CC-K12 Circulating Bath Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of CC-K12
  2. Spec pp. 1-2
    Technical data, performance and dimensions per DIN 12876
  3. Ordering p. 2
    Ordering information and available variants
  4. p. 2 table, 9 rows
  5. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
max. current refrigerated bath1,5 A
max. current immersion thermostat10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S395758

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-20...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 20°C0,25 kW
Cooling power at 0°C0,2 kW
Cooling power at -10°C0,12 kW
Cooling power at -20°C0,05 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,041 kg
CO2 equivalent0,0 t
Pressure / Suction pump max. delivery27 l/min
max. delivery pressure0,7 bar
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume10,5 l
Width bath opening WxD290x152 mm
Bath depth150 mm
Height of bath opening265 mm
Overall dimensions WxDxH **350x560x430 mm
Net weight29 kg
Power supply requirement220-240V 1~/2~ 50/60Hz

** см. сноску в оригинале (в тексте страницы пояснение не приведено)

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit uses flammable natural refrigerant R-290 (A3, H220), so the installation location must be assessed for flammable-refrigerant requirements.

Refrigerant safety class A3 affects site risk assessment, ventilation and compliance approvals.

Source
“R-290 (A3, H220)”
Installation

The power supply must be 220-240 V single- or two-phase at 50/60 Hz, so the site must provide a matching circuit.

Mismatched supply voltage or phase configuration prevents correct commissioning.

Source
“220-240V 1~/2~ 50/60Hz”
Installation

The circuit must be protected with a fuse between 10 A and 16 A.

Correct upstream protection is required for safe electrical installation.

Source
“min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances must be applied together: ±5% voltage with simultaneous ±2% frequency, and some combinations are not allowed.

Supply quality outside these combined limits can cause malfunction or damage.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water ingress.

IP20 is not water-protected and constrains the installation environment.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C room temperature.

Exceeding ambient limits reduces performance and may void safe operation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity falls from 0.25 kW at 20 °C to 0.05 kW at -20 °C, so low-temperature duty needs careful sizing.

Available cooling capacity at the required working temperature determines process feasibility.

Source
“at -20°C 0,05 kW”
Operation

The pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min suction at 0.4 bar.

Pump flow and pressure determine whether the external circuit can be supplied adequately.

Source
“max. delivery 27 l/min”
Operation

When using the Huber pump adapter, polyglycol must not be used as heat transfer fluid.

Using the wrong fluid with the pump adapter can damage the pump or compromise safety.

Source
“Polyglycol is not permissible to be used as a heat transfer fluid”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide the plug.

Planning must include a suitable plug and connection for three-phase units.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento

The equipment is classified as Class A, Group 1 for electromagnetic disturbance, which has implications for residential EMC environments.

Class A equipment may not be suitable for residential use without additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Correct setting of overtemperature protection is essential for safe unattended operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”