Sommaire

  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 12 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
max. cooling water pressure6 bar
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight56 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current14,5 A
Fuse16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S592998

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P010w-H — Technical data according to DIN 12876

ParameterValue
Order-No.3050.0007.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C at 15°C1 kW
Cooling power at ambient temperature +20°C at 0°C0,8 kW
Cooling power at ambient temperature +20°C at -10°C0,5 kW
Cooling power at ambient temperature +20°C at -20°C0,15 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,2 kg
CO2 equivalent0,3 t
UN-numberUN 2857
Circulation pumpB
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
min. cooling water differential pressure1 bar

Cooling power values are given at ambient temperature +20°C for the respective coolant temperatures 15°C, 0°C, -10°C and -20°C.

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a minimum differential pressure of 1 bar to operate correctly.

Ensures the cooling water supply can provide adequate flow through the water-cooled refrigeration circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake temperatures between 5°C and 20°C, and pipes must be insulated if condensation occurs.

Prevents condensation damage and ensures specified cooling performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Operation

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Affects sizing of the cooling water supply and expected performance at elevated water temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump ordered.

Prevents chemical attack and premature pump failure when selecting heat transfer fluid.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Avoids freezing and damage in the cooling circuit when operating near or below 0°C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the power supply stays within safe operating limits and avoids invalid combinations.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without a plug.

Requires the installer to provide a suitable plug and connection for three-phase power.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit has IP20 protection, so it must be installed in a dry environment protected from water ingress.

Prevents electrical hazards and ensures compliance with environmental conditions.

Source
“Degree of Protection IP20”
Selection

The refrigerant R-449A has a GWP of 1397 and a CO2 equivalent of 0.3 t, which matters for environmental compliance.

Affects refrigerant handling regulations and environmental reporting obligations.

Source
“Global Warming Potential (GWP) 1397”
Operation

The pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar, so system pressure drop directly reduces flow.

Helps evaluate whether the pump can meet the flow demand of a high-pressure-drop application.

Source
“Delivery at 0,5 bar 21 l/min”
Installation

The maximum cooling water pressure is 6 bar, so the supply must be regulated below this limit.

Prevents damage to the cooling water circuit from overpressure.

Source
“max. cooling water pressure 6 bar”
Operation

The ambient temperature must remain between 5°C and 40°C for safe and reliable operation.

Ensures the unit operates within its designed environmental envelope.

Source
“min. ambient temperature 5 °C”