Unichiller P230

Sommaire

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

Unichiller P230 — Technical data according to DIN 12876

ParameterValue
Order-No.3039.0023.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 20°C23 kW
at 0°C13,5 kW
at -10°C9 kW
at -20°C5,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,49 kg
CO2 equivalent2,1 t
Circulation pump:E2
max. delivery140 l/min
max. delivery pressure5,3 bar
Delivery at 2,8 bar126 l/min
Delivery at 3,0 bar124 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar92 l/min
Delivery at 4,5 bar83 l/min
Delivery at 5,0 bar43 l/min
Pump connectionG1 1/4 male
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **938x1288x2003 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so process requirements outside this window cannot be met.

Defines the application envelope and excludes processes needing lower or higher temperatures.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity drops from 23 kW at 20 °C ambient to 5.5 kW at -20 °C, so sizing must use the capacity at the actual operating point.

Prevents undersizing when the process runs at low temperature or elevated ambient.

Source
“at 20°C 23 kW at 0°C 13,5 kW at -10°C 9 kW at -20°C 5,5 kW”
Installation

The unit requires a 460 V three-phase 60 Hz supply, so site power must match this configuration.

Determines electrical infrastructure and whether a transformer or different model is needed.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without plug, so the installer must provide a suitable connection.

Affects electrical installation planning and required site hardware.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Conformité

External branch circuit protection according to UL 489 is required, so the panel must include a UL 489 listed breaker.

Mandatory for compliant installation and acceptance by electrical inspection.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±5% and ±2% but cannot be combined in the unfavorable direction, so supply quality must be verified.

Avoids invalid supply conditions that could damage the compressor or void warranty.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is IP20 and must be installed in an environment between 5 and 40 °C ambient, so it needs a protected indoor location.

Prevents installation in dusty, wet, or hot areas that would compromise operation.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The refrigerant is R-449A with GWP 1397 and a CO2 equivalent of 2.1 t, which matters for environmental reporting and leak regulations.

Affects F-gas compliance, leak checks, and sustainability assessment.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 1,49 kg CO2 equivalent 2,1 t”
Installation

The pump connection is G1 1/4 male, so hoses and fittings must match this thread.

Ensures correct plumbing hardware is ordered before commissioning.

Source
“Pump connection G1 1/4 male”
Operation

The expansion tank volume is 48 l and minimum filling capacity is 18 l, so the external loop volume and fluid charge must be planned accordingly.

Prevents overflow or insufficient fluid volume in the external circuit.

Source
“min. filling capacity 18 l Volume of expansion 48 l”
Installation

Overall dimensions are 938x1288x2003 mm, so access and floor space must accommodate this footprint and height.

Needed for room planning, transport route, and service clearance.

Source
“Overall dimensions WxDxH ** 938x1288x2003 mm”
Performance

The pump delivers 140 l/min maximum and 5.3 bar maximum pressure, with flow falling to 43 l/min at 5.0 bar, so system pressure drop must be checked against the pump curve.

Ensures the external circuit does not exceed available pump head and flow.

Source
“max. delivery 140 l/min max. delivery pressure 5,3 bar”
Operation

Cooling capacity data is valid at 20 °C room temperature and may drop at higher ambient, so performance must be derated for warm rooms.

Avoids unexpected shortfall when the chiller is installed in a warm environment.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Conformité

The equipment is EMC class A group 1, so it may cause radio interference in residential areas and is intended for industrial use.

Informs site EMC planning and suitability for non-residential locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”