Agilent Technologies Analyzers to measure and monitor GHG emissions
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Analyzers to measure and monitor GHG emissions

Agilent Technologies Analyzers to measure and monitor GHG emissions — Gas chromatography with Motor power 2.35 kW.

The Agilent 8890-series Environmental GC Analyzer is a 54 × 50 × 54 cm, 50 kg benchtop gas chromatograph configured for greenhouse-gas emission measurement. It draws 2350 W from a 120 VAC line at 19.6 A, requiring a dedicated 20 A outlet. It is the GHG-analysis modification of the 8890 platform, distinct from the trace-sulfur sibling.

Motor power
2.35 kW

Agilent Technologies Environmental GC Analyzers Series — Model Comparison & Selection

Model Motor power, kWDepth, mmFrequency, HzHeight, cmMaximum altitude, mPower outlet current, Amp
Analyzers to measure and monitor GHG emissions Current 2.355400.550461520
Trace sulfur analysis 2.355400.550461520

Full specifications

Core Performance & Functional Specifications
Motor power 2350 W
Dimensions, Weight & Mechanical Design
Depth 54 cm
Height 50 cm
Weight 50 kg
Width 54 cm
Electrical, Power & Drive Specifications
Frequency 50 % Hz
Power Outlet Current 20 Amp
Line voltage mfr 120 VAC
Current rating mfr 19.6 amps
Operating Conditions, Environment & Protection
Maximum Altitude 4615 m
Operating humidity range mfr 5...95
Storage humidity range mfr 5...95

Description

Construction and positioning

Electrical behavior

Environment and installation

Integration & Operating Notes

Electrical
Dedicate a 20 A branch circuit at 120 VAC

The instrument draws 19.6 A at 120 VAC, so it must sit on a circuit rated 20 A with no other heavy loads sharing it. Undersized wiring or a shared circuit will trip or sag during oven ramps, and the 2350 W motor power leaves almost no headroom on a 15 A circuit.

Factory Specification Sheet; 8890 GC Site Preparation Checklist D0007040
Mounting
Keep ambient between 15 and 35 °C

Operation outside 15–35 °C or 5–95 % RH invalidates the thermal environment the GC oven and detectors were characterized in. Plan the lab HVAC and bench placement accordingly; storage outside 40–70 °C is also outside specification.

Factory Specification Sheet
Control & I/O
Data-system PC and power cord are separate line items

The site preparation checklist treats the data-system PC and the power cord as separate configuration items from the GC itself. Confirm the PC specification and the correct cord/plug type for the 120 VAC build before the site is signed off.

8890 GC Site Preparation Checklist D0007040, Rev. 3.0
Process Limits
Do not install above 4615 m altitude

The maximum operating altitude is 4615 m. Above this the oven cooling and detector electronics are outside their qualified envelope; high-altitude sites need to be checked against this limit before specifying the instrument.

Factory Specification Sheet

Selection & Integration Q&A

When should I specify this environmental GHG analyzer rather than the trace-sulfur sibling in the 8890 series?

Pick this model when the analytical target is greenhouse-gas emission measurement. The sibling listed in the series is the trace-sulfur analysis configuration; it shares the 2350 W drive but is set up for sulfur speciation, so the choice is driven by the analyte, not by the platform.

What power supply and circuit does the 120 VAC build of this analyzer require?

A 120 VAC line rated 19.6 A on a 20 A outlet. The 2350 W motor power is the oven drive figure; size the branch circuit from the 19.6 A current rating and keep the circuit dedicated so oven ramps do not share capacity with other equipment.

What ambient and altitude limits govern where this GC can be installed?

Operating ambient 15–35 °C at 5–95 % RH, storage 40–70 °C at 5–95 % RH, and a maximum operating altitude of 4615 m. Sites outside these envelopes need re-evaluation before the instrument is placed.

What is not included in the base delivery and must be ordered or prepared separately?

The data-system PC and the power cord are handled as separate configuration items in the 8890 site preparation checklist (D0007040), not as part of the GC itself. Confirm PC specification and cord/plug type for the 120 VAC build during site preparation.