Minichiller 800w-H OLÉ Technical Data

Contenido

  1. Overview p. 1
    Product name and cover page for the Minichiller unit
  2. Spec p. 2
    Technical data table according to DIN 12876
  3. p. 2 table, 10 rows

Technical data according to DIN 12876

ParameterValue
max. cooling water pressure6 bar
min. filling capacity2,8 l
expansion tank2,2 l
Overall dimensions WxDxH **280x490x424 mm
Power supply requirement208-240V 1~/2~ 50/60Hz
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/23

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is water-cooled, so a cooling water supply with at least 3 bar differential pressure and 5–20 °C inlet temperature must be available at the installation site.

Without adequate cooling water pressure and temperature the specified cooling capacity cannot be achieved.

Source
“min. cooling water differential pressure”
Installation

Cooling water connections are G1/2 male, so mating hoses/fittings must be G1/2 female.

Ensures correct mechanical connection and avoids adapters or leaks.

Source
“Cooling water connection”
Installation

The process pump connection is M16x1 male, so process hoses must match this thread.

Prevents mismatch with process loop fittings and ensures leak-free circulation.

Source
“Pump connection”
Operation

The refrigerant is R-290 (A3, flammable), so ignition sources and ventilation must be considered in the installation area.

A3 refrigerant requires attention to safety classification and local flammable-refrigerant regulations.

Source
“R-290 (A3, H220)”
Operation

Cooling water inlet must stay between 5 °C and 20 °C; below the dew point the cooling water pipes inside the unit must be insulated.

Operating outside this range or without insulation causes condensation and reduced performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Sizing must account for actual site cooling water temperature, not only the 15 °C reference.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Selection

The plastic pump is not resistant to acetone and polyglycols, so these media must be avoided or a more resistant pump specified.

Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than using pure water below 0 °C.

Avoids freezing and damage in the cooling circuit when operating at sub-zero temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination -5% voltage with +2% frequency is not allowed.

Site power quality must be checked; some voltage/frequency combinations are outside the permitted envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is IP20, so it must be installed in a dry, protected location and not exposed to water spray.

IP20 limits installation to indoor, non-wet environments.

Source
“Degree of Protection”
Installation

Ambient temperature must be between 5 °C and 40 °C for proper operation.

Installation room conditions affect cooling performance and reliability.

Source
“max. ambient temperature”
Installation

The unit requires 208–240 V single/two-phase 50/60 Hz supply; three-phase devices below 63 A are supplied with cable but without plug.

Electrical preparation must include a suitable plug and supply matching the delivered cable configuration.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Maintenance

Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.

Guides water quality and treatment to avoid corrosion and scaling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

Confirms built-in safety limiting for unattended operation and selection against process risk.

Source
“With adjustable overtemperature protection according to DIN 12876.”