Agilent Technologies 5500 Series FTIR
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 5500 Series FTIR

Agilent Technologies 5500 Series FTIR — Molecular spectroscopy.

The Agilent 5500 Series FTIR is a compact 20.3 × 20.3 × 11.4 cm, 3.6 kg spectrometer platform configured with one of three sampling accessories: TumblIR liquid transmission (5500t), diamond ATR (5500a), or DialPath variable-pathlength transmission. This card covers the series as a whole; the specific accessory determines sample compatibility, pathlength, and pH limits. It is intended for onsite lubricant analysis and general mid-IR sample identification in non-hazardous, indoor environments.

Full specifications

Core Performance & Functional Specifications
Atr Sample Press Force 15 pounds
Dimensions 20.3 cm
Single Reflection Atr Active Area 200 µm
Single Reflection Atr Sampling Surface Diameter 1 mm
Triple Reflection Atr Active Area 200 µm
Triple Reflection Atr Sampling Surface Diameter 2 mm
Tumblir Bottom Window Diameter 2 mm
Tumblir Pathlength 100 micrometers
ATR window thickness mfr 0.5 less
ATR sample press sample thickness assumption mfr 0.03 inch
DialPath pathlengths mfr three factory-set pathlengths pathlengths
Warmup period before analysis mfr 5 minutes
Warmup period before performance validation tests mfr 30 minutes
Stability test regions mfr 3000
Stability test sample interval mfr one sample every minute minute
Diagnostic value Energy (centerburst) optimal mfr 26000...21,000
Diagnostic value Energy (centerburst) marginal mfr 30000
Diagnostic value Energy (centerburst) critical mfr 31000
Diagnostic value Laser signal optimal mfr 15000...3,000
Diagnostic value Laser signal marginal mfr 15000
Diagnostic value Laser signal critical mfr 17000
Energy (centerburst) optimal mfr 26000...21,000
Energy (centerburst) marginal mfr 30000
Energy (centerburst) critical mfr 31000
Laser signal optimal mfr 15000...3,000
Laser signal marginal mfr 15000
Laser signal critical mfr 17000
System OFF mfr Red Red
System startup mfr Red/green Red/green
System ON mfr Red/green Red/green
Firmware update mfr Red/green Red/green
ATR sample press assumed sample thickness mfr 0.03 inch
Dimensions, Weight & Mechanical Design
Single Reflection Atr Depth Of Penetration 2 micron
Triple Reflection Atr Effective Penetration Depth 6 micron
Weight 3.6 kg
Electrical, Power & Drive Specifications
Laser frequency calibration reference mfr 1921 film
Diagnostic value Battery life optimal mfr 30 min
Diagnostic value Battery life marginal mfr 30...5 min
Diagnostic value Battery life critical mfr 5 min
Diagnostic value Source optimal mfr 1.9 A
Diagnostic value Source marginal mfr 2.2 A
Diagnostic value Source critical mfr 2.5 A
Battery life optimal mfr 30 min
Battery life marginal mfr 30...5 min
Battery life critical mfr 5 min
Source optimal mfr 1.9 A
Source marginal mfr 2.2 A
Source critical mfr 2.5 A
Low battery mfr Green Green
Critically low battery mfr Red Red
Dead battery mfr No LED illumination illumination
Operating Conditions, Environment & Protection
Signal-to-noise test regions mfr 2500
Signal-to-noise test resolution mfr 4
Signal-to-noise test collection time mfr one-minute minutes minutes
Signal-to-noise test duration mfr 2 test
Signal-to-noise minimum number of tests mfr 10 tests
Minimum number of signal-to-noise tests mfr 10 tests

Description

Construction and series positioning

Behavior under load and in different modes

Installation, control, and operation

Integration & Operating Notes

Electrical
Surge strip and connection order are mandatory

Connect the power cord to the spectrometer first, then to the AC outlet, with a UL-approved surge protection strip in between. Reversing the order or omitting the strip exposes the instrument to power startup damage and surge events; the blue cap must be threaded clockwise until finger tight.

User Manual, Connecting power, p.9
Process Limits
ZnSe accessories accept only pH 4–9 liquid samples

TumblIR and DialPath use ZnSe windows that tolerate only pH 4–9 and must not see solid or powder samples such as tablets or hard graphite lubricant. Solids damage the windows or alter the fixed 100 µm pathlength; route extreme-pH or solid samples to the diamond ATR instead.

User Manual, Lubrication samples p.18–19; DialPath p.32
Process Limits
Diamond ATR window can crack under sharp samples

The single-reflection ATR diamond is 0.5 mm or less thick; the spring-loaded press applies 15 lb force assuming 0.03 in sample thickness. Avoid using the press on sharp or pointed samples, and do not leave extremely acidic samples on the metal mounting plate for extended periods even though the diamond accepts pH 1–14.

User Manual, Collecting a sample spectrum, p.29
Control & I/O
Instrument blocks data collection on gross diagnostic faults

If diagnostic values (energy, source current, laser signal, detector temperature, CPU temperature, battery life) are grossly out of specification, the instrument will not collect data. Check the diagnostic page daily or weekly; a yellow or red status circle requires contacting Agilent technical support before critical measurements.

User Manual, Performance verification, p.13–14
Maintenance
Laser calibration only under Agilent guidance

Laser frequency calibration uses NIST SRM 1921 polystyrene film and should only be set after consultation with an Agilent technical support engineer. Independent adjustment affects the frequency (X-axis) accuracy of all spectra and cannot be corrected by routine performance tests.

User Manual, Laser Frequency Calibration test, p.15
Mounting
Indoor, non-hazardous location only

The system is not intrinsically safe and must only be used in atmospheres tested for flammable materials. It is intended for indoor use; allow the shipment to reach room ambient temperature for several hours before opening to avoid condensation.

User Manual, Introduction p.5; Unpacking

Selection & Integration Q&A

When should I choose the 5500a with single-reflection ATR over the 5500t with TumblIR or the triple-reflection ATR?

Pick single-reflection ATR for solids, powders, rubber, polymers, paints, fibers, or extreme-pH samples (diamond accepts pH 1–14) and limited sample amounts; its ~2 µm penetration suits higher-absorbing materials. Choose TumblIR for pH 4–9 liquids needing a fixed 100 µm transmission path; choose triple-reflection ATR for lower-absorbing liquids/pastes only (no press, ~6 µm penetration).

What adjacent equipment and connections are required to put the 5500 Series into operation?

A PC with MicroLab PC software, the supplied USB cable to the upper-left rear USB port, and the supplied power cord to the lower-left rear connector with a UL-approved surge protection strip between the outlet and the cord. Connect the cord to the spectrometer first, then the outlet. First logon is Admin/admin (case-sensitive).

What mounting, environment, and duty-cycle limits apply to the 5500 Series?

Indoor use only; not intrinsically safe, so only in atmospheres tested for flammable materials. Allow the shipment to reach room temperature for several hours before opening. A 5-minute warmup precedes analysis; validation tests need 30 minutes warmup. Detector temperature optimal 35–44 (critical <30 or >48) and CPU temperature optimal 10–75 (critical >80) constrain ambient conditions and ventilation.

What maintenance schedule and consumables/accessories are needed beyond the base delivery?

Check diagnostic values daily or weekly; run the performance test initially and validation tests (signal-to-noise, stability, laser frequency calibration) as needed. Laser calibration requires NIST SRM 1921 polystyrene film and Agilent consultation. Base delivery includes spectrometer, accessory, manual, power cord, USB cable, installation disk; options not included: laptop, surfactant kit, extra cords, single-bounce ATR with sample press, multi-bounce ATR.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent 5500 Series FTIR Operational Manual
Manual ·40 pp.
View PDF