CB S (E7.1) 02 2026

Technical Reference Summary

Covered Models & Document Scope

This operating manual (CB-S / CB-S-UL (E7.1), issue 02/2026, Art. No. 7001-0763) covers the BINDER CO2 Incubator series CB-S and CB-S-UL with FPI-sensor system and RD4 controller. Models: CB-S 170 (CBS170-230V, Art. 9040/9140-0219), CB-S 170 with left-hinged door (CBS170-230V-L, 200-230 V, Art. 9040/9140-0220), CB-S 170-UL (CBSUL170-120V, Art. 9040/9140-0221), CB-S 170-UL with left-hinged door (CBS170UL-120V-L, 100-120 V, Art. 9040/9140-0222), CB-S 260 (CBS260-230V, Art. 9040/9140-0223), and CB-S 260-UL (CBSUL260-120V, Art. 9040/9140-0224).

Section-by-Section Technical Breakdown

pp. 20-24
Chamber description

Describes the chamber overview, inner chamber, rear connection panel, chamber doors, and instrument panel. The inner chamber accommodates a maximum of 6 shelves for CB-S 170 and 8 shelves for CB-S 260. The chamber can be connected to an intranet with the APT-COM 4 Multi Management Software (accessory).

  • Maximum 6 shelves in CB-S / CB-S-UL 170; maximum 8 shelves in CB-S / CB-S-UL 260.
  • Connection panel located on the rear of the chamber.
  • Instrument panel and chamber doors described.
  • Intranet connection possible with APT-COM 4 Multi Management Software (accessory, chap. 20.4).
pp. 28-33
Installation and connections

Covers shelf installation, CO2 sensor connection, water pan, gas connection including CO2 gas cylinder and gas hose, and electrical connection. The CO2 sensor is temperature resistant up to a maximum of 60 °C / 140 °F. Gas connection includes a gas cylinder connection kit accessory.

  • CO2 sensor maximum temperature: 60 °C / 140 °F.
  • Gas connection includes connecting the CO2 gas cylinder and gas hose to the chamber rear.
  • Gas cylinder connection kit available as accessory.
  • Electrical connection described in chap. 4.5.
pp. 34-36
Functional overview of the RD4 chamber controller

Describes the menu structure of the RD4 controller, access levels, and performance during and after power failures and shut down. The controller manages temperature and CO2 set-points, alarms, and network settings.

  • Menu structure and access levels described.
  • Performance during and after power failures and shut down described.
  • Controller supports Ethernet network settings and data recorder functions.
pp. 36-38
Start up

Covers turning on the chamber, preset factory parameters, performance after turning on, and altitude of the installation site above sea level. Factory parameters are preset; performance after power-on is described.

  • Turning on the chamber via main power switch.
  • Preset factory parameters listed.
  • Performance after turning on the chamber described.
  • Altitude of installation site above sea level must be considered.
pp. 39
Temperature and CO2 set-point entry

Describes how to enter temperature and CO2 set-points on the RD4 controller. These set-points define the incubation conditions.

  • Temperature set-point entry procedure.
  • CO2 set-point entry procedure.
pp. 41-42
Setting special controller functions

Covers Idle mode and deactivated CO2 control. Idle mode can be activated or deactivated; when active, set-point values are not equilibrated and icon '1' is displayed.

  • Idle mode activation/deactivation (chap. 9.1).
  • Deactivated CO2 control (chap. 9.2).
  • When Idle mode is active, set-point values are not equilibrated and icon '1' is displayed.
pp. 42-43
Password

Describes password request and how to assign and modify User and Admin passwords. Access to certain controller functions is protected by passwords.

  • Password request procedure.
  • Assign and modify User password.
  • Assign and modify Admin password.
pp. 44-46
Temperature safety devices

Describes the over temperature protective device (class 1) and the overtemperature safety controller class 3.1. Includes setting the safety controller mode and value, messages and measures in the state of alarm, and function check.

  • Over temperature protective device class 1 (chap. 11.1).
  • Overtemperature safety controller class 3.1 (chap. 11.2).
  • Setting safety controller mode and value.
  • Message and measures in the state of alarm (chap. 11.2.3).
  • Function check (chap. 11.2.4).
pp. 47-50
General controller settings

Covers selecting menu language, temperature unit, setting date and time, language selection at restart, chamber address, and display brightness.

  • Menu language selection.
  • Temperature unit selection.
  • Setting current date and time.
  • Function 'Language selection at restart'.
  • Setting chamber address.
  • Display brightness adjustment.
pp. 51-52
Tolerance range settings

Describes setting the delay time for tolerance range alarm, temperature tolerance range, CO2 tolerance range, and delay time for door open alarm.

  • Delay time for tolerance range alarm.
  • Temperature tolerance range setting.
  • CO2 tolerance range setting.
  • Delay time for door open alarm.
pp. 53
Chamber settings (for experienced users only)

Covers setting the humidity control. Intended for experienced users only.

  • Setting the humidity control (chap. 14.1).
pp. 53-62
Notification and alarm functions

Describes alarm messages, information messages, activating/deactivating the audible alarm, and measures in case of alarm including door open alarm, safety controller temperature alarm, temperature tolerance range alarm, CO2 tolerance range alarm, CO2 pressure alarm, power failure alarm, temperature sensor failure alarms, CO2 sensor failure alarms, and alarm on ineffective sterilization. Also covers the zero-voltage relay alarm output.

  • Alarm messages and information messages listed.
  • Audible alarm (alarm buzzer) can be activated/deactivated.
  • Measures for door open alarm, safety controller temperature alarm, temperature tolerance range alarm, CO2 tolerance range alarm, CO2 pressure alarm, power failure alarm, temperature sensor failures, CO2 sensor failure, and ineffective sterilization.
  • Zero-voltage relay alarm output (chap. 15.5).
pp. 63-68
Ethernet network settings

Covers showing network settings (MAC address, IP address, subnet mask, standard gateway, DNS server address, DNS chamber name) and changing the configuration (automatic/manual IP assignment, DNS server assignment, IP address, subnet mask, standard gateway, DNS server address).

  • Show MAC address, IP address, subnet mask, standard gateway, DNS server address, DNS chamber name.
  • Select automatic or manual IP address assignment.
  • Select automatic or manual DNS server address assignment.
  • Assign IP address, subnet mask, standard gateway, DNS server address.
pp. 68-69
Data recorder

Describes recorded data, storage capacity, setting the storage rate for the 'DL1' recorder data, and deleting the data recorder.

  • Recorded data described.
  • Storage capacity specified.
  • Setting storage rate for 'DL1' recorder data.
  • Deleting the data recorder.
pp. 70-72
USB menu: data transfer via USB interface

Covers connecting a USB stick, import function, export functions, ongoing data transfer, error during data transmission, and removing the USB stick.

  • Connecting USB stick.
  • Import function.
  • Export functions.
  • Ongoing data transfer.
  • Error during data transmission.
  • Removing the USB stick.
pp. 72-74
Reference measurements

Describes CO2 reference measurement indirectly via pH of cell medium, directly via chemical indicator tubes, and directly with an electronic infrared measuring device; also temperature reference measurement.

  • CO2 reference measurement via pH of cell medium.
  • CO2 reference measurement via chemical indicator tubes.
  • CO2 reference measurement with electronic infrared measuring device.
  • Temperature reference measurement.
pp. 75-76
Options and accessories

Lists silicone access ports (30 mm / 1.18 in, closable with 2 silicone plugs; Art. 8012-0558 rear, 8012-0559 left, 8012-0560 right), base on castors, flat stacking frame, APT-COM 4 Multi Management Software, and analog outputs for temperature and CO2.

  • Silicone access ports 30 mm / 1.18 in, closable with 2 silicone plugs: 8012-0558 rear, 8012-0559 left, 8012-0560 right (option).
  • Base on castors (accessory).
  • Flat stacking frame (accessory).
  • APT-COM 4 Multi Management Software (accessory).
  • Analog outputs for temperature and CO2 (option).
pp. 77-79
Avoiding microbial contamination

Covers cells and media, laboratory conditions/equipment around the CO2 incubator, working and behavior in the lab, chamber design and equipment, and handling the CO2 incubator to avoid microbial contamination.

  • Cells and media handling.
  • Laboratory conditions and equipment around the CO2 incubator.
  • Working and behavior in the lab.
  • Chamber design and equipment.
  • Handling the CO2 incubator.
pp. 80-83
Cleaning, decontamination / disinfection, and sterilization

Describes cleaning, decontamination/chemical disinfection of the chamber, and disinfection of the CO2 sensor.

  • Cleaning procedure.
  • Decontamination / chemical disinfection of the chamber.
  • Disinfection of the CO2 sensor.
pp. 84-89
Hot-air sterilization

Covers overview, preparations, starting and running the hot-air sterilization cycle, performance during sterilization, completing the cycle, and premature termination effects and procedure. Premature termination after less than 4 hours is ineffective; after more than 4 hours (during cooling-down phase) sterilization is successful.

  • Preparations for hot-air sterilization.
  • Starting sterilization (chap. 23.3.1).
  • Performance during sterilization (chap. 23.3.2).
  • Completing the entire sterilization cycle (chap. 23.3.3).
  • Premature termination after less than 4 hours: ineffective sterilization (chap. 23.4.1).
  • Premature termination after more than 4 hours, i.e., during cooling-down phase: successful sterilization (chap. 23.4.2).
  • Cancelling sterilization via controller menu, opening outer door, or turning off chamber (chap. 23.5).
pp. 90-96
Maintenance, and service, troubleshooting, repair, testing

Covers general information and personnel qualification, maintenance intervals and service, service reminder, gas inlet fine filter, simple troubleshooting (general, temperature, CO2, humidity, controller, sterilization), and sending the chamber back to BINDER GmbH. Electrical testing requires annual repeat inspection and test before initial startup and before resuming operation after maintenance or repair, recommended under EN 50678:2020 and EN 50699:2020.

  • Maintenance intervals and service (chap. 24.2).
  • Service Reminder (chap. 24.3).
  • Gas inlet fine filter: aseptic filter, filtration efficiency 99.99%, particle size 0.45 µm; checked at each maintenance interval and replaced at least once a year (chap. 24.4).
  • Simple troubleshooting tables for general, temperature, CO2, humidity, controller, and sterilization (chap. 24.5).
  • Annual repeat electrical inspection and test before initial startup and before resuming operation after maintenance or repair; recommended under EN 50678:2020 and EN 50699:2020 (chap. 24.1).
  • Sending the chamber back to BINDER GmbH (chap. 24.6).
pp. 101-107
Technical description

Covers factory calibration and adjustment, over current protection, definition of usable volume, CB-S / CB-S-UL technical data, important conversion data for non-SI units, conversion table for gas inlet pressures (bar – psi), equipment and options (extract), options/accessories/spare parts (extract), and dimensions.

  • Factory calibration and adjustment (chap. 26.1).
  • Over current protection (chap. 26.2).
  • Definition of usable volume (chap. 26.3).
  • CB-S / CB-S-UL technical data (chap. 26.4).
  • Conversion data for non-SI units (chap. 26.5).
  • Conversion table for gas inlet pressures, bar – psi (chap. 26.6).
  • Equipment and options extract (chap. 26.7).
  • Options, accessories, and spare parts extract (chap. 26.8).
  • Dimensions (chap. 26.9).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
CO2 sensor is temperature resistant up to a maximum temperature of 60 °C / 140 °F. p.24 ↗ Do not expose the CO2 sensor to temperatures above 60 °C / 140 °F; relevant during hot-air sterilization and cleaning procedures. CB-S / CB-S-UL series
Gas inlet fine filter: aseptic filter with filtration efficiency 99.99% and particle size 0.45 µm; checked at each maintenance interval and replaced at least once a year. p.24 ↗ Maintenance plan must include annual replacement of the gas inlet fine filter to prevent dirt accumulation in gas inlet valves and tubes. CB-S / CB-S-UL series
Electrical testing: annual repeat inspection as well as test prior to initial startup and prior to resuming operation after maintenance or repair; recommended under EN 50678:2020 and EN 50699:2020. p.24 ↗ Schedule annual electrical safety tests and post-maintenance tests before returning chamber to operation. CB-S / CB-S-UL series
Maximum shelves: 6 for CB-S / CB-S-UL 170 and 8 for CB-S / CB-S-UL 260. p.2 ↗ Do not exceed shelf count limits when loading the chamber. CB-S 170 / CB-S 260 variants
Nominal temperature 180 °C / 356 °F; nominal power 1.30 kW; nominal current 5.7 A for 200-230 V models and 10.9 A for 100-120 V models; IP protection 20; temperature safety device class 3.1 per DIN 12880. p.1 ↗ Electrical supply and overcurrent protection must be sized for 5.7 A (230 V models) or 10.9 A (120 V models); IP20 requires dry indoor installation. CB-S 170 (9040-0219) and CB-S 170-UL (9040-0221) as shown on type plates
Nominal voltage range 200-230 V / 50 Hz or 60 Hz (+/- 10%) for 230 V models; 100-120 V / 50 Hz or 60 Hz for UL models; current type 1 N ~. p.1 ↗ Verify supply voltage and frequency within tolerance before connection; UL models require 100-120 V supply. CB-S 170/260 (200-230 V) and CB-S 170-UL/260-UL (100-120 V)
Premature termination of hot-air sterilization after less than 4 hours is ineffective; after more than 4 hours (during cooling-down phase) sterilization is successful. p.23 ↗ Do not abort sterilization before 4 hours have elapsed if sterilization effectiveness is required. CB-S / CB-S-UL series
Silicone access ports 30 mm / 1.18 in, closable with 2 silicone plugs; part numbers 8012-0558 (rear), 8012-0559 (left), 8012-0560 (right). p.20 ↗ Order correct port part number for desired mounting side when specifying options. CB-S / CB-S-UL series (option)
Idle mode: when activated, set-point values are not equilibrated and icon '1' is displayed. p.9 ↗ Deactivate Idle mode (chap. 9.1) if set-point values fail to equilibrate and icon '1' appears. CB-S / CB-S-UL series with RD4 controller
Over temperature protective device (class 1) responds when nominal temperature is exceeded by 10 °C due to chamber failure. p.24 ↗ If chamber without function and nominal temperature exceeded by 10 °C, the class 1 over temperature protective device has responded; contact BINDER service. CB-S / CB-S-UL series
Miniature fuse for overcurrent protection may trigger; check chamber fuse and replace if appropriate; if it responds again, contact BINDER service. p.24 ↗ Repeated fuse blowing indicates a fault requiring service intervention. CB-S / CB-S-UL series
Zero-voltage relay alarm output available (chap. 15.5). p.15 ↗ Can be used to connect external alarm devices or monitoring systems. CB-S / CB-S-UL series with RD4 controller
Analog outputs for temperature and CO2 available as option (chap. 20.5). p.20 ↗ Option enables external data logging or integration with monitoring systems. CB-S / CB-S-UL series (option)
Data recorder storage rate for 'DL1' recorder data can be set; data recorder can be deleted. p.17 ↗ Configure storage rate according to required recording interval and storage capacity. CB-S / CB-S-UL series with RD4 controller
Ethernet network settings include MAC address, IP address, subnet mask, standard gateway, DNS server address, and DNS chamber name; IP and DNS assignment can be automatic or manual. p.16 ↗ Chamber can be integrated into laboratory network for remote monitoring via APT-COM 4. CB-S / CB-S-UL series with RD4 controller
CO2 reference measurement can be performed indirectly via pH of cell medium, directly via chemical indicator tubes, or directly with an electronic infrared measuring device. p.19 ↗ Multiple methods available for verifying CO2 concentration accuracy. CB-S / CB-S-UL series
Temperature reference measurement described (chap. 19.2). p.19 ↗ Use for verifying chamber temperature accuracy against independent reference. CB-S / CB-S-UL series
Gas inlet pressure conversion table bar – psi provided (chap. 26.6). p.26 ↗ Use to convert gas supply pressure requirements between bar and psi. CB-S / CB-S-UL series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Chamber without function. p.92 ↗ No power supply. Check connection to power supply. Check whether the chamber is turned on at the main power switch. chap. 24.5.1
Chamber without function. p.92 ↗ Wrong voltage. Check power supply for voltage of 120 V / 230 V. chap. 24.5.1
Chamber without function. p.92 ↗ Nominal temperature exceeded by 10 °C due to chamber failure. Over temperature protective device (class 1) responds (chap. 11.1). Contact BINDER service. chap. 24.5.1
Chamber without function. p.92 ↗ The miniature fuse for overcurrent protection has triggered (chap. 26.2). Check chamber fuse and replace it if appropriate. If it responds again, contact BINDER service. chap. 24.5.1
Chamber without function. p.92 ↗ Controller defective. Contact BINDER service. chap. 24.5.1
Set-point values are not equilibrated. Icon “1” is displayed. p.92 ↗ Operating mode “Idle mode” activated. Deactivate operating mode “Idle mode” (chap. 9.1). chap. 24.5.1
Alarm message “Door open”. p.92 ↗ The outer chamber door is open. Close chamber door. chap. 24.5.1