Unistat 527w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat 527w unit
  2. Spec p. 2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 20 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C880 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,2 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight448 kg
Power supply requirement400V 3~ 50Hz
max. current31,5 A
Fuse3x32A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S550886

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterUnistat 527w
Order-No.5001.0001.01
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 250°C12 kW
Cooling power with Thermooil at 200°C12 kW
Cooling power with Thermooil at 100°C12 kW
Cooling power with Thermooil at 20°C12 kW
Cooling power with Ethanol at 0°C12 kW
Cooling power with Ethanol at -20°C6 kW
Cooling power with Ethanol at -40°C2 kW
Cooling power with Ethanol at -50°C1 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,9 kg
CO2 equivalent6,2 t
Circulation pump: MK pump max. delivery196 l/min
Circulation pump: MK pump max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x32 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable cross-section and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 31,5 A Fuse 3x32A”
Installation

The cooling water circuit must be supplied at 5–20 °C with 1–4 bar differential pressure and no more than 6 bar static pressure.

Cooling water supply conditions directly affect achievable cooling capacity and safe operation.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Installation

Cooling water pipes must be insulated if the cooling water temperature can fall below the condensation point.

Prevents condensation and corrosion on the cooling water circuit at low water temperatures.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity is rated at 20 °C room temperature and 15 °C cooling water inlet with 1 bar differential pressure, and drops as cooling water temperature rises.

The published 12 kW cooling capacity is only valid under the stated water conditions; higher water temperature reduces usable capacity.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The pump connection is M38x1.5 male with a Y-shaped bore (60°) per DIN 3863, requiring matching ball-socket fittings and sleeve nuts per DIN 3870.

Ensures correct mating hoses and fittings are ordered and leak-free connections are made.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Safety

Leakage current exceeds 3.5 mA, which may require a suitable residual current device or protective earthing arrangement.

Affects electrical protection design and compliance with local installation rules.

Source
“Attention: leakage current > 3,5mA”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Site power quality must be checked; some voltage/frequency combinations are outside the permitted operating envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

The installer must plan and supply the correct three-phase plug for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C, so warmer water supplies are outside the intended operating range.

Cooling water above 20 °C may prevent the unit from reaching specified performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Selection

The refrigerant is R-452A with GWP 2141 and a charge of 2.9 kg, giving a CO2 equivalent of 6.2 t.

Relevant for refrigerant regulations, leak reporting thresholds and environmental compliance.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 2,9 kg CO2 equivalent 6,2 t”
Selection

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.

IP20 offers no water protection, and ambient limits constrain where the unit can be installed.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The minimum filling capacity is 8.2 l and the expansion tank holds 19.9 l, which sets the heat transfer fluid volume to prepare.

Needed to plan heat transfer fluid quantity and expansion volume for the loop.

Source
“min. filling capacity 8,2 l Filling capacity expansion tank 19,9 l”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or damage the cooling circuit materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”