Induztr
Resumen técnico Visor de documentos Modelos relacionados

Unichiller 050w-H Technical Datasheet

Contenido

  1. Overview p. 1
    Product introduction and general description of the chiller unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters
  3. Spec p. 2
    Detailed specifications according to DIN 12876 standard
  4. p. 2 table, 16 rows
  5. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -20°C137 l/h
min. cooling water differential pressure0,4 bar
max. cooling water pressure6 bar
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1050 mm
Net weight290 kg
sound pressure level +/- 4 dB(A)62 dB(A)
Power supply (3 Phase)460V 3~ 60Hz
max. current (3 Phase)10,5 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S474338

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 050w-H (Order-No.: 3038.0061.01)

CharacteristicValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C
at 100°C5 kW
at 20°C5 kW
at 0°C4,6 kW
at -10°C3,4 kW
at -20°C2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,85 kg
CO2 equivalent1,2 t
Circulation pumpE1
max. delivery56 l/min
max. delivery pressure3,8 bar
Delivery at 0,3 bar53 l/min
Delivery at 0,5 bar51 l/min
Delivery at 1,0 bar46 l/min
Delivery at 1,5 bar40 l/min
Delivery at 2,0 bar34 l/min
Delivery at 2,5 bar27 l/min
Delivery at 3,0 bar18 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C240 l/h
Consumption at water 15°C, flow 0°C216 l/h
Consumption at water 15°C, flow -10°C167 l/h

Строка «Cooling power at ambient temperature +20°C» является заголовком группы для последующих строк at 100°C … at -20°C; значение в ней пустое.

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 460 V 60 Hz supply with a 3x16 A fuse, so the site must provide a suitable branch circuit.

Determines electrical infrastructure and protection sizing before installation.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

External branch circuit protection according to UL 489 is mandatory, so a listed breaker must be included in the installation.

Affects electrical panel design and compliance for North American installations.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices drawing less than 63 A are supplied with cable but without a plug, so the customer must provide the final connection.

Clarifies what is in the scope of delivery and what the installer must supply.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation conditions occur.

Defines the cooling water supply requirements and condensation risk at the site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The cooling water circuit must provide at least 0.4 bar differential pressure and must not exceed 6 bar.

Sets the acceptable cooling water supply pressure window for correct operation.

Source
“min. cooling water differential pressure 0,4 bar”
Installation

The unit needs at least 18 l filling capacity and has 48 l expansion volume, which matters for loop design and fluid charge.

Informs the required external loop volume and expansion tank sizing.

Source
“min. filling capacity 18 l”
Installation

The chiller is water-cooled and requires a cooling water flow of 240 l/h at 15 °C water and 20 °C flow temperature.

Determines cooling water supply capacity needed at the installation site.

Source
“Consumption at water 15°C, flow 20°C 240 l/h”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.

Guides water quality and material compatibility to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C, so the room must stay within this range.

Ensures the installation location meets the environmental operating envelope.

Source
“min. ambient temperature 5 °C”
Operation

Cooling power drops as cooling water temperature rises and flow rate may need to increase, so performance must be derated accordingly.

Prevents overestimating cooling capacity when site water is warmer than 15 °C.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation

Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in opposite directions.

Defines acceptable supply quality limits for reliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant R-449A is classified A1 with GHS H280, and the unit contains 0.85 kg with 1.2 t CO2 equivalent.

Informs refrigerant safety classification and environmental reporting obligations.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Safety

The unit has safety classification III / FL according to DIN 12876 and adjustable overtemperature protection.

Confirms the safety category relevant for flammable liquids and protection design.

Source
“Safety classification III / FL”
Cumplimiento

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which affects EMC planning.

Determines EMC environment suitability and possible interference measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”