Agilent Technologies Multiplexing LC/MS, multichannel, multiple stream StreamSelect
Agilent Technologies Multiplexing LC/MS, multichannel, multiple stream StreamSelect — Liquid chromatography mass spectrometry lc ms.
The Agilent StreamSelect multiplexing LC/MS system is a multichannel, multiple-stream front end that time-slices a single triple quadrupole MS across four parallel UHPLC streams. It is defined by the StreamSelect Console software and the attached application methods, not by a single hardware rating. Its purpose is to raise sample throughput by overlapping chromatographic runs from several streams into one MS acquisition cycle.
Agilent Technologies High Throughput LC MS Series — Model Comparison & Selection
| Model | Sample volume, milliliter | Autosampler temperature, °C | Capillary voltage, V | Column temperature, °C | Drying gas temperature, °C | Fragmentor voltage, V |
|---|---|---|---|---|---|---|
| Multiplexing LC/MS, multichannel, multiple stream StreamSelect Current | 0.003 | 4 | 500 | Ambient | 250 | 200 |
| Rapid fire mass spectrometry, high throughput screening | — | — | — | — | — | — |
Full specifications
| Column temperature | Ambient |
| Gas Flow | 14 L/min |
| Sample volume | 3 μL |
| Sheath Gas Flow | 12 liter/minute |
| Analytical column mfr | 120 μm |
| Stop time mfr | 5 minutes |
| Isocratic separation mfr | 5 minutes |
| Instrument mfr | 6490 LC/MS |
| Ion mode mfr | Positive Positive |
| Nebulizer mfr | 20 psi |
| Intra-assay imprecision ADMA mfr | 1.7 |
| Intra-assay imprecision SDMA mfr | 2.4 |
| Inter-assay imprecision ADMA mfr | 3.0 |
| Inter-assay imprecision SDMA mfr | 2.8 |
| Mean bias at 30 ng/mL ADMA mfr | 0.7 |
| Mean bias at 30 ng/mL SDMA mfr | 2.5 |
| Mean bias at 90 ng/mL ADMA mfr | 1.2 |
| Mean bias at 90 ng/mL SDMA mfr | 3.9 |
| Needle wash mfr | 50 % Methanol |
| Nebulizer pressure mfr | 45 psi |
| Q1/Q3 resolution mfr | 0.7 unit |
| Dwell time mfr | 18 ms |
| Calibrator range mfr | 5...5,000 ng/mL |
| Chromatographic method duration mfr | six-minute minute |
| Calibration linearity (R2) mfr | 0.996 |
| Retention time reproducibility (%RSD) mfr | 0.43...1.54 |
| Quantitation accuracy mfr | 95.3...109.1 |
| Capillary Voltage | 500 V |
| Fragmentor Voltage | 200 V |
| Mobile phase A mfr | 5 |
| Mobile phase B mfr | 0.01 acid |
| EMV mfr | 100 V |
| Autosampler Temperature | 4 °C |
| Drying Gas Temperature | 250 °C |
| Sheath Gas Temperature | 300 °C |
Description
Architecture and series positioning
- The StreamSelect configuration is a multiplexing LC/MS front end: several LC streams are coupled to one mass spectrometer through a stream-selection interface, and acquisition is scheduled by the MassHunter StreamSelect Console (version 8.0 Quick Start Guide).
- The attached application notes instantiate the concept with a four-stream Agilent 1290 Infinity II UHPLC setup feeding an Agilent 6470 triple quadrupole, and with a four-stream StreamSelect system on an Agilent 6490 triple quadrupole. These are application configurations, not fixed model ratings.
- Within the series, this is the multichannel/multiple-stream variant; the sibling 'Rapid fire mass spectrometry, high throughput screening' entry is a different high-throughput screening approach, not a stream-multiplexed LC front end.
Behavior under load and in different methods
- Throughput comes from overlapping streams: the ADMA/SDMA method uses a six-minute chromatographic method with a 5-minute stop time and 5-minute isocratic separation, while the cannabinoid note runs six analytes plus deuterated internal standards across four streams.
- The ADMA/SDMA application reports intra-assay imprecision of 1.7 (ADMA) and 2.4 (SDMA), inter-assay imprecision of 3.0 and 2.8, calibration linearity R² 0.996 over 5–5,000 ng/mL, retention-time reproducibility 0.43–1.54 %RSD, and quantitation accuracy 95.3–109.1%.
- Source and MS settings in the ADMA/SDMA method include drying gas 250 °C, sheath gas 300 °C at 12 L/min, capillary 500 V, fragmentor 200 V, nebulizer 20–45 psi, Q1/Q3 resolution 0.7, dwell time 18 ms, and positive ion mode.
Control, installation and operation
- Method setup and batch sample submission are performed in the MassHunter StreamSelect Console; the Quick Start Guide covers its windows and the acquisition-method workflow.
- The application notes do not define electrical supply, utility, or mounting requirements for the StreamSelect hardware; those must come from the base LC and MS platform documentation.
- Column temperature is stated as ambient in the ADMA/SDMA method; autosampler temperature is 4 °C and sample volume 3 µL, with a needle wash of 50% methanol.
Integration & Operating Notes
Acquisition across streams is coordinated by the MassHunter StreamSelect Console, so the batch and acquisition method must be built and submitted there; the LC and MS methods alone do not define the stream timing. Plan method development around the Console workflow described in the Quick Start Guide.
The ADMA/SDMA application uses a six-minute chromatographic method with a 5-minute stop time and 5-minute isocratic separation; stream overlap only works when each stream's run length is matched to the acquisition schedule. Methods that overrun the allotted window will collide with the next stream's injection.
Drying gas 250 °C, sheath gas 300 °C at 12 L/min, capillary 500 V, fragmentor 200 V and nebulizer 20–45 psi appear in the ADMA/SDMA method; the cannabinoid note uses a different MS platform (6470). Treat these as starting method values to be tuned per assay, not as instrument ratings.
Selection & Integration Q&A
Choose StreamSelect when you need chromatographic separation on multiple parallel LC streams feeding one triple quadrupole, as in the four-stream ADMA/SDMA and cannabinoid methods. Rapid Fire is a different high-throughput screening approach without stream-multiplexed LC separation; pick it only when chromatographic resolution is not the bottleneck.
The application notes pair StreamSelect with Agilent 1290 Infinity II UHPLC streams and either a 6490 or 6470 triple quadrupole MS, coordinated through the MassHunter StreamSelect Console. Confirm your LC pumps, autosampler and MS model are supported by the Console version in use before committing the configuration.
The method requires a 50% methanol needle wash and 4 °C autosampler operation, so wash solvent and autosampler cooling are routine consumables/checks. Analytical column is 120 µm with mobile phases 5 (A) and 0.01 acid (B). The Quick Start Guide covers Console software setup; hardware service intervals are not defined in the attached documents and must come from the LC/MS platform manuals.