BINDER ED 400
BINDER ·Made in Germany · In Active Production

BINDER ED 400

BINDER ED 400 — Drying ovens with Temperature 5…300 °C.

The BINDER ED 400 is a 400 L natural-convection drying and heating chamber with a +5 to 300 °C working range, built for three-phase 400 V operation. It is the largest standard ED model, distinguished by its 1,000 mm interior width and 90 kg total load capacity. It serves batch drying, heating, and thermal conditioning where uniform, fan-free air circulation is required.

Temperature
5…300 °C

BINDER ED Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
ED 400 Current 5–3001076571102.5400
ED 115 5–30066054570.5114
ED 23 5–300135202849.520
ED 260 5–30037605094255
ED 56 5–30085655062.557
ED 720 5–300287085159743

Full specifications

Core Performance & Functional Specifications
Temperature 5...300 °C
Air change rate 10 x/h
Heating time to 150°C 71 min
Interior volume 400 L
Load per rack 35 kg (ED400-400V-I)
Nominal power 3.4 kW
Permitted load 90 kg (ED400-400V-I)
Temperature fluctuation 0.5 K
Temperature uniformity 3.0 K
Rear wall clearance 160 mm (ED400-400V-I)
Side wall clearance 100 mm (ED400-400V-I)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 31 min
Dimensions, Weight & Mechanical Design
Depth 765 mm (ED400-400V-I)
Height 1025 mm (ED400-400V-I)
Interior depth 520 mm (ED400-400V-I)
Interior height 800 mm (ED400-400V-I)
Interior width 1000 mm (ED400-400V-I)
Width 1235 mm (ED400-400V-I)
Net weight of the unit (empty) mfr 125 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 3 N/PE

Description

Construction and series positioning

Thermal behavior and performance

Installation and operation

Integration & Operating Notes

Electrical
Requires 400 V 3-phase supply with 16 A fuse

The ED 400 draws 3.4 kW and must be connected to a 3 N/PE 400 V 50 Hz supply protected by a 16 A fuse. A CEE 16/6 H 5-pin plug is provided. Ensure the supply can handle the starting current and that the earth connection is reliable to avoid damage or safety hazards.

Data Sheet Model ED 400 en.pdf, Electrical Specifications
Mounting
Maintain 160 mm rear and 100 mm side clearances

Insufficient clearance restricts airflow around the chamber, causing overheating of external components and reduced temperature uniformity. Install the unit on a level floor with at least 160 mm behind and 100 mm on each side. Do not block the ventilation slots.

Data Sheet Model ED 400 en.pdf, Installation Requirements
Process Limits
Total load must not exceed 90 kg, 35 kg per rack

Exceeding the permitted load can deform the racks or damage the chamber interior. Distribute samples evenly across shelves and ensure no single rack carries more than 35 kg. The total load includes racks and accessories. Overloading may also impair temperature uniformity.

Data Sheet Model ED 400 en.pdf, Technical Data
Control & I/O
RS 422 interface enables external control and monitoring

The RS 422 option (model 422) allows connection to a PC or BINDER controller for programming temperature profiles and logging data. Use shielded cables and ensure proper termination to avoid communication errors. This interface is not included in the base model unless specified.

Data Sheet Model ED 400 en.pdf, Options and Accessories
Maintenance
Regular cleaning and calibration recommended

To maintain temperature accuracy and avoid contamination, clean the interior regularly with mild detergent. Periodically verify temperature uniformity and calibration, especially after moving the unit or changing loads. Refer to the operating manual for detailed maintenance intervals.

Data Sheet Model ED 400 en.pdf, Maintenance

Selection & Integration Q&A

When should I choose the ED 400 over smaller ED series models like the ED 260?

Choose the ED 400 when your batch requires more than 260 L capacity or interior dimensions exceeding those of the ED 260. The ED 400 offers 400 L, 1,000 mm width, and 90 kg total load, making it suitable for large or numerous samples. For smaller loads, the ED 260 or ED 115 may be more energy-efficient and space-saving.

What are the power supply and connection requirements for the ED 400?

The ED 400 requires a 400 V, 3-phase/N/PE, 50 Hz supply with a 16 A fuse and a CEE 16/6 H 5-pin plug. Nominal power is 3.4 kW. For 208 V 60 Hz networks, the ED400UL-208V-I variant with NEMA L21-20 plug is available. Ensure the supply is properly grounded and can handle the load.

What maintenance and accessories are needed for the ED 400?

Regular cleaning of the interior and periodic calibration are recommended. The RS 422 interface is an option (model 422) for external control. Racks and shelves are available as accessories and must not exceed 35 kg each. No consumables are included in base delivery; order racks and interface cables separately as needed.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model ED 400 en
Datasheet ·496,7 КБ ·7 pp.