Peter Huber Kältemaschinenbau CC-E xd
Peter Huber Kältemaschinenbau CC-E xd — Thermostatic baths.
The CC-E xd is a 195 mm immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for 200–240 V single- or two-phase supply. Its 25–200 °C base range extends to 20–200 °C with water cooling or −30–200 °C with a refrigerator, positioning it as the cold-capable member of the Einhaengethermostate series. It circulates heat transfer fluid through external circuits at up to 22 l/min.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Einhaengethermostate Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Weight, kg | Ambient temperature limit, °C | Current consumption, A |
|---|---|---|---|---|---|---|
| CC-E xd Current | Pilot ONE | Ethernet, USB (Host u. D | 17 | 5 | 5–40 | 10 |
| KISS E | KISS | USB (Device), RS232 Inte | 14 | 4 | 5–40 | 10 |
Full specifications
| Maximum delivery flow rate | 22 l/min (pressure pump max.; 0, 4 bar) suction pump max.; 0, 25 bar: 17 l/min |
| Pump connection | M16x1 AG male (Option, #39593) (Option, #39593) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 01 K |
| Dimensions (W x D x H) mfr | 132 x 159 x 360 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Weight | 5 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz V, Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-E xd is an immersion circulator (Einhaengethermostat) with a 195 mm immersion depth, 132 × 159 × 360 mm housing and 5 kg weight. Unlike the plain CC-E sibling, the xd modification carries a variable-speed pressure and suction pump, enabling both delivery and return-side suction in closed external circuits. The KISS E is the simpler neighbor in the same series. Heating capacity scales with site voltage: 1.5 kW at 200 V, 1.6 kW at 208 V, 1.8 kW at 220 V, 2.0 kW at 230 V and 2.1 kW at 240 V.
Thermal behavior and media
- Base operating range 25…200 °C; 20…200 °C with water cooling; −30…200 °C with a refrigerator. The cooling method must be fixed at selection because it defines the achievable low limit.
- Temperature stability is stated as ±0.01 K, with a Pt100 external sensor connection available for measuring at the load rather than in the bath.
- Pump performance: pressure side up to 22 l/min at 0.4 bar; suction side up to 17 l/min at 0.25 bar. Speed is adjustable.
- Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is fitted.
Control, installation and operation
The Pilot ONE controller provides Ethernet, USB host/device and RS232 interfaces. Supply is 200–240 V 1~/2~ 50/60 Hz with maximum current consumption of 10 A; voltage tolerance is ±10% provided the frequency tolerance does not run in the opposite direction. External branch circuit protection to UL 489 is required, and single/two-phase units ship with a power cable and country-specific plug that must be specified at order. Enclosure rating is IP20, safety class III/FL, and EMC classification is Class A, Group 1 to EN55011. Ambient operating temperature must remain between 5 and 40 °C.
Integration & Operating Notes
The unit does not carry its own branch circuit protection to UL 489; the electrical panel design must include it. Plan the feeder and protective device accordingly, since the unit draws up to 10 A.
Without auxiliary cooling the range starts at 25 °C; water cooling lowers it to 20 °C and a refrigerator to −30 °C. The cooling option must be decided at selection, because it defines the achievable process temperature and cannot be inferred from the base unit.
When the Huber pump adapter is used, polyglycol must not be selected as heat transfer fluid. Confirm fluid compatibility with the pump circuit before filling, since an incompatible fluid can damage the pump.
Single- and two-phase units are delivered with a power cable and a country-specific plug. State the destination country when ordering so the correct plug is supplied; otherwise the unit may arrive with a plug unsuitable for the site.
Selection & Integration Q&A
Pick the xd when the process needs external circuit circulation with suction-side return and cold operation down to −30 °C with a refrigerator. The plain CC-E and KISS E are the simpler neighbors; the xd's variable-speed pressure/suction pump and Pilot ONE interfaces are its differentiators.
Provide 200–240 V 1~/2~ 50/60 Hz with ±10% voltage tolerance not opposed by frequency deviation, plus external UL 489 branch protection; the unit draws up to 10 A. Specify the country plug at order. Do not use polyglycol with the Huber pump adapter.