BINDER CB-S 170
BINDER ·Made in Germany · In Active Production

BINDER CB-S 170

BINDER CB-S 170 — Incubators with Temperature 6…50 °C.

The CB-S 170 is a 170 L CO2 incubator with IR CO2 sensing and hot-air sterilization, configured for 200–230 V single-phase supply. It holds +6 to 50 °C with 0.1 K fluctuation and 0.3 K uniformity, and recovers to setpoint in 6 min after a 30 s door opening at 37 °C. It serves as the standard-capacity chamber for cell culture where 170 L and six shelves suffice.

Temperature
6…50 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER CB-S Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LPermitted load, kgPower frequency, Hz
CB-S 170 Current 6–5071.5871703050–60
CB-S 260 6–5078.51022674050

Full specifications

Core Performance & Functional Specifications
Temperature 6...50 °C
CO2 measuring technology IR (all variants)
CO2 recovery time 6 min (all variants)
Interior volume 170 L (all variants)
Load per rack 10 kg (all variants)
Nominal power 1.3 kW (all variants)
Permitted load 30 kg (all variants)
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Back wall clearance 100 mm (all variants)
Side wall clearance 50 mm (all variants)
Article Number mfr 9040
Recovery time after door was opened for 30 s at 37°C mfr 6 min
Dimensions, Weight & Mechanical Design
Depth 715 mm (all variants)
Height 870 mm (all variants)
Width 680 mm (all variants)
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (all variants)
Rated voltage 200...230 V (CBS170-230V / CBS170-230V-L)
Unit fuse 10.0 A (CBS170-230V / CBS170-230V-L)
Operating Conditions, Environment & Protection
Humidity range 90...95 % RH

Description

Construction and series positioning

Behavior under load and in operation

Installation, control, and operation

Integration & Operating Notes

Electrical
Supply must be 200–230 V, 50/60 Hz, 1 N~ with 10 A fuse

The 230 V model draws 5.7 A nominal and is protected by a 10.0 A unit fuse; connecting to a 100–120 V supply or a circuit rated below the fuse will prevent startup or cause repeated fuse trips. Verify voltage and frequency within ±10 % before first power-on.

Operating Manual CB-S (E7.1), chap. 1.6, pages 10–11
Mounting
Keep 100 mm rear and 50 mm side clearance for airflow

The chamber requires 100 mm clearance at the back and 50 mm at each side; insufficient clearance restricts heat dissipation and can raise internal temperatures or trip the over-temperature protective device. IP20 rating also mandates a dry indoor location.

Operating Manual CB-S (E7.1), Technical Data and Installation sections
Process Limits
Do not exceed six shelves or 30 kg total load

The CB-S 170 inner chamber accepts a maximum of six shelves, with 10 kg per rack and 30 kg total permitted load. Exceeding these limits can deform shelves or compromise temperature uniformity; the CB-S 260 is the alternative when more capacity is needed.

Operating Manual CB-S (E7.1), chap. 2, page 20, and Technical Data
Maintenance
Replace gas inlet fine filter at least once per year

The aseptic gas inlet fine filter (0.45 µm, 99.99 % efficiency) must be checked at each maintenance interval and replaced at least annually to prevent dirt accumulation in gas inlet valves and tubes. Schedule this with the annual electrical safety test per EN 50678:2020 and EN 50699:2020.

Operating Manual CB-S (E7.1), chap. 24.4, page 92, and chap. 24.1, page 90
Control & I/O
Zero-voltage relay can drive external alarm or monitoring

The RD4 controller provides a zero-voltage relay alarm output (chap. 15.5) that can be wired to an external alarm device or building monitoring system. Plan the contact rating and wiring during installation; optional analog outputs for temperature and CO2 are separate accessories.

Operating Manual CB-S (E7.1), chap. 15.5, page 62, and chap. 20.5, page 76
Process Limits
Hot-air sterilization must run at least 4 hours to be effective

Premature termination of the hot-air sterilization cycle after less than 4 hours is ineffective; only after more than 4 hours (during the cooling-down phase) is sterilization successful. The CO2 sensor must not exceed 60 °C during any procedure, so verify sensor protection before starting a cycle.

Operating Manual CB-S (E7.1), chap. 23.4, pages 87–88, and chap. 24, page 90

Selection & Integration Q&A

When should I choose the CB-S 170 over the CB-S 260 or the left-hinged/UL variants?

Pick the CB-S 170 when 170 L and six shelves cover your workload; the CB-S 260 adds volume and two more shelves at the same temperature range. Choose the -L variant if door swing must be left-hand, and the UL variant only for 100–120 V supplies.

What supply and gas connections does the CB-S 170 require for integration?

It needs 200–230 V, 50/60 Hz, 1 N~ with a 10.0 A fuse (5.7 A nominal). CO2 is supplied from a gas cylinder via the gas hose and connection kit (accessory); the inlet fine filter is 0.45 µm. Ethernet and zero-voltage relay are available for monitoring.

What mounting and duty-cycle limits apply to the CB-S 170?

Install indoors (IP20) with 100 mm rear and 50 mm side clearance. Load max 10 kg per rack and 30 kg total across six shelves. Hot-air sterilization must run at least 4 hours; the CO2 sensor must stay below 60 °C.

What maintenance and accessories are not included with the base CB-S 170?

Base delivery excludes the CO2 gas cylinder connection kit, silicone access ports (8012-0558/0559/0560), base on castors, stacking frame, APT-COM 4 software, and analog outputs. Plan annual gas inlet filter replacement and electrical safety testing per EN 50678/50699.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model CB S 170 en
Datasheet ·417,5 КБ ·8 pp.