booster pump.

Cooling water supply must be verified before installation; insufficient pressure degrades cooling.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water must be between 5 and 20 °C, and pipes must be insulated if water is below dew point to avoid condensation.

Determines cooling water source selection and condensation protection measures on site.

Source
“The temperature control unit is designed for cooling water temperatures between 5 °C and 20 °C.”
Safety

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, so the installation location must respect flammable-refrigerant rules.

Flammability class A3 affects room classification, ventilation and safety planning.

Source
“R-1270 (A3, H220)”
Cumplimiento

The unit is classified Class A, Group 1 for EMC, meaning it is industrial-grade equipment and may cause interference in residential environments.

Affects where the device may legally be installed and EMC planning in the lab.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Voltage and frequency tolerances are coupled: ±5% voltage only with ±2% frequency, and a -5% voltage with +2% frequency combination is not allowed.

Guides power supply quality checks and generator/UPS compatibility at the installation site.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase devices are delivered with a country-specific plug, while three-phase units below 63 A come with cable but no plug.

Determines what electrical termination hardware the installer must provide.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The pump delivers up to 25 l/min at 0.7 bar and also offers suction up to 18.5 l/min at 0.4 bar, enabling both pressure and suction circulation.

Needed to match external loop pressure drop and decide on pressure vs. suction operation.

Source
“max. delivery pressure (suction) 0,4 bar”
Operation

Maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, limiting fluid choice at low temperatures.

Prevents selecting a fluid too viscous for the pump at the cold end of operation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The device requires ambient temperature between 5 and 40 °C, so the room climate must be controlled.

Room conditions affect cooling performance and must be checked during site planning.

Source
“max. ambient temperature 40 °C”
Selection

Bath volume is 22 l with a minimum filling of 17 l, and only 8.5 l with a displacement rack, which affects fluid cost and thermal response.

Determines thermofluid quantity, bath dynamics and accessory selection.

Source
“Bath capacity with displacement rack 8,5 l”
Operation

The unit is IP20, so it must be kept dry and free of water ingress in the installation area.

Restricts placement near splashes or washdown areas.

Source
“Degree of Protection IP20”
Maintenance

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion and fouling of the circuit.

Source
“Please use suitable cooling water.”