Sound Pressure Levels (EN ISO 2151:2008) (ISO 4871:1996) (ISO 11201:2010)
59 ± 2 dB(A)
Vibration level at inlet (per ISO 10816-1)
6.3 mm/sec
Installation
Indoor use, Equipment Class 1, Installation Category II, Pollution Degree 2
Altitude
2000 m (6562 ft)
Safety
Machinery Directive 2006/42/EC: 2006 EN 1012-2:1996+A1: 2010 EN ISO 12100:2012 Low Voltage Directive 2006/95/EC: 2007 EC 61010-1 Ed. 3.1 b:2018 CSA certified for US and Canada CE, ROHS
* If the pump is new or has been exposed to atmosphere for an extended period of time, it is possible that water can be present which can affect the pump’s ability to achieve ultimate pressure. To remove water from the pump and system, activate the gas ballast using the procedure in the gas ballast paragraph on page 164. Ultimate pressure will vary for different configuration and RPM setting. ** Mains supply voltage fluctuations shall not exceed ±10% of the nominal voltage.
Table 2 Full Load Motor Current, Amp
1 phase motor
220V
230V
50Hz
4.65
4.60
Table 7 Accessories
Part Number
Description
S4807001
Exhaust Extension
PTS600EXFIL
Exhaust Filter Kit
PTSPURGEKIT
Purge Kit (contains a flow meter and all necessary valving and tubing.)
Table 6 Troubleshooting Chart
Problem
Possible Causes
Corrective Actions
Will not start.
Motor thermal protector open
Allow the pump to cool. Identify the cause of overload.
Excessive voltage drop in the power cord.
Check the size and length of the cable.
Defective motor
Inspect. Contact Agilent.
Pump jammed
Inspect. Contact Agilent.
Poor ultimate pressure
System leak
Locate and repair the leak.
Water vapor in the pump or system
Flush pump with air or dry nitrogen. Activate the Gas Ballast.
Water in the pump
Activate the Gas Ballast.
Gas ballast plugged
Replace breather vent. Contact Agilent.
Solvent in the pump
Activate the Gas Ballast. Install trap or filter.
Worn tip seal
Replace tip seals.
Poor conductance to pump
Replumb with shorter and/or larger diameter tubing.
Low pumping speed.
Replace tip seals.
Hammering noise
Pump overheated
Check the ventilation to pump. Check ambient temperature.
Check the ambient temperature
Check the intake screen. Flush pump. Disassemble the pump head and inspect.
Debris in pump
Check intake screen. Flush pump. Disassemble pump head and inspect.
Pump runs intermittently
Motor thermal protector is cycling open and closed. *
Allow the pump to cool. Identify the cause of the overload. Check for proper operation of the fan. Ensure there is adequate space for air to circulate at inlets.
* Сноска в тексте не приведена.
Health and safety — The product has been exposed to the following substances: (by selecting ‘YES’ you MUST complete the table to the right)
Substances (please refer to MSDS forms)
Trade name
Chemical name
Chemical Symbol
CAS Number
Toxic
Harmful
Corrosive
Reactive
Flammable
Explosive (*)
Radioactive (*)
Biological (*)
Oxidizing
Sensitizer
Other dangerous substances
* Agilent will not accept delivery of any product that is exposed to radioactive, biological, explosive substances or dioxins, PCB’s without written evidence of decontamination.
Engineering Facts
12 facts · every claim linked to a page
Installation
The pump starts immediately when the power cord is plugged in, so the outlet must be treated as the on/off control and kept accessible.
Prevents unexpected startup during installation and requires an accessible disconnect means.
Source
“Pump will turn on immediately upon plugging the power cord to the electrical outlet.”
Installation
The pump is rated for indoor use only in clean, dry environments and is not certified for outdoor operation.
Rules out outdoor siting and requires a protected indoor installation location.
Source
“This pump is not designed or certified for operation outdoors.”
Installation
A minimum 20 cm clearance on all sides is required for ventilation, maintenance access and access to the power cord disconnect.
Sets the minimum installation envelope and prevents overheating from blocked airflow.
Source
“This pump requires a clear space of at least 20 cm (8 in.) on all sides for ventilation,”
Operation
Operating ambient must stay within +5 °C to +40 °C and 5–95% non-condensing relative humidity.
Defines the allowable site conditions for reliable operation and warranty compliance.
Source
“Temperature: +5 °C to +40 °C (41 °F to 104°F)”
Operation
Adsorbed water vapor is the most common cause of poor initial vacuum performance, and gas ballast may need 30 minutes to 24 hours to reach full performance.
Prevents misdiagnosing a new or long-idle pump as faulty during commissioning.
Source
“it can typically take between ~30 minutes to ~24 hours of gas ballast activation to achieve the full vacuum performance.”
Installation
The pump weighs about 33 kg unpacked, so proper lifting aids and techniques are mandatory to avoid injury.
Drives the need for lifting equipment and two-person handling during installation.
Source
“wiegen ohne Verpackung ca. 33 kg (70 Pfund).”
Installation
When unpacking, the pump must be lifted with a hoist hooked to the eyebolt, and the hoist must be rated for 100 kg.
Specifies the minimum lifting capacity required for safe unpacking and positioning.
Source
“Das Hebezeug sollte eine Tragfähigkeit von 100 kg haben.”
Cumplimiento
The pump is configured for 220–230 V 50 Hz only and cannot be run at 60 Hz or at 100–115 V without damage.
Confirms supply compatibility before selection and prevents destructive miswiring.
Source
“cannot be configured to operate at either 60hz or at low voltage, 100V-115V.”
Safety
The product must be grounded; the supplied cord has a grounding wire and plug that must go into a properly grounded outlet.
Grounding is mandatory for electrical safety and reduces shock risk on short circuit.
Source
“This product should be grounded.”
Operation
Exhaust back pressure must not exceed 1.5 psig when an exhaust hose or muffler is used.
Sets the maximum allowable exhaust restriction to avoid overloading the pump.
Source
“排気抵抗によって背圧が1.5 psigを超えることがないようにしてください。”
Maintenance
Before servicing or disconnecting the pump, wait for it to stop completely and cool to room temperature because of sharp edges and hot surfaces.
Prevents burns and cuts during maintenance and system disconnection.
Source
“wait for the pump to come to a complete stop and cool down to room temperature.”
Safety
Do not block the cooling vents; a pump surface temperature above 80 °C is potentially damaging and requires shutdown and inspection.
Gives a concrete over-temperature limit for safe operation and troubleshooting.
Source
“A pump surface temperature in excess of 80 °C (176 °F) is potentially damaging.”