Agilent Technologies TwisTorr 84 FS — Turbomolecular Vacuum Pumps.
The TwisTorr 84 FS is a turbomolecular pump with self-positioning aluminium alloy stator disks and parallel spiral pumping channels, available with ISO 63, CFF 4.5", CFF 2.75" or KF 40 inlet flanges. It delivers 67 l/s N2 and 53 l/s H2 at 81,000 rpm, with foreline tolerance of 14 mbar N2/Ar, 12 mbar He and 4 mbar H2. It is intended for high-throughput N2, Ar and lighter gases in leak-free, baked systems requiring base pressure below 5×10⁻¹⁰ mbar.
Direct drop-in replacements
Full specifications
| Base Pressure With Recommended Forepump | 5 mbar |
| Pumping Speed Ar | 66 l/s |
| Pumping Speed H2 | 53 l/s |
| Pumping Speed He | 63 l/s |
| Pumping Speed N2 | 67 l/s |
| Rotational speed | 81000 rpm |
| Run-up time | 2 minutes |
| Max gas throughput N2 mfr | 100 sccm |
| Max gas throughput Ar mfr | 70 sccm |
| Compression ratio N2 mfr | 1 |
| Compression ratio He mfr | 2 |
| Compression ratio H2 mfr | 5 |
| Compression ratio Ar mfr | 1 |
| Foreline Tolerance N2 mfr | 14 mbar |
| Foreline Tolerance He mfr | 12 mbar |
| Foreline Tolerance H2 mfr | 4 mbar |
| Foreline Tolerance Ar mfr | 14 mbar |
| Operating position mfr | Any Any |
| Lubricant mfr | Permanent lubrication lubrication |
| Installation category mfr | II II |
| Max gas throughput at full rotational speed N2 mfr | 100 sccm |
| Max gas throughput at full rotational speed Ar mfr | 70 sccm |
| N2 mfr | 14 mbar |
| He mfr | 12 mbar |
| H2 mfr | 4 mbar |
| Ar mfr | 14 mbar |
| X3508-64001 mfr | 84 |
| X3508-64002 mfr | 84 Profibus |
| X3510-64000 mfr | 84 PCB |
| Description of the TwisTorr 84 FS mfr | 291 |
| Pump Description mfr | 293 |
| Technical Specification mfr | 295 |
| TwisTorr 84 FS Outline mfr | 298 |
| Interconnections mfr | 314 |
| Earth (Ground) Connection mfr | 314 |
| Connection C – Electrical mfr | 315 |
| Soft Start mfr | 316 |
| Vent Accessories mfr | 326 |
| Vibration Isolator Installation mfr | 328 |
| Purge Valve Installation mfr | 329 |
| Vibration isolator CFF 4.5” mfr | 9699376 |
| Vent Valve N.O. 0,5 mm orifice (0,5 m cable) mfr | 9699844 |
| Vent Valve extension cable (5 m) mfr | 9699941 |
| Vent Screw M5 mfr | 3502 |
| Purge Screw mfr | 3502 |
| Purge valve 10 SCCM NW16KF - M12 mfr | 9699239 |
| Purge valve 10 SCCM 1⁄4 Swagelok - M12 mfr | 9699240 |
| Purge valve 20 SCCM NW16KF - M12 mfr | 9699241 |
| Purge valve 20 SCCM 1⁄4 Swagelok - M12 mfr | 9699242 |
| Purge valve 10 SCCM 1⁄4 Swagelok - 1⁄4 Swagelok mfr | 9699232 |
| Purge valve 20 SCCM 1⁄4 Swagelok - 1⁄4 Swagelok mfr | 9699236 |
| Metric screws kit mfr | 3502 |
| American screws kit mfr | 3502 |
| Pump extension cable (3 m) mfr | 9699942 |
| Maximum pressure above atmospheric mfr | 2 bar |
| Maximum storage period mfr | 12 meses |
| Weight | 2.05 kg |
| Bottom Mounting mfr | 330 |
| CFF4.5 Mounting kit mfr | 3502 |
| Input Voltage mfr | 76 Vac |
| Max input pwr mfr | 130 VA |
| Stand-by pwr mfr | 8 W |
| Max oper. power mfr | 70 W |
| X3509-64000 mfr | 84 V |
| X3509-64001 mfr | 84 V |
| Inlet flange | 4.5 O.D. |
| Foreline flange mfr | 16 NW |
| Inlet screen installation mfr | 317 |
| TwisTorr 84 FS On Board Controller Installation mfr | 330 |
| Connection of the High Vacuum Flange mfr | 333 |
| Installation of ISO-K Flange mfr | 333 |
| Installation of ConFlat Flange (CFF 2 3⁄4) mfr | 335 |
| Installation of ConFlat Flange (CFF 4 1⁄2) mfr | 335 |
| Mains cable NEMA plug, 3 m long mfr | 9699958 |
| Mains cable European plug, 3 m long mfr | 9699957 |
| Serial cable and A-Plus software mfr | 9699883 |
| Inlet screen CFF4.5” mfr | 3502 |
| Inlet screen, KF40 mfr | 9699309 |
| Inlet screen CFF 2.75” mfr | 9699328 |
| Controller side mounting bracket mfr | 3502 |
| Counter-flange torque rating mfr | 250 Nm |
| Coolant Water Minimum Flow | 65 l/h |
| Coolant water pressure mfr | 2...4 bar |
| Bakeout Temperature | 80 max |
| Operating Altitude | 3000 m |
| Operating temperature | 5 °C |
| Storage temperature | 40 °C |
| Cooling requirements mfr | 35 °C |
| Noise Pressure level (pump only) mfr | 40 |
| Pollution degree mfr | 2 |
| Air cooling mfr | 100 sccm |
| Water cooling mfr | 100 sccm |
| Air Cooling Kit Installation mfr | 321 |
| Water Cooling Kit Connection mfr | 323 |
| Metal Water Cooling kit mfr | 3502 |
| Plastic Water cooling kit mfr | 3502 |
| Air cooling kit (0,5 m cable) mfr | 9699290 |
| Air cooling kit extension cable (5 m) mfr | 9699940 |
| Inlet screen ISO63 mfr | 3502 |
| Vibration isolator ISO 63 mfr | 9699375 |
| ISO 63 K clamp tightening torque mfr | 22 Nm |
Description
Construction and series positioning
- The pump uses self-positioning aluminium alloy stator disks with parallel spiral pumping channels, driven by a 54 Vac three-phase, 1350 Hz motor at 81,000 rpm.
- Four inlet flange variants exist: ISO 63 (2.05 kg), CFF 4.5" (3.50 kg), CFF 2.75" (3.34 kg) and KF 40 (2.37 kg). The CFF 4.5" variant permits bakeout to 120 °C, while ISO 63 is limited to 80 °C.
- No sibling models are listed; this card covers the TwisTorr 84 FS only.
Behavior under load and media
- Pumping speeds are 67 l/s N2, 66 l/s Ar, 63 l/s He and 53 l/s H2. Maximum gas throughput at full speed is 100 sccm N2 and 70 sccm Ar.
- Foreline tolerance is 14 mbar for N2 and Ar, 12 mbar for He and 4 mbar for H2; the backing pump must keep foreline pressure below these values to maintain compression.
- Base pressure with the recommended forepump is <5×10⁻¹⁰ mbar (DIN 28 428, leak-free test dome, 48 h after bake-out).
- The pump is designed for high throughput of N2, Ar and lighter gases; it must not be used with corrosive gases or in explosive environments.
Installation, cooling and control
- The pump can be installed in any position and must be fixed by its inlet flange with a connection able to withstand 250 Nm around its axis; ISO 63 uses 4 × M10 clamps at 22 Nm.
- Cooling is mandatory before start: either air cooling kit 9699290 (24 Vdc fan, 12.8 l/s, connects to controller P4) or water cooling kit X3502-68002/X3502-68003 (65 l/h min, 15–25 °C, 2–4 bar).
- A thermistor near the upper bearing trips Error Code 2 above 60 °C and Error Code 1 below 0 °C; if the sensor is disconnected the pump will not start.
- Compatible controllers: X3508-64001 (AG Rack RS232/485), X3508-64002 (AG Rack Profibus), X3509-64000 (on board 110/220 V), X3509-64001 (on board 24 V), X3510-64000 (PCB).
- Soft Start is disabled by default and uses frequency steps 450, 1100, 1200, 1300 and 1350 Hz; enable it after protracted stops to improve bearing grease distribution.
- Protective earth requires an M5 screw, AWG 16 green/yellow wire, 2 Nm torque and impedance <0.1 Ω.
Integration & Operating Notes
The pump must always be cooled before use; running without the air cooling kit 9699290 or a water cooling kit X3502-68002/X3502-68003 can damage internal components. Water cooling requires at least 65 l/h at 15–25 °C and 2–4 bar. During bakeout, water cooling must remain active.
The backing pump must maintain foreline pressure below 14 mbar for N2 and Ar, 12 mbar for He and 4 mbar for H2; exceeding these values causes loss of compression. Recommended forepumps are Agilent DS 42, DS 102, IDP-7 and IDP-3. Maximum gas throughput at full speed is 100 sccm N2 and 70 sccm Ar.
The pump can be installed in any position but must be fixed by its inlet flange with a connection able to withstand 250 Nm around its axis. For ISO 63, use 4 × M10 high-strength steel clamps (e.g. IC63250DCMZ) at 22 Nm fixing torque to prevent detachment in case of rotor failure.
The 6-pin controller connector uses pins B, C, D for the 3-phase motor supply and pin E for pump ground. If the temperature sensor is disconnected, the pump will not start; the sensor automatically cuts off power above 60 °C. Verify sensor wiring if the pump fails to start.
Soft Start uses frequency steps 450, 1100, 1200, 1300 and 1350 Hz and is disabled by default. Enable it via software or Remote I/O command after a protracted stop to improve grease distribution in the bearings before reaching full rotational speed.
The TwisTorr 84 FS requires no maintenance. Any work must be carried out by authorized personnel. Before any work, disconnect from mains, vent the pump, wait for the rotor to stop and for surface temperature to fall below 50 °C. A Request for Return sheet is required before returning the pump for repair or advanced exchange.
Selection & Integration Q&A
Choose CFF 4.5" when you need bakeout to 120 °C and metal-sealed UHV joints for base pressure below 5×10⁻¹⁰ mbar; it weighs 3.50 kg. Choose ISO 63 for lower bakeout (80 °C max) and lighter weight (2.05 kg) with elastomer seals. KF 40 and CFF 2.75" are also available for other foreline/inlet standards.
Use only a compatible Agilent controller providing 54 Vac three-phase at 1350 Hz: X3508-64001 (AG Rack RS232/485), X3508-64002 (AG Rack Profibus), X3509-64000 (on board 110/220 V), X3509-64001 (on board 24 V) or X3510-64000 (PCB). Recommended forepumps are Agilent DS 42, DS 102, IDP-7 or IDP-3; the foreline must stay below 14 mbar N2/Ar, 12 mbar He and 4 mbar H2.
Install indoors at +5 °C to +35 °C ambient, 0–90 % RH non-condensing, max altitude 3000 m. Vent port pressure must be less than 2 bar above atmospheric. Bakeout is limited to 80 °C (ISO 63) or 120 °C (CFF 4.5") with water cooling active. Storage is -40 °C to +70 °C with 12-month shelf life.
Cooling is mandatory but not included: order air cooling kit 9699290 or water cooling kit X3502-68002/X3502-68003. Inlet screens X3502-68000, X3502-68001, 9699309 or 9699328 are optional and reduce pumping speed by 20% Ar, 18% N2, 7% He, 5% H2. The pump requires no maintenance; service only by authorized personnel after venting and cooling below 50 °C.