KB (E7) 09 2026
Technical Reference Summary
Covered Models & Document Scope
Covers BINDER KB and KB-UL (E7) Cooling Incubators with compressor technology and RD4 controller, sizes 65, 130, 260, 470, 720 and 1060. Each size has a standard (KB) and a UL variant (KB-UL). Article numbers: KB 65 9020-0471/9120-0471, KB 65-UL 9020-0491/9120-0491, KB 130 9020-0472/9120-0472, KB 130-UL 9020-0500/9120-0500, KB 260 9020-0473/9120-0473, KB 260-UL 9020-0501/9120-0501, KB 470 9020-0474/9120-0474, KB 470-UL 9020-0502/9120-0502, KB 720 9020-0475/9120-0475, KB 720-UL 9020-0503/9120-0503, KB 1060 9020-0476/9120-0476, KB 1060-UL 9020-0504/9120-0504. Options include analog temperature outputs, zero-voltage alarm contacts, door lock, internal LED light, flexible Pt 100 object temperature sensor, APT-COM 4 software, racks with U-rails or telescopic rails, and a flexible tilt protection kit for KB/KB-UL 65, 130 and 470.
Section-by-Section Technical Breakdown
Describes the chamber overview, the connection panel on the chamber rear and the triangular instrument panel. The chambers are equipped with a temperature safety device and interfaces for computer communication (Ethernet) and USB; APT-COM 4 software permits networking up to 100 chambers.
- Connection panel is located on the chamber rear
- Triangular instrument panel houses the RD4 controller
- Ethernet and USB interfaces for computer communication
- APT-COM 4 Multi Management Software can network up to 100 chambers
Covers location and ambient conditions, wall spacer from size 260, mounting of the flexible tilt protection kit for KB/KB-UL 65, 130 and 470, the condensate collection pan, rack installation and electrical connection. Minimum distances and ventilation requirements must be observed.
- Spacer for wall distance required from size 260 on
- Flexible tilt protection kit mounted on KB/KB-UL 65, 130 and 470
- Condensate collection pan must be installed
- Racks available with U-rails or telescopic rails (accessory)
- Electrical connection per type plate
Explains the menu structure and access levels of the RD4 controller and its performance during and after power failures. This is the basis for set-point entry and all special controller functions.
- Menu structure with access levels
- Defined behavior during and after power failures
- Controller display with text field for messages
Describes initial start up and entry of the temperature set-point and fan speed set-point. These are the basic operating settings of the chamber.
- Temperature set-point entry
- Fan speed set-point entry
Covers special controller functions and automatic setpoint switchover controlled by date or by regular cyclic control, including entry of the alternative temperature set-point and activation/deactivation of the functions.
- Automatic setpoint switchover by date
- Regular cyclic setpoint switchover
- Entry of alternative temperature set-point
- Activating/deactivating special controller functions
Describes preparations, starting, the process itself, completion and premature termination of hot-air disinfection.
- Preparations before hot-air disinfection
- Starting and running the hot-air disinfection process
- Completing or prematurely terminating hot-air disinfection
Covers the password request and assignment/modification of the User password and Admin password.
- Password request on the controller
- Assign and modify User password
- Assign and modify Admin password
Describes the over temperature protective device (class 1) and the safety controller class 2 / class 3.3, including selection between temperature limiter and temperature protection, safety controller mode, limit and offset values, class 3.2 undertemperature protection, alarm state measures and function check.
- Over temperature protective device class 1
- Safety controller class 2 (temperature limiter) or class 3.3 (temperature protection)
- Setting safety controller mode, limit value and offset value
- Safety controller class 3.2 undertemperature protection
- Message and measures in the state of alarm
- Function check of the safety devices
Covers menu language, temperature unit, date and time, language selection at restart, chamber address and display brightness.
- Selecting menu language and temperature unit
- Setting current date and time
- Language selection at restart
- Setting the chamber address
- Display brightness
Covers the delay time for the tolerance range alarm and the temperature tolerance range settings.
- Setting delay time for tolerance range alarm
- Setting temperature tolerance range
Describes information messages, alarm messages and activation/deactivation of the audible alarm (alarm buzzer).
- Information messages
- Alarm messages
- Activating/deactivating the audible alarm buzzer
Covers the start time program, start week program, stopping the program and performance after power failures.
- Start time program
- Start week program
- Stopping the program
- Performance after power failures
Covers showing and changing network settings: MAC address, IP address, subnet mask, standard gateway and DNS server address, plus selection of automatic or manual IP/DNS assignment.
- Showing MAC address, IP address, subnet mask, standard gateway and DNS server address
- Selecting automatic or manual IP address assignment
- Selecting automatic or manual DNS server address assignment
- Assigning IP address, subnet mask, standard gateway and DNS server address
Describes recorded data, storage capacity, setting the storage rate for the DL1 recorder data and deleting the data recorder.
- Recorded data and storage capacity
- Setting storage rate for DL1 recorder data
- Deleting the data recorder
Covers connecting a USB stick, import/export functions, ongoing data transfer, errors during data transmission and removing the USB stick.
- Connecting the USB stick
- Import/export functions
- Ongoing data transfer
- Error during data transmission
- Removing the USB stick
Describes activating and deactivating the self-test function of the chamber.
- Activating the self-test function
- Deactivating the self-test mode
Covers defrosting during refrigerating operation and the anti-condensation protection function.
- Defrosting at refrigerating operation
- Anti-condensation protection
Describes APT-COM 4 Multi Management Software, analog outputs for temperature, zero-voltage alarm contacts for collective alarm output, door lock, internal LED light and object temperature control with flexible Pt 100 temperature sensor including sensitivity and maximum deviation settings.
- APT-COM 4 Multi Management Software (accessory)
- Analog outputs for temperature (option)
- Zero-voltage alarm contacts for collective alarm output (option)
- Door lock (option)
- Internal LED light (option)
- Object temperature control with flexible Pt 100 sensor (option)
- Setting sensitivity and maximum deviation of object temperature control
Covers cleaning, decontamination/chemical disinfection and hot-air disinfection procedures.
- Cleaning
- Decontamination / chemical disinfection
- Hot-air disinfection
Covers personnel qualification, maintenance intervals and service, the service reminder, simple troubleshooting, and returning the chamber to BINDER GmbH. Annual maintenance and annual electrical repeat inspection are required.
- Maintenance at least once a year
- Annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702
- Service reminder shown on controller after one year of operation
- Simple troubleshooting by operating personnel without disassembly
- Detailed troubleshooting and repair by BINDER Service or qualified partners
- Chamber must be tested before resuming operation after maintenance or repair
Covers factory calibration and adjustment, over current protection, definition of usable volume, KB/KB-UL technical data, equipment and options, accessories and spare parts, and dimensions.
- Factory calibration and adjustment
- Over current protection
- Definition of usable volume
- KB / KB-UL technical data
- Equipment and options (extract)
- Accessories and spare parts (extract)
- Dimensions
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal temperature of the chambers is 100 °C / 212 °F. p.10 ↗ | Maximum rated operating temperature for selection and application planning. | KB / KB-UL series |
| IP protection is IP 20 according to EN 60529. p.10 ↗ | Chamber must be installed splash-proof and must not become wet during operation, cleaning or maintenance. | KB / KB-UL series |
| Temperature safety device according to DIN 12880:2007, class 2/3.3 (adjustable). p.10 ↗ | Safety controller can be configured as class 2 temperature limiter or class 3.3 temperature protection. | KB / KB-UL series |
| KB 260 (example) nominal power 1.30 kW, nominal current 5.4 A at 220-240 V, 50/60 Hz, 1 N~. p.10 ↗ | Electrical supply and fusing must be sized for the nominal current of the specific model. | KB 260 (9020-0473) |
| KB 260-UL (example) nominal power 1.30 kW, nominal current 10.8 A at 120 V, 50/60 Hz, 1 N~. p.10 ↗ | UL variant requires a 120 V supply sized for 10.8 A. | KB 260-UL (9020-0501) |
| Maximum permissible operating overpressure in the refrigerating system is 11 bar. p.10 ↗ | Refrigerant circuit must not be exposed to higher pressure; relevant for service work. | KB / KB-UL series |
| Refrigerant R600A with filling weight 0.095 kg (example KB 260); contains hydrocarbon gases. p.10 ↗ | Flammable refrigerant; service work on the refrigeration system only by personnel trained per EN 13313:2010. | KB / KB-UL series |
| Intended temperature application range 4 °C / 39.2 °F to 37 °C / 98.6 °F for cultivation of microorganisms. p.1 ↗ | Typical applications: long-term storage at 4 °C, refrigerated incubation 20-25 °C, incubation at 37 °C, alternating temperatures. | KB / KB-UL series |
| Maintenance work at the refrigeration system only by qualified personnel trained according to EN 13313:2010 (e.g. refrigeration technician per Regulation (EC) no. 303/2008). p.25 ↗ | Service personnel qualification must be verified before any refrigeration work. | KB / KB-UL series |
| Regular maintenance work at least once a year; all work on the refrigeration system must be documented. p.25 ↗ | Plan annual maintenance and keep documentation. | KB / KB-UL series |
| Annual repeat electrical inspection and test before initial startup and before resuming operation after maintenance or repair, recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.25 ↗ | Electrical safety testing is mandatory before first use and after service. | KB / KB-UL series |
| With increased dust in ambient air, clean the condenser of the refrigeration machine by suction several times a year; condenser is behind the tool-free removable machine room cover. p.25 ↗ | Dusty environments require more frequent condenser cleaning to maintain performance. | KB / KB-UL series |
| Replace the door gasket only when cold. p.25 ↗ | Replacing a warm gasket may damage it. | KB / KB-UL series |
| Service reminder: keep OK button pressed for 5 seconds to display remaining days until maintenance is due; after one year of operation the message 'Service due!' appears and reappears after one week if not confirmed. p.25 ↗ | Controller provides a maintenance countdown and recurring due message. | KB / KB-UL series |
| APT-COM 4 Multi Management Software permits networking up to 100 chambers and connecting them to a PC. p.2 ↗ | Network planning must account for up to 100 chambers per APT-COM 4 installation. | KB / KB-UL series |
| Spacer for wall distance required from size 260 on. p.4 ↗ | Sizes 260 and larger need the wall spacer installed to maintain ventilation clearances. | KB / KB-UL 260, 470, 720, 1060 |
| Flexible tilt protection kit is mounted on KB/KB-UL 65, 130 and 470. p.4 ↗ | These sizes require the tilt protection kit during installation. | KB/KB-UL 65, 130, 470 |
| Chamber must not be installed in unventilated recesses; sufficient ventilation for heat dispersal and prescribed minimum distances must be observed. p.1 ↗ | Prevents overheating and chamber damage. | KB / KB-UL series |
| Do not remove the rear panel of the chamber; disconnect power before maintenance by turning off the On/Off switch (H) and pulling the power plug. p.25 ↗ | Live maintenance work is prohibited; rear panel removal is not permitted. | KB / KB-UL series |
| Loading material must not contain corrosive ingredients such as acids and halides that may damage stainless steel, aluminum and copper components. p.1 ↗ | Corrosive damage from such ingredients is excluded from BINDER liability. | KB / KB-UL series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Service due! p.78 ↗ | Recommended maintenance interval (one year of operation) has elapsed. | Press the OK button to confirm the message. The message reappears after one week of operation. | chap. 25.3 |