BINDER KB 65
BINDER KB 65 — Incubators with Temperature 0…70 °C.
The BINDER KB 65 is a 63 L compressor-cooled incubator with a 0…70 °C working range, 0.2 K uniformity and fluctuation, and a 160 W max heat compensation. As the smallest standard variant in the KB series, it is distinguished by a 30 kg per-rack load and 90 kg total permitted load. It is intended for refrigerated incubation, long-term storage at 4 °C and alternating-temperature cultivation of microorganisms.
Direct drop-in replacements
BINDER KB Series — Model Comparison & Selection
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 63 L |
| Load per rack | 30 kg |
| Max. heat compensation | 160 W |
| Nominal power | 0.9 kW |
| Permitted load | 90 kg (KB065-230V / KB065UL-120V) |
| Temperature fluctuation | 0.2 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 100 mm (KB065-230V / KB065UL-120V) |
| Side wall clearance | 100 mm (KB065-230V / KB065UL-120V) |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 3 min |
| Depth | 710 mm (KB065-230V / KB065UL-120V) |
| Height | 905 mm (KB065-230V / KB065UL-120V) |
| Interior width | 406 mm (KB065-230V / KB065UL-120V) |
| Width | 650 mm (KB065-230V / KB065UL-120V) |
| Net weight of the unit (empty) mfr | 95 kg |
| Power frequency | 50 Hz |
| Rated voltage | 220...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The KB 65 is a forced-convection cooling incubator using compressor technology (R600A refrigerant) and the RD4 chamber controller. Its 63 L interior (406 mm wide, 710 mm deep, 905 mm high) is the smallest standard size in the KB series; siblings KB 130, KB 260, KB 470 and KB 1060 share the same 0…70 °C range but offer larger volumes. The KB 65 is the only size that combines the smallest footprint with a 90 kg total load rating and a 30 kg per-rack limit.
- Unlike sizes 260 and above, the KB 65 does not require the wall spacer; instead, it uses the flexible tilt protection kit (shared with KB 130 and KB 470). Rear and side wall clearances of 100 mm must be maintained.
Behavior under load and in different media
- Temperature fluctuation and uniformity are both specified at 0.2 K, supporting stable long-term cultivation. Recovery after a 30 s door opening at 37 °C is 3 minutes. Maximum heat compensation is 160 W, which defines the permissible heat load from samples and equipment inside the chamber.
- The intended application range is 4 °C to 37 °C for microorganism cultivation, covering storage at 4 °C, refrigerated incubation at 20–25 °C, incubation at 37 °C and alternating temperature profiles. The chamber is not intended for operation outside 0…70 °C.
Installation, control and operation
- Electrical supply is 220…240 V, 1~, 50 Hz, nominal power 0.9 kW, with a 16.0 A unit fuse. The RD4 controller provides set-point entry for temperature and fan speed, automatic setpoint switchover by date or cyclic control, program start/stop, a data recorder (DL1) and Ethernet/USB interfaces. APT-COM 4 software can network up to 100 chambers.
- Temperature safety is provided by an over-temperature protective device (class 1) plus an adjustable safety controller per DIN 12880:2007 class 2/3.3, configurable as temperature limiter or temperature protection. Hot-air disinfection and defrosting functions are available. The chamber is IP 20 and must be installed splash-proof.
Selection & Integration Q&A
Choose the KB 65 when 63 L interior volume and a 90 kg total load (30 kg per rack) cover your workload; it is the smallest standard KB size and needs no wall spacer. Step up to KB 130 or larger when you need more chamber volume or higher total load, accepting the wall-spacer requirement from size 260.
The standard KB065-230V needs 220…240 V, 1~, 50 Hz, 0.9 kW, with a 16.0 A unit fuse and CEE 7/7 plug. For networking, the RD4 controller provides Ethernet and USB; APT-COM 4 can manage up to 100 chambers on one PC installation.
Maintain 100 mm clearance at the rear and sides, mount the flexible tilt protection kit, and do not install in an unventilated recess. The chamber is IP 20 and must be kept splash-proof. Ambient dust requires more frequent condenser cleaning to maintain refrigeration performance.
Annual maintenance and annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required; refrigeration work must be by EN 13313:2010-trained personnel. Optional accessories not included: APT-COM 4 software, analog temperature outputs, zero-voltage alarm contacts, door lock, internal LED light, Pt 100 object sensor, racks with U-rails or telescopic rails.