Agilent Technologies Quadrupole Time-of-Flight GC/MS, Mass Screening GC/Q-TOF
Agilent Technologies Quadrupole Time-of-Flight GC/MS, Mass Screening GC/Q-TOF — Gas chromatography mass spectrometry gc ms.
The Agilent 7250 Series GC/Q-TOF is a quadrupole time-of-flight GC/MS configured for accurate-mass screening (50–950 m/z) with 70 eV EI at a 200 °C source and 150 °C quadrupole. It is distinguished within the GC/MS series by accurate-mass, retention-time-locked library workflows (900+ entries) rather than single-quadrupole or triple-quadrupole targeted analysis. It serves nontargeted and targeted screening of complex extracts in metabolomics, natural-products and E&L work.
Agilent Technologies GC MS Instruments Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Carrier gas | Transfer line temperature, °C | Sample volume, milliliter | Ion source temperature, °C | Analysis time, min |
|---|---|---|---|---|---|---|
| Quadrupole Time-of-Flight GC/MS, Mass Screening GC/Q-TOF Current | — | 1.5 | 280 | — | — | — |
| 5977C GC/MSD | 1 | Hydrogen | 230 | 3×10⁻⁴ | 230 | 7.5 |
| 7000E GC/TQ | 6×10⁻⁵ | Helium | 300 | 0.002 | 280 | 10 |
| 7010D GCMS | — | Helium | 300 | 0.002 | 280 | — |
Full specifications
| Carrier gas | 1.5 Helium |
| Electron Energy | 70 eV |
| Spectral Acquisition Rate | 50 Hz |
| Wavelength range | 50...1,000 m/z |
| Column mfr | -5 μm |
| Oven program mfr | 50 minutes |
| Source mode mfr | 70 eV |
| Library entries mfr | 900 entries |
| Library type mfr | accurate mass retention time locked EI library l |
| Compounds identified in mouse tissues mfr | 200 compounds |
| Compounds identified in plasma samples mfr | 139...168 compounds |
| Retention time range mfr | 6.48...29.48 min |
| Match score range mfr | 80.3...99.3 |
| Library score range mfr | 80...99.9 |
| Modulation period mfr | 6 s |
| Cold jet flow mfr | 13 L/min |
| Hot jet duration mfr | 320 ms |
| Mass range | 50...950 m/z |
| Quadrupole Temperature | 150 °C |
| Source Temperature | 200 °C |
| Transfer Line Temperature | 280 °C |
| Inlet mfr | 310 °C |
| Transferline mfr | 280 °C |
Description
Construction and series positioning
- This is the accurate-mass Q-TOF modification of the Agilent GC/MS line; the sibling 5977C GC/MSD is a single-quadrupole instrument, the 7000E GC/TQ and 7010D GCMS are triple-quadrupole targeted platforms.
- Choose the Q-TOF when accurate mass and library-based nontargeted identification are required; choose the single quadrupole for routine targeted quantitation and the triple quadrupole for high-sensitivity MRM work.
- The system is documented in application notes coupled to an Agilent 7890B GC, including 1D GC and GC×GC configurations.
Behavior and operating envelope
- Mass range is stated as 50–950 m/z in the matched specs and 50–1,000 m/z in the wavelength-range field; these two figures conflict, so no single upper mass limit is asserted here.
- Ionization is 70 eV EI; source 200 °C, quadrupole 150 °C, transfer line 280 °C, inlet 310 °C.
- Spectral acquisition rate is 50 Hz.
- GC×GC operation is described with a 6 s modulation period, 13 L/min cold jet, 300 °C hot jet and 320 ms hot jet duration.
- Library workflows use an accurate-mass, retention-time-locked EI library of over 900 entries; reported screening identified ~200 compounds in mouse tissue and 131–157 in plasma, with match scores 80.3–99.3 and library scores 80–99.9.
Installation and operation
- The document digests are application notes, not an installation manual; they do not provide electrical, utility, site-preparation or control-I/O specifications, so those must be taken from the separate hardware manual.
- Documented configurations pair the Q-TOF with a 7890B GC; GC×GC operation requires the modulator hardware and the jet parameters above.
Integration & Operating Notes
For 2D GC×GC the documented conditions are a 6 s modulation period, 13 L/min cold jet flow, 300 °C hot jet and 320 ms hot jet duration. These are coupled settings; changing the modulation period without re-checking jet flows and durations will degrade peak capacity and library match quality.
Selection & Integration Q&A
Pick the Q-TOF for accurate-mass nontargeted screening and library identification (900+ entry accurate-mass EI library). The 5977C single quadrupole suits routine targeted work; the 7000E and 7010D triple quadrupoles suit high-sensitivity targeted MRM quantitation.
The application note describes the Q-TOF coupled to an Agilent 7890B GC with a modulator set to a 6 s period, 13 L/min cold jet, 300 °C hot jet and 320 ms hot jet duration. Plan the GC, modulator and jet gas supply as part of the system, not as an accessory.
The documents are application notes only; they do not list a base scope. The GC (e.g. 7890B), GC×GC modulator and jet gases, the accurate-mass PCDL library, and consumables such as the -5 µm column are described as separate items and must be ordered and specified separately.