Induztr
Résumé technique Visionneuse de documents Modèles associés

CC-212B Immersion Circulator

Sommaire

  1. Overview p. 1
    Introduces the CC-212B immersion circulator device
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pump max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume10,5 l
min. filling capacity7 l
Width bath opening WxD/ bath depth290x152/ 150 mm
Overall dimensions WxDxH **350x375x375 mm
Net weight11 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The usable temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.

Determines whether an external cooler or refrigerator is required for the intended process temperature.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit accepts 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so branch protection must be sized accordingly.

Needed to plan the electrical supply and breaker rating before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A”
Installation

External branch circuit protection according to UL 489 is mandatory.

Dictates the type of upstream protection device that must be installed for compliance.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not combine in opposite directions, e.g. -10% voltage with +3% frequency is not allowed.

Prevents specifying a supply that appears within tolerance but is actually outside the permitted combined envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order.

Avoids receiving a unit with the wrong plug for the site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The bath must be filled with at least 7 l of heat transfer fluid within its 10.5 l volume.

Sets the minimum fluid charge and expansion volume needed for safe pump operation.

Source
“Bath volume 10,5 l min. filling capacity 7 l”
Operation

The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar.

Defines the external circuit flow and head available for sizing hoses and connected equipment.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Safety

The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL.

Confirms the protective cut-out feature required for unattended operation and flammable-liquid classification.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

When using the Huber pump adapter, polyglycol must not be used as heat transfer fluid.

Prevents selecting an incompatible fluid that could damage the pump adapter.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Installation

The bath bridge has a hole for a cooling probe such as an immersion cooler TC45–TC100E.

Allows planning of an immersion cooler for operation below ambient without extra modification.

Source
“Bath bridge with hole for cooling probe (e.g. for immersion cooler TC45-TC100E).”
Operation

The ambient temperature during operation must stay between 5 °C and 40 °C.

Restricts where the unit can be installed and how the room must be conditioned.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Conformité

The equipment is classified as Class A, Group 1 for electromagnetic disturbance per EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Performance

Heating power scales with supply voltage from 1.5 kW at 200 V to 2.1 kW at 240 V.

Shows that actual heating capacity depends on the site voltage, affecting heat-up time estimates.

Source
“Heating power at 240V 2,1 kW”
Maintenance

A drain valve with cap is available as an accessory for emptying the bath.

Simplifies fluid changes and maintenance planning.

Source
“Drain valve with cap #6839”