BINDER UF V 350
BINDER ·Made in Germany · In Active Production

BINDER UF V 350

BINDER UF V 350 — Ultra-low temperature freezers with Temperature -90…-40 °C.

The UF V 350 is a 402 L ultra-low temperature freezer with a two-stage hydrocarbon refrigerant cascade (R290/R170) reaching -90 °C, distinguished within the UF V series by its support feet and absence of CO2 emergency cooling or door access options. It serves as a standalone -80 °C-class storage unit in laboratory cold chains, with a 210 min cooling-down time and 1.6 kW single-phase 230 V supply.

Temperature
-90…-40 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER UF V Series — Model Comparison & Selection

Model Temperature, °CCooling down time, minInterior volume, LInterior width, mmTemperature fluctuation at 80 c, KTemperature uniformity at 80 c, K
UF V 350 Current -90–-4021040270643.5
UF V 500 -90–-403605018261.52.5
UF V 700 -90–-4045073611101.52.5

Full specifications

Core Performance & Functional Specifications
Temperature -90...-40 °C
Interior volume 402 L (all variants)
Nominal power 1.6 kW
Temperature fluctuation (-80 °C) 4.0 K
Temperature uniformity (-80 °C) 3.5 K
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 938 mm (all variants)
Height 1966 mm (all variants)
Interior width 706 mm (all variants)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse mfr 10.0 A
Phase (Nominal voltage) mfr 1
Operating Conditions, Environment & Protection
Cool-down time 210 min

Description

Construction and series positioning

Behavior under load and in operation

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must stay within ±10% of 230 V at 50 Hz

The chamber is rated 230 V 1~ 50 Hz with a 10.0 A unit fuse and 1.6 kW nominal power. Operation outside the ±10% voltage tolerance risks compressor and controller malfunction; where the site supply is unstable, the optional advanced voltage booster should be specified.

Type plate / p.11; Advanced voltage booster 4.7 / p.42
Mounting
Rear-wall spacers and support feet set the installation envelope

The UF V 350 uses support feet (unique to this model in the series) for leveling, and rear-wall spacers must be fitted to maintain the required distance from the wall. IP 20 rating prohibits installation in damp areas or where water exposure is possible.

Installation and Connections 4.3 / p.34; Type plate / p.11
Control & I/O
Wire the volt-free alarm relay to the site monitoring system

The RD4 provides a zero-voltage relay alarm output plus Ethernet and USB interfaces. Alarm conditions (safety controller, tolerance range, door open, power failure, sensor failure) can be forwarded to external monitoring via the relay contact or network; configure tolerance range and alarm delays accordingly.

Alarm Functions 14.5 / p.68; Ethernet 15 / p.69-74

Selection & Integration Q&A

When should I choose the UF V 350 over the UF V 500 or UF V 700?

Pick the 350 for 402 L capacity where CO2 emergency cooling and the door access system are not needed — those options exist only on the 500/700. The 350 is also the only UF V model with support feet for leveling; larger siblings suit higher inventory volumes.

What supply and monitoring infrastructure does the UF V 350 require?

A 230 V 1~ 50 Hz supply within ±10%, fused at 10.0 A, with CEE 7/7 plug, sized for 1.6 kW. For remote monitoring, use the Ethernet port (configurable IP/DNS) or the volt-free alarm relay; an optional analog temperature output is available.

What are the mounting and environmental limits for this freezer?

IP 20 per EN 60529 means no damp areas or water exposure. Fit rear-wall spacers and level via the support feet. The R290/R170 cascade is flammable, so the room needs ventilation and no ignition sources; max system pressure is 28 bar.

What maintenance and accessories are required beyond the base unit?

Annual cleaning and medical-vaseline lubrication of door handle, hinges and locking counterpart, plus annual electrical safety testing per EN 50678/50699. Storage racks and cryo boxes, DL1 data recorder, analog output and voltage booster are options not included in base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model UF V 350 en
Datasheet ·723,4 КБ ·8 pp.
View PDF
PDF
UF V (E3) 09 2026 en
Manual ·10,2 МБ ·127 pp.
View PDF