Agilent Technologies Analyzers to measure and monitor GHG emissions
Agilent Technologies Analyzers to measure and monitor GHG emissions — Gas chromatography with Motor power 2.35 kW.
The Agilent 8890-series Environmental GC Analyzer is a 54 × 50 × 54 cm, 50 kg benchtop gas chromatograph configured for greenhouse-gas emission measurement. It draws 2350 W from a 120 VAC line at 19.6 A, requiring a dedicated 20 A outlet. It is the GHG-analysis modification of the 8890 platform, distinct from the trace-sulfur sibling.
Agilent Technologies Environmental GC Analyzers Series — Model Comparison & Selection
| Model | Motor power, kW | Depth, mm | Frequency, Hz | Height, cm | Maximum altitude, m | Power outlet current, Amp |
|---|---|---|---|---|---|---|
| Analyzers to measure and monitor GHG emissions Current | 2.35 | 540 | 0.5 | 50 | 4615 | 20 |
| Trace sulfur analysis | 2.35 | 540 | 0.5 | 50 | 4615 | 20 |
Full specifications
| Motor power | 2350 W |
| Depth | 54 cm |
| Height | 50 cm |
| Weight | 50 kg |
| Width | 54 cm |
| Frequency | 50 % Hz |
| Power Outlet Current | 20 Amp |
| Line voltage mfr | 120 VAC |
| Current rating mfr | 19.6 amps |
| Maximum Altitude | 4615 m |
| Operating humidity range mfr | 5...95 |
| Storage humidity range mfr | 5...95 |
Description
Construction and positioning
- Benchtop GC, 54 cm wide × 50 cm high × 54 cm deep, 50 kg.
- This model is the environmental/GHG-emissions configuration of the 8890 series; the sibling listed is the trace-sulfur analysis variant. Choose this one for GHG monitoring, the sibling for sulfur speciation.
- The 8890 platform is offered with standard or fast oven ramp and in 120 VAC (Americas) or 220/230/240 VAC configurations; the electrical figures below apply to the 120 VAC build.
Electrical behavior
- Motor power 2350 W; at 120 VAC the current rating is 19.6 A, so the branch circuit must be rated 20 A.
- The 2350 W figure is the drive/motor power of the oven system, not a total instrument rating; size the circuit from the 19.6 A current rating.
Environment and installation
- Operating ambient 15–35 °C, 5–95 % RH; storage 40–70 °C, 5–95 % RH.
- Maximum operating altitude 4615 m.
- Site preparation follows the 8890 GC site preparation checklist (D0007040), which also covers the data-system PC and power-cord options.
Integration & Operating Notes
The instrument draws 19.6 A at 120 VAC, so it must sit on a circuit rated 20 A with no other heavy loads sharing it. Undersized wiring or a shared circuit will trip or sag during oven ramps, and the 2350 W motor power leaves almost no headroom on a 15 A circuit.
Operation outside 15–35 °C or 5–95 % RH invalidates the thermal environment the GC oven and detectors were characterized in. Plan the lab HVAC and bench placement accordingly; storage outside 40–70 °C is also outside specification.
The site preparation checklist treats the data-system PC and the power cord as separate configuration items from the GC itself. Confirm the PC specification and the correct cord/plug type for the 120 VAC build before the site is signed off.
The maximum operating altitude is 4615 m. Above this the oven cooling and detector electronics are outside their qualified envelope; high-altitude sites need to be checked against this limit before specifying the instrument.
Selection & Integration Q&A
Pick this model when the analytical target is greenhouse-gas emission measurement. The sibling listed in the series is the trace-sulfur analysis configuration; it shares the 2350 W drive but is set up for sulfur speciation, so the choice is driven by the analyte, not by the platform.
A 120 VAC line rated 19.6 A on a 20 A outlet. The 2350 W motor power is the oven drive figure; size the branch circuit from the 19.6 A current rating and keep the circuit dedicated so oven ramps do not share capacity with other equipment.
Operating ambient 15–35 °C at 5–95 % RH, storage 40–70 °C at 5–95 % RH, and a maximum operating altitude of 4615 m. Sites outside these envelopes need re-evaluation before the instrument is placed.
The data-system PC and the power cord are handled as separate configuration items in the 8890 site preparation checklist (D0007040), not as part of the GC itself. Confirm PC specification and cord/plug type for the 120 VAC build during site preparation.