BINDER KB 130
BINDER ·Made in Germany · In Active Production

BINDER KB 130

BINDER KB 130 — Incubators with Temperature 0…70 °C.

The BINDER KB 130 is a 127 L compressor-cooled incubator with a 0–70 °C working range and RD4 controller, positioned between the KB 65 and KB 260 in the KB series. Its 270 W heat compensation and 0.1 K fluctuation make it suited for long-term storage at 4 °C and incubation at 37 °C. The 130 size requires the flexible tilt protection kit during installation, as do the KB 65 and KB 470.

Temperature
0…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB 130 Current 0–7071100.512765630
KB 1060 0–70117.5195.11070101140
KB 260 0–7090139.626265640
KB 470 0–7090195.147265640
KB 65 0–707190.56340630
KB 720 0–7090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 127 L
Load per rack 30 kg
Max. heat compensation 270 W
Nominal power 1.2 kW
Permitted load 150 kg (KB130-230V / KB130UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm (KB130-230V / KB130UL-120V)
Side wall clearance 100 mm (KB130-230V / KB130UL-120V)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Depth 710 mm (KB130-230V / KB130UL-120V)
Height 1005 mm (KB130-230V / KB130UL-120V)
Interior width 656 mm (KB130-230V / KB130UL-120V)
Width 900 mm (KB130-230V / KB130UL-120V)
Net weight of the unit (empty) mfr 124 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 220...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

The KB 130 (article 9020-0472) is a floor-standing cooling incubator with 127 L interior volume, 656 mm interior width, and 124 kg empty weight. It uses compressor technology with R600A refrigerant and the RD4 chamber controller. Within the KB series, the 130 size sits between the KB 65 and KB 260; unlike sizes 260 and larger, it does not require a wall spacer, but it does require the flexible tilt protection kit (shared with KB 65 and KB 470). The chamber is IP 20 and must be installed splash-proof.

Behavior under load and in different modes

The intended temperature application range is 4 °C to 37 °C for cultivation of microorganisms, though the full operating range is 0–70 °C. Temperature fluctuation is 0.1 K and uniformity 0.2 K. Maximum heat compensation is 270 W, and recovery time after a 30 s door opening at 37 °C is 3 minutes. The chamber supports hot-air disinfection, automatic setpoint switchover by date or cyclic control, and defrosting during refrigerating operation with anti-condensation protection. The temperature safety device is adjustable per DIN 12880:2007 class 2/3.3, configurable as a temperature limiter or temperature protection.

Installation, connection and control

Electrical supply is 220–240 V, 1~, 50 Hz, nominal power 1.2 kW, unit fuse 16.0 A. The RD4 controller provides Ethernet and USB interfaces; APT-COM 4 software can network up to 100 chambers. The rear connection panel includes interfaces for computer communication. Minimum wall clearances are 100 mm at the back and 100 mm at the side. The chamber must not be installed in unventilated recesses. The condenser is behind a tool-free removable machine room cover for cleaning.

Integration & Operating Notes

Mounting
Tilt protection kit is mandatory for KB 130 installation

The flexible tilt protection kit must be mounted on KB/KB-UL 65, 130 and 470 during installation. Omitting it leaves the chamber without the prescribed tip-over safeguard, which is a safety-relevant deviation from the manual.

Instruction Manual, chap. 4.2, p.25
Mounting
Maintain 100 mm clearance at rear and sides

The chamber requires 100 mm wall clearance at the back and sidewise to ensure sufficient ventilation for heat dispersal. Installing it in an unventilated recess or closer than these distances risks overheating and chamber damage.

Instruction Manual, chap. 1.8, p.12; Technical Data
Electrical
Supply must be sized for 1.2 kW and 16 A fusing

The KB 130 draws 1.2 kW nominal at 220–240 V, 1~, 50 Hz, with a 16.0 A unit fuse. The branch circuit and protective device must be coordinated with this nominal current; the UL variant (KB130UL-120V) uses a different supply and plug and is not interchangeable.

Factory Specification Sheet; Instruction Manual, Type plate p.10
Process Limits
Loading material must be free of corrosive halides and acids

Materials containing acids or halides that may attack stainless steel, aluminum or copper components must not be loaded. Corrosive damage from such ingredients is excluded from BINDER liability, so process compatibility must be checked before use.

Instruction Manual, chap. 1.9, p.14
Maintenance
Annual maintenance and electrical repeat inspection required

Regular maintenance is required at least once a year, and an annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended before initial startup and after any service. The controller shows 'Service due!' after one year and repeats it weekly until confirmed.

Instruction Manual, chap. 25.1–25.3, p.77–78
Cooling & Media
Clean condenser several times a year in dusty environments

With increased dust in ambient air, the condenser behind the tool-free removable machine room cover must be cleaned by suction several times a year to maintain refrigerating performance. Refrigeration system work is restricted to personnel trained per EN 13313:2010 due to flammable R600A refrigerant.

Instruction Manual, chap. 25.2, p.77–78

Selection & Integration Q&A

When should I choose the KB 130 rather than the KB 65 or KB 260?

Choose the KB 130 for 127 L interior volume when the KB 65 is too small and the KB 260 would require the wall spacer. The 130 shares the tilt protection kit requirement with the 65 and 470, but offers more usable volume than the 65 without the larger footprint of the 260.

What supply and fusing does the KB 130 require, and can I use the UL variant on the same circuit?

The KB 130 needs 220–240 V, 1~, 50 Hz, 1.2 kW, with a 16.0 A unit fuse. The KB130UL-120V is a separate 120 V variant with NEMA 5-20 plug and different current; it cannot share the same branch circuit or plug configuration.

What installation clearances and safeguards apply to the KB 130?

Maintain 100 mm clearance at the rear and sides, do not install in unventilated recesses, and mount the flexible tilt protection kit required for sizes 65, 130 and 470. The chamber is IP 20 and must be kept splash-proof.

What maintenance and accessories are needed beyond the base delivery?

Plan annual maintenance and electrical repeat inspection; the controller shows a service reminder after one year. Optional accessories include APT-COM 4 software, analog temperature outputs, zero-voltage alarm contacts, door lock, internal LED light, Pt 100 object sensor, and racks with U-rails or telescopic rails.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB 130 en
Datasheet ·475,4 КБ ·7 pp.
View PDF
PDF
KB (E7) 09 2026 en
Manual ·12,4 МБ ·107 pp.
View PDF