Agilent Technologies AdvanceBio Gly-X Technology | N-glycan Analysis
Agilent Technologies AdvanceBio Gly-X Technology | N-glycan Analysis — Biopharma hplc analysis.
The AdvanceBio Gly-X Technology is a kit-based N-glycan sample preparation and analysis workflow, not a standalone instrument. It combines rapid enzymatic deglycosylation and fluorescent labeling with HILIC-UHPLC separation and fluorescence or MS detection. This modification is distinguished by its 2-AB Express and InstantPC labeling options, enabling a total workflow time of approximately 2 hours.
Agilent Technologies Glycan Preparation Series — Model Comparison & Selection
| Model | Mass range | Nozzle voltage, V | Sheath gas flow, liter/minute | Fluorescence detector maximum data rate, Hz | Glycan sample preparation time, h | Pumping speed, m³/h |
|---|---|---|---|---|---|---|
| AdvanceBio Gly-X Technology | N-glycan Analysis Current | 600...3,000 | 500 | 10 | 74 | 1 | — |
| Sialic Acid Profiling and Quantitation Kits | Analysis | — | 1800 | 12 | 74 | 1 | 2.4×10⁻⁵ |
Full specifications
| Column temperature | 40 °C |
| Fluorescence Detector Maximum Data Rate | 74 Hz |
| Glycan Sample Preparation Time | 1 hour |
| Sheath Gas Flow | 10 L/min |
| Denature time (GlykoPrep with 2-AB) mfr | 5 min |
| Deglycosylation time (GlykoPrep with 2-AB) mfr | 15...60 min |
| Dry down time (GlykoPrep with 2-AB) mfr | 30...60 min |
| Labeling time (GlykoPrep with 2-AB) mfr | 60 min |
| Cleanup time (GlykoPrep with 2-AB) mfr | 15...30 min |
| Total workflow time (GlykoPrep with 2-AB) mfr | 3.5 hours |
| Denature time (Gly-X with 2-AB Express) mfr | 3 min |
| Deglycosylation time (Gly-X with 2-AB Express) mfr | 5 min |
| Labeling time (Gly-X with 2-AB Express) mfr | 60 min |
| Cleanup time (Gly-X with 2-AB Express) mfr | 10...15 min |
| Total workflow time (Gly-X with 2-AB Express) mfr | 2 hours |
| Fluorescence detection excitation/emission mfr | 360 nm |
| Cleanup time mfr | 10...15 minutes |
| Protein amount per preparation mfr | 40 μg |
| Mass range | 600...3,000 |
| Nozzle Voltage | 500 V |
| Vcap mfr | 3000 V |
| Fragmentor mfr | 120 V |
| Skimmer mfr | 65 V |
Description
Construction and Positioning
- The Gly-X Technology is a consumable kit and protocol set for N-glycan sample preparation, designed for use with standard HPLC/UHPLC systems and fluorescence or mass spectrometry detectors.
- Within the Glycan Preparation series, it is positioned as the rapid workflow option: the Gly-X with 2-AB Express achieves a total workflow time of 2 hours, compared to 3.5 hours for the GlykoPrep with 2-AB workflow.
- Sibling products include GlykoPrep-plus for N-glycan sample prep with AssayMAP Bravo (automation-focused) and Sialic Acid Profiling and Quantitation Kits (different analyte focus).
Workflow Behavior and Media
- The workflow involves sequential steps: denaturation (90 °C), deglycosylation (50 °C), labeling (65 °C for 2-AB Express or 50 °C for InstantPC), and cleanup (10–15 min).
- Protein amount per preparation is 40 µg, suitable for monoclonal antibody and fusion protein samples such as rituximab and etanercept.
- Labeled glycans are analyzed by HILIC-UHPLC with fluorescence detection (excitation/emission 360 nm) or MS detection; the mass range covered is 600–3,000 Da.
Installation and Operation
- The kit requires a thermocycler or heating blocks capable of precise temperatures (90 °C, 50 °C, 65 °C) and a centrifuge or vacuum manifold for cleanup steps.
- LC separation typically uses a 2.1 mm internal diameter AdvanceBio Amide HILIC column at 40 °C, with fluorescence detector data rate up to 74 Hz.
- For MS detection, typical source parameters include nozzle voltage 500 V, sheath gas flow 10 L/min, fragmentor 120 V, skimmer 65 V, and Vcap 3000 V.
Integration & Operating Notes
The protocol is validated for 40 µg protein per preparation; exceeding this load may overload the cleanup resin and reduce labeling efficiency, leading to incomplete glycan profiles or inconsistent fluorescence response.
Denaturation requires 90 °C, deglycosylation 50 °C, and labeling 65 °C for 2-AB Express or 50 °C for InstantPC. Temperature deviations can cause incomplete deglycosylation or label degradation, affecting glycan quantitation.
The Gly-X kits (p/n GX96-IPC and GX96-2AB) are consumables with defined expiry; using expired kits may result in reduced enzyme activity or label degradation, compromising glycan profiles. Store according to the kit instructions.
Selection & Integration Q&A
Choose Gly-X for a rapid, manual or semi-automated workflow with 2-hour total time and 40 µg protein per prep. Choose GlykoPrep-plus with AssayMAP Bravo when full automation and higher throughput are required, accepting a longer 3.5-hour workflow.
The base kit includes enzymes, labels, and cleanup plates, but requires additional consumables such as pipette tips, microcentrifuge tubes, and LC mobile phases (e.g., AdvanceBio Ammonium Formate). Regular maintenance of the LC/FLD/MS system per manufacturer schedules is necessary.