BINDER CB-S 260
BINDER CB-S 260 — Incubators with Temperature 6…50 °C.
The CB-S 260 is a 267 L CO₂ incubator with IR CO₂ sensing and hot-air sterilization, configured for 200–230 V single-phase supply (CBS260-230V). It extends the CB-S series to 8 shelves and 40 kg total load, with 0.1 K fluctuation and 0.4 K uniformity at +6 to 50 °C. It serves as the controlled incubation stage for cell culture where contamination control and CO₂ recovery (5 min) matter.
Direct drop-in replacements
BINDER CB-S Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Permitted load, kg | Power frequency, Hz |
|---|---|---|---|---|---|---|
| CB-S 260 Current | 6–50 | 78.5 | 102 | 267 | 40 | 50 |
| CB-S 170 | 6–50 | 71.5 | 87 | 170 | 30 | 50–60 |
Full specifications
| Temperature | 6...50 °C |
| CO2 measuring technology | IR |
| Interior volume | 267 L (CBS260-230V / CBS260UL-120V) |
| Load per rack | 10 kg (both variants) |
| Nominal power | 1.5 kW |
| Permitted load | 40 kg (both variants) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.4 K |
| Back wall clearance | 100 mm (both variants) |
| Side wall clearance | 50 mm (both variants) |
| Article Number mfr | 9040 |
| Recovery time after door was opened for 30 s at 37°C mfr | 8 min |
| CO2 recovery time after door was opened for 30 s at 5 vol. % CO2 mfr | 5 min |
| Depth | 785 mm (both variants) |
| Height | 1020 mm (both variants) |
| Width | 740 mm (both variants) |
| Power frequency | 50 Hz |
| Rated voltage | 200...230 V |
| Unit fuse | 10.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Humidity range | 90...95 % RH |
Description
Construction and series positioning
- The CB-S 260 (CBS260-230V, Art. 9040-0223) is the larger chamber of the CB-S series; the sibling CB-S 170 shares the +6 to 50 °C range but holds 6 shelves versus 8 here.
- Interior volume is 267 L with net dimensions 740 mm wide × 785 mm deep × 1,020 mm high; the chamber accepts up to 8 shelves at 10 kg per rack and 40 kg total permitted load.
- CO₂ measurement uses an IR sensor; humidity is held at 90–95 % RH via the water pan. Hot-air sterilization is integrated for chamber decontamination.
Behavior under load and in operation
- Temperature control is specified at 0.1 K fluctuation and 0.4 K uniformity across the +6 to 50 °C working range.
- After a 30 s door opening at 37 °C, temperature recovers in 8 min; CO₂ recovers in 5 min at 5 vol. % CO₂.
- The IR CO₂ sensor is rated to a maximum of 60 °C / 140 °F, so it must not be exposed above this during sterilization or cleaning.
- Hot-air sterilization is ineffective if terminated before 4 hours; termination after 4 hours (cooling phase) still yields successful sterilization.
Installation, control and operation
- The RD4 controller manages temperature and CO₂ set-points, alarms, tolerance ranges, and network settings; access levels are password-protected (User/Admin).
- Electrical supply: 200–230 V, 1~, 50/60 Hz, 1.5 kW nominal, 10.0 A unit fuse; the 230 V variant uses a CEE 7/7 plug.
- Rear and side wall clearances are 100 mm and 50 mm respectively. The chamber can join an Ethernet network for APT-COM 4 monitoring; a zero-voltage relay alarm output and optional analog outputs for temperature and CO₂ are available.
- A gas inlet fine filter (0.45 µm, 99.99 % efficiency) protects the gas path and is replaced at least annually.
Integration & Operating Notes
The CBS260-230V draws 1.5 kW nominal and requires a 200–230 V single-phase 50/60 Hz supply protected by a 10.0 A fuse. Verify voltage and frequency tolerance before connection; the 100–120 V UL variant is a separate model (CBS260UL-120V) and cannot be fed from the same circuit.
The IR CO₂ sensor is temperature resistant only up to 60 °C / 140 °F. During hot-air sterilization and any cleaning procedure, ensure the sensor is not exposed above this limit; exceeding it risks sensor damage and loss of CO₂ control.
Hot-air sterilization is ineffective if terminated before 4 hours have elapsed. If aborted after more than 4 hours (during the cooling-down phase), sterilization is still considered successful. Plan the duty cycle so the chamber is not needed for incubation during this window.
The aseptic gas inlet fine filter (0.45 µm, 99.99 % efficiency) is checked at each maintenance interval and replaced at least once a year. Skipping replacement allows dirt to accumulate in gas inlet valves and tubes, compromising CO₂ supply and sterility.
An annual repeat electrical inspection is required, plus a test before initial startup and before resuming operation after maintenance or repair, recommended under EN 50678:2020 and EN 50699:2020. Schedule these tests so the chamber is not returned to service without a valid safety check.
The RD4 controller provides a zero-voltage relay alarm output (chap. 15.5) that can be wired to external alarm devices or a monitoring system. Use it to propagate door-open, temperature, CO₂, or power-failure alarms to a central station; verify contact rating before connecting loads.
Network settings include MAC address, IP address, subnet mask, gateway, and DNS; IP assignment can be automatic or manual. Integrating the chamber into the lab network enables remote monitoring and data logging with APT-COM 4 software, which is an accessory and not included in base delivery.
Selection & Integration Q&A
Pick the CB-S 260 when you need 267 L interior volume and up to 8 shelves (40 kg total load) rather than the CB-S 170's 6 shelves. Both share the +6 to 50 °C range and IR CO₂ sensing; the 260 suits larger culture batches, while the 170 fits smaller benches.
The chamber needs a CO₂ gas cylinder and gas hose (gas cylinder connection kit is an accessory). The gas inlet fine filter (0.45 µm) is built in and must be replaced annually. Verify inlet pressure using the bar–psi conversion table in the manual before connecting.
Base delivery excludes APT-COM 4 software, analog outputs, silicone access ports (e.g., 8012-0558 rear), base on castors, and stacking frame. Plan for annual gas inlet fine filter replacement and annual electrical safety testing per EN 50678/50699.