Agilent Technologies 1260 Infinity III SFC/UHPLC Hybrid System
Agilent Technologies 1260 Infinity III SFC/UHPLC Hybrid System — Liquid chromatography with Pumping speed 3×10⁻⁵ m³/h.
The 1260 SFC/UHPLC hybrid system combines supercritical fluid chromatography with ultra-high-performance liquid chromatography in a single instrument. It uses a 2.1 mm column with 2.7 µm particles, a 4 µL injection volume, and a 10 mm flow cell, achieving a 5.5-minute total run time. This configuration is suited for rapid method development and orthogonal separations in pharmaceutical and food analysis.
Agilent Technologies 1260 Infinity III Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Column particle size, µm | Filter membrane pore size, µm | Flow cell path length, mm | Sample volume, milliliter | Total run time, minutes |
|---|---|---|---|---|---|---|
| 1260 Infinity III SFC/UHPLC Hybrid System Current | 3×10⁻⁵ | 2.7 | 0.22 | 10 | 0.004 | 5.5 |
| 1260 Infinity III SFC System | 3×10⁻⁵ | 1.8 | 0.22 | 10 | — | — |
Full specifications
| Pumping speed | 0.5 mL/min |
| Column Particle Size | 2.7 µm |
| Filter Membrane Pore Size | 0.22 µm |
| Sample volume | 4 µL |
| Total Run Time | 5.5 minutes |
| BPR pressure mfr | 150 bar |
| Sample concentration mfr | 100 mg/L |
| Flow gradient rate mfr | 2.5...4 mL/min |
| Column dimensions mfr | 2.1 mm |
| Resolution (Rs) mfr | 1.69 |
| Solvent A mfr | Water Water |
| Solvent B mfr | ACN ACN |
| Software version mfr | 2.8 later |
| Gradient mfr | 5...95 |
| Flow Cell Path Length | 10 mm |
| Peak height increase mfr | 37.6 mAu |
| Column part number mfr | 959758 |
Description
System Architecture and Series Positioning
- The 1260 SFC/UHPLC hybrid system integrates an SFC control module (G4301A), binary pump (G4782A), multisampler (G4767A), diode array detector (G7115A) with a high-pressure SFC flow cell, and a multicolumn thermostat (G7116A) with a four-column selection valve.
- Within the 1260 Infinity II series, this hybrid model is distinguished by its dual-mode capability (SFC and UHPLC), whereas sibling models such as the 1220 Infinity II LC System and 1260 Analytical-Scale LC Purification System are dedicated to either analytical HPLC or purification. The hybrid is chosen when both SFC and UHPLC separations are required on one platform.
Separation Behavior and Performance
- The system employs a 2.1 mm internal diameter column with 2.7 µm particles, enabling high-resolution separations. A typical injection volume is 4 µL, and the flow cell path length is 10 mm for sensitive detection.
- Method conditions from the technical overview (5991 8492EN) include a flow gradient rate of 2.5–4 mL/min, a backpressure regulator (BPR) set to 150 bar and 60 °C, and a solvent gradient from 5% to 95% B (water/ACN). Under these conditions, a resolution (Rs) of 1.69 and a peak height increase of 37.6 mAu were observed, with a total run time of 5.5 minutes.
- The system supports isocratic, flow-gradient, and backpressure-gradient modes, allowing optimization of retention and selectivity for a wide range of analytes.
Installation, Control, and Operation
- The system requires connection to a suitable LC instrument controller and software (version 2.8 or later). The multicolumn thermostat accommodates up to four columns, enabling automated column selection.
- The SFC control module requires a supply of CO2 (grade 4.5 or better) and a compatible backpressure regulator. The binary pump delivers mobile phase at flow rates up to 4 mL/min. The multisampler supports sample volumes from 0.1 to 100 µL.
- The flow cell is rated for pressures up to 150 bar, and the BPR maintains system pressure to prevent phase separation in SFC mode. The system should be installed in a well-ventilated area due to CO2 use.
Integration & Operating Notes
The SFC control module requires CO2 of at least 4.5 grade purity. Moisture or particulates in the CO2 supply can cause baseline noise, restrictor clogging, or damage to the pump seals. Install a suitable filter and moisture trap in the CO2 line.
The 1260 SFC/UHPLC hybrid system requires Agilent OpenLab CDS or compatible software version 2.8 or later to coordinate SFC and UHPLC modes. Earlier versions may not support the four-column selection valve or the BPR control, leading to communication errors or incomplete method execution.
The 0.22 µm filter membrane in the solvent delivery path should be replaced at least every three months or when backpressure increases by 10 bar. A clogged membrane can cause erratic flow and damage the pump. Use only Agilent-certified replacement membranes.
Selection & Integration Q&A
Choose the hybrid when you need both supercritical fluid chromatography (SFC) and ultra-high-performance liquid chromatography (UHPLC) on one platform, especially for orthogonal separations or rapid method development. The 1220 Infinity II LC System is limited to analytical HPLC and cannot perform SFC. The hybrid also offers higher pressure capability (150 bar BPR) and a four-column selection valve for automated method scouting.
The SFC control module requires CO2 of at least 4.5 grade purity, supplied at a pressure of 5–7 bar. The backpressure regulator must be set to 150 bar and 60 °C to maintain supercritical conditions. Ensure the CO2 line is equipped with a filter and moisture trap. The BPR must be compatible with the flow cell pressure rating and connected downstream of the detector.