Agilent Technologies 5900-icp-oes
Agilent Technologies 5900-icp-oes — Atomic spectroscopy.
The Agilent 5900 ICP-OES is a dual-view (DV) atomic emission spectrometer built around a freeform echelle polychromator with a 50% reduced optical volume and 167–785 nm coverage. Its 1.8 mm injector and 35% TDS ceiling define the standard aqueous/acidic torch configuration, while the 40% detection-limit improvement over the preceding generation positions it for trace-element work in complex matrices.
Agilent Technologies ICP Oes Instruments Series — Model Comparison & Selection
| Model | Wavelength range | Detection limit improvement, % | Injector internal diameter id, mm | Maximum total dissolved solids tds, % | Polychromator volume reduction, % |
|---|---|---|---|---|---|
| 5900-icp-oes Current | 167...785 | 40 | 1.8 | 35 | 50 |
| 5800-icp-oes | 167...785 | 40 | 1.8 | 35 | 50 |
Full specifications
| Wavelength range | 167...785 nm |
| Detection Limit Improvement | 40 % |
| Injector Internal Diameter (id) | 1.8 mm |
| Maximum Total Dissolved Solids (tds) | 35 % |
| Polychromator Volume Reduction | 50 % |
| Ba 614.171 detection limit mfr | 32 |
| Most aqueous or acidic samples mfr | Standard quartz quartz |
| Volatile and semi-volatile organic samples mfr | 'Organic' high purity quartz quartz |
| As 188.980 mfr | 6.5 |
| Mo 202.032 mfr | 7 |
| Zn 213.857 mfr | 7.5 |
| Pb 220.353 mfr | 7.5 |
| Cr 267.716 mfr | 9.5 |
| Cu 327.395 mfr | 13 |
| Ba 614.171 mfr | 32 |
| pixel count mfr | 70000 |
| DLA count mfr | 70 |
| full spectrum measurement time mfr | less than half a second second |
| injector inner diameter mfr | 2.4 mm |
| optical focal length mfr | 253 mm |
| High-TDS matrices rich in alkali metals (including Li battery materials), hydrofluoric (HF) acid digests or fusions mfr | Ceramic outer outer |
| Cooling time before torch removal mfr | at least five minutes minutes |
| cooling time before torch removal mfr | cinque minuti minuti |
Description
Construction and series positioning
The 5900 is the higher-sensitivity member of the 5800/5900 ICP-OES pair; the sibling 5800 shares the freeform-optics architecture but the 5900 is specified with a 40% detection-limit improvement and a polychromator volume reduced by 50%. The instrument is configured for Dual View (DV) operation, which requires the slotted, longer DV outer tube-set rather than the shorter, slot-free Radial View (RV) tube-set. Torch hardware follows the Easy-fit family: one-piece, semi-demountable, and fully demountable variants are all supported, and the fully demountable torch can be switched between RV and DV by changing only the outer tube-set.
Torch and sample-path behavior
- Standard quartz outer tube-set with 1.8 mm id injector: general aqueous and acidic samples, most economical configuration.
- Organic high-purity quartz outer tube-set with 0.8 mm id (volatile) or 1.4 mm id (semi-volatile) injector: improved resistance to solarization for organic matrices.
- Ceramic outer tube-set with 2.4 mm id inert alumina injector (G8020-47004): longest service life and resistance to devitrification for HF digests, fusions, and high-TDS matrices up to 35% TDS.
The 35% TDS ceiling is tied specifically to the 2.4 mm alumina injector; it is not a blanket limit for every injector diameter.
Installation, control and operation
The fully demountable torch ships without the injector installed. The injector is inserted until the bulge contacts the retaining clamp and clicks into position, then the locking nut is fitted by hand only. During reassembly the groove on the torch body must align with the 'T' mark on the outer tube-set, and the tube-set is pressed fully into the body. The ceramic outer tube-set carries a fixed polymer adapter on its quartz intermediate tube that must not be removed; the supplied tube-set tool is used for assembly and disassembly of the ceramic/quartz joint. Torch installation and removal must follow a straight, parallel path to avoid breaking quartz or ceramic components. After plasma switch-off, the torch and compartment must cool for at least five minutes before the torch is removed. Detailed maintenance and cleaning procedures are delegated to the ICP Expert Help and Learning Center and publication 5994-8864EN.
Integration & Operating Notes
The 35% total-dissolved-solids ceiling applies to the G8020-47004 2.4 mm id inert alumina injector used with the ceramic outer tube-set. Higher TDS loads risk salt deposition and devitrification-related failure; for matrices above this level, dilution or a different sample-introduction strategy is required before the torch is installed.
Any angular motion while sliding the torch or its quartz/ceramic components into the compartment can crack the outer tube-set or injector and create a hand-injury hazard from broken glass. The torch must be guided straight in and out; the same rule applies during reassembly of the fully demountable torch.
The locking nut (G8020-60810) must be tightened by hand. Using tools or over-torquing can damage both the injector and the torch body, and a damaged body is not field-repairable without replacement parts. The same hand-tight rule applies at first injector installation and at every reassembly.
The quartz intermediate tube of the ceramic outer tube-set carries a fixed polymer adapter. It is not a serviceable joint and must not be removed during cleaning or maintenance; detaching it compromises the tube-set and requires replacement of the assembly.
Selection & Integration Q&A
The 5900 is specified with a 40% detection-limit improvement and a 50% smaller polychromator volume than the preceding design; the 5800 shares the freeform-optics platform. Choose the 5900 when trace-level detection limits in complex matrices drive the method; the 5800 covers routine work where those limits are not the constraint.
Use the ceramic outer tube-set with the 2.4 mm id inert alumina injector G8020-47004, rated to 35% TDS. The ceramic tube-set kit does not include an injector, so a compatible quartz or alumina injector must be ordered separately; the supplied tube-set tool is required for assembly.
The fully demountable torch ships without an injector installed. Stock replacement injectors (quartz or inert alumina, 0.8/1.4/1.8/2.4 mm id), outer tube-sets for RV or DV, top seal pack G8014-60023, and injector locking nut G8020-60810. Cleaning procedures are in publication 5994-8864EN.