Turbo-V 81-T Turbomolecular Pump Instruction Manual

Contenido

  1. Overview pp. 189-192
    Safety notes, product description and general warnings for the pump
  2. Procedure pp. 193-194
    Pre-installation checks, unpacking and site preparation instructions
  3. Procedure pp. 195-198
    Mechanical and electrical installation steps and precautions
  4. Procedure pp. 197-198
    Routine maintenance, cleaning and service procedures
  5. Spec pp. 199-218
    Performance curves, dimensions, specifications and operating parameters
  6. Procedure pp. 209-210
    Instructions for fitting the inlet screen accessory
  7. Config. pp. 219-220
    Vent valve and accessory options with ordering information
  8. Config. pp. 221-222
    System layout showing pump connections and accessories
  9. Procedure pp. 223-224
    Connection details for the high vacuum flange
  10. Procedure pp. 225-226
    Connection details for the fore-vacuum pump
  11. Ordering pp. 227-228
    Ordering codes and lists of accessories and spare parts
  12. p. 204 table, 17 rows
  13. p. 205 table, 13 rows
  14. p. 227 table, 21 rows
  15. p. 228 table, 4 rows
  16. p. 195 table, 1 rows

Tab. 2 Technical Specification

ParameterValue
Pumping speed With ISO 63 or CFF 4.5”N2: 77 l/s He: 65 l/s H2: 50 l/s
Pumping speed With KF 40 NW or CFF 2.75”N2: 50 l/s He: 56 l/s H2: 46 l/s
Compression ratioN2: >7 x 106 He: 5 x 103 H2: 3 x 102
Base pressure * (with recommended forepump)mechanical: 5 x 10-9 mbar (3.8 x 10-9 Torr) diaphragm: 5 x 10-8 mbar (3.8 x 10-8 Torr)
Inlet flangeISO 63 KF 40 NW CFF 4.5” O.D. CFF 2.75” O.D.
Foreline flangeKF16 NW
Rotational speed1350 Hz (max)
Start-up time< 60 seconds
Cooling requirementsNatural air convection Forced air or water optional
Recommended forepumpmechanical: Agilent DS 42 – DS 102 dry pump: Agilent SH 100
Operating positionany
Coolant waterflow: 10 l/h (0.05 GPM) temperature: + 15 °C to + 35 °C pressure: 2 to 4 bar (30 to 60 Psi)
Operating ambient temperature with natural air convection+5 °C to +25 °C
Operating ambient temperature with forced air cooling or water+5 °C to +35 °C
Bakeout temperature120 °C at inlet flange max. (CF flange) 80 °C at inlet flange max. (ISO flange)
Noise level≤ 45 dB (A) at 1 meter
Compliance with:UNI EN 292-1 UNI EN 292-2

* with recommended forepump

Technical Specification

ParameterValue
EN-CENELEC 55011
IEC 1000-4-2 (ex 801-2)
IEC 1000-4-3 (ex 801-3)
IEC 1000-4-4 (ex 801-4)
EN 61010-1 (IEC 1010-1)
EN 1012-2
Storage temperature-20 °C to +70 °C
Input60 Vac, three phase, 1350 Hz (max) 50 W max
Lubricantpermanent lubrication
Installation categoryII
Pollution degree2
Storage temperature- 20 °C to + 70 °C
Weight kg (lbs):ISO 63: 1.83 (4.03) KF 40: 1.80 (3.97) CFF 4.5”: 2.62 (5.78) CFF 2.75”: 2.59 (5.71)

* (According to standard DIN 28 428, the base pressure is that measured in a leak-free test dome, 48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange and using the recommended pre-vacuum pump)

Tab. 3 Accessories and Spare Parts

DESCRIPTIONPART NUMBER
Inlet screen, DN 409699309
Inlet screen, DN 639699300
Heater band 220 V9699801
Heater band 120 V9699802
Metallic water cooling kit9699823
Plastic water cooling kit9699824
Air cooling kit (0.5 m cable)9699290
Air cooling kit extension cable (5 m)9699940
Pump extension cable9699942
Vibration damper DN 639699375
Vibration damper CFF 4.5"9699376
Mechanical pump DS 1029499315
Mechanical pump DS 429499309
Dry scroll SH 100SH01001 UNIV
Dry scroll SH 110SH01101 UNIV
Rack controller 81-AG base9698988
Rack controller 81-AG RS232/4859698989
Rack controller 81-AG Profibus9698990
PCB 24 V9699538
Turbo-V 81-AG Navigator Controller 24 Vdc9698995
Turbo-V 81-AG Navigator Controller 100-240 Vac9698996

Tab. 4 Vent valve model selection table

CONTROL UNIT MODELVENT VALVE PART NUMBER
Turbo-V 81-AG Navigator Controller (cable 0.7 m) (any version)N.O. 9699844
Turbo-V 81-AG Rack Controller (cable 0.7 m) (any version)N.O. 9699844
Controller TV70 Navigator (P/N 9698970, 9698971)9699834
Turbo-V 70 Rack controller (P/N 9699405, 9699505)9699843

Tab. 1

FLANGEFIXING DEVICEN.FIXING TORQUE
ISO 63M10 clamps422 Nm

Engineering Facts

14 facts · every claim linked to a page
Installation

The protective cap and adhesive must remain in place until the pump is actually connected to the vacuum system, preventing contamination during handling.

Removing the cap early exposes the pump interior to dust and moisture, which degrades ultimate vacuum and can damage bearings.

Source
“Do not remove the adhesive and protective cap before connecting the turbopump to the system.”
Operation

Operating ambient temperature is limited to +5 °C to +35 °C, and relative humidity must be 0–95 % non-condensing.

Exceeding these limits can overheat the pump or cause condensation, leading to electrical faults or bearing damage.

Source
“temperature: from +5 °C to +35 °C (see the diagram pressure-temperature in the appendix “Technical Information”)”
Installation

The pump can be mounted in any orientation but must be fixed to a counter-flange able to withstand 250 Nm torque around its axis.

Insufficient fixing torque risks the pump rotating or detaching under rotor crash torque, causing equipment damage or injury.

Source
“Fix the Turbo-V 81-T in a stable position connecting the inlet flange of the turbopump to a fixed counter-flange capable of withstanding a torque of 250 Nm around its axis.”
Installation

The Turbo-V 81-T cannot be fixed by means of its base; it must be supported via the inlet flange connection.

Attempting base mounting will not provide adequate support and may lead to vibration, misalignment, or pump damage.

Source
“The Turbo-V 81-T cannot be fixed by means of its base.”
Operation

During bakeout, inlet flange temperature must not exceed 120 °C for ConFlat or 80 °C for ISO flanges, and water cooling must always be used.

Exceeding these temperatures can damage the pump's internal components or seals; water cooling is required to protect the pump during bakeout.

Source
“While heating the vacuum chamber, the temperature of the inlet flange must not exceed 120 °C for a ConFlat flange and 80 °C for a ISO flange.”
Operation

Never operate the pump with the inlet flange open to atmosphere; it must be connected to the vacuum chamber or closed with a blank flange.

Running the pump without a sealed inlet causes overspeed, overheating, and potential rotor damage.

Source
“Never use the turbopump when the inlet flange is not connected to the vacuum chamber.”
Operation

Vent pressure at the vent port must be less than 2 bar above atmospheric pressure, and vent gas must be free of dust or particles.

Excessive vent pressure can damage the rotor or bearings; particulate contamination can degrade vacuum performance.

Source
“The pressure at the vent port must be less than 2 bar (above atmospheric pressure).”
Maintenance

Before any maintenance, disconnect mains power, vent the pump, wait for the rotor to stop, and wait until surface temperature falls below 50 °C.

Ignoring these steps risks electric shock, burns, or injury from a spinning rotor.

Source
“Before carrying out any work on the system, disconnect it from the mains, vent the pump by opening the appropriate valve, wait until the rotor has stopped turning and wait until the surface temperature of the pump falls below 50 °C.”
Selection

The pump requires a suitable Agilent controller and must be connected to a primary (backing) pump for proper operation.

Without the correct controller and backing pump, the turbopump cannot achieve its specified performance and may be damaged.

Source
“The Turbo-V 81-T pumps must be used in conjunction with one of the suitable Agilent controller and they must be connected to a primary pump (see "Technical Information").”
Selection

The inlet screen reduces pumping speed by about 10 % and protects blades from debris larger than 0.7 mm.

When selecting the pump for a process, account for this 10 % speed loss if an inlet screen is required for debris protection.

Source
“The inlet screen, however, does reduce the pumping speed by about 10 %.”
Operation

The pump must be soft-started when first received and operated by the customer.

Soft-starting prevents excessive inrush current and mechanical stress on bearings during initial run-in.

Source
“the turbomolecular pumps must be always soft-started when received and operated for the first time by the customer”
Selection

The shelf life of the turbomolecular pump is 10 months from the shipping date; exceeding this requires return to factory.

Planning installation within the shelf life avoids costly returns and ensures bearing lubricant integrity.

Source
“the shelf life of a turbomolecular pump is 10 months from the shipping date.”
Installation

The pump belongs to installation category II per EN 61010-1 and must be connected to a mains line meeting that category.

Connecting to an inadequate mains category risks electrical safety and compliance with EN 61010-1.

Source
“The Turbo-V 81-T belongs to the second installation (or overvoltage) category as per directive EN 61010-1.”
Operation

When pumping aggressive gases, a special purge port must be used to supply inert gas (nitrogen or helium) to protect the bearings.

Without inert gas purge, aggressive gases can corrode bearings and shorten pump life.

Source
“For pumping aggressive gases, these pumps are fitted with a special port to allow a steady flow of inert gas (like Nitrogen or Helium) for pump bearing protection (see the appendix "Technical Information").”