Shimadzu Nexera Lite Inert
Shimadzu ·Made in Japan · In Active Production

Shimadzu Nexera Lite Inert

Shimadzu Nexera Lite Inert — Chromatographs with Flow rate 0.5 milliliter/minute.

The Shimadzu Nexera Lite Inert is an inert high-performance liquid chromatograph (HPLC) configured for bio-inert separations, with a maximum operating pressure of 20 MPa. As a member of the Nexera Lite series, it prioritizes inert sample pathways for sensitive analytes while maintaining standard HPLC performance. It serves as the core separation module in analytical workflows requiring metal-sensitive compound compatibility.

Flow rate
0.5 milliliter/minute

Full specifications

Core Performance & Functional Specifications
Flow rate 0.5 mL/min (analytical conditions, gel filtration)
Column temperature 25 °C (analytical conditions)
Detection wavelength 280 nm (UV detection)
UV detector, deuterium lamp: 190 to 700 nm · UV detector, deuterium and tungsten lamps: 190 to 1000 nm · fluorescence detector, xenon lamp: 200 to 650 nm
Number of degassed solvents 3 (SCL-40)
CBM-40: 5
Retention time repeatability mfr 0.08 % RSD (n=5 injections)
Analog board channels mfr Up to 2 channels (option) (SCL-40)
Analog board channels mfr Up to 1 channel (CBM-40)
Injection volume accuracy mfr ≤ ±1 % (μL injection)
Containable column size and number mfr Up to 250 mm L. column × 6 or 300 mm L. column × (standard equipment)
Containable column size and number mfr Up to 100 mm L. column × 6 or 300 mm L. column × (without sample cooler)
Dimensions mfr W 260 × D 500 × H 415 mm (autosampler, standard equipment)
Dimensions mfr W 260 × D 509 × H 415 mm (autosampler without sample cooler)
Dimensions, Weight & Mechanical Design
Weight 27 kg (autosampler, standard equipment)
autosampler without sample cooler: 30 kg
Operating Conditions, Environment & Protection
Cell temperature setting range mfr 4 to 40 in 1 °C steps °C (fluorescence detector, low-pressure mercury lamp)

Description

Construction and Positioning

Integration & Operating Notes

Process Limits
Do not exceed 20 MPa system pressure

Operating above the rated maximum pressure risks damage to the inert flow path components and invalidates results. Ensure method parameters, including flow rate and column selection, keep backpressure below this limit under all expected conditions.

Factory Specification Sheet
Detector Integration
pH Monitor pHM-40 is a supported detector

The system lists the pH Monitor pHM-40 as a detector type, so integration with this module is anticipated. Verify compatibility of the monitor's flow cell and connections with the inert flow path to maintain inertness throughout the system.

Factory Specification Sheet

Selection & Integration Q&A

What detector options are compatible with the Nexera Lite Inert?

The pH Monitor pHM-40 is explicitly listed as a detector type. Ensure the detector's flow path materials match the inertness requirements of the system to avoid introducing metal contact.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
c190-e292