M (E3.1) 06 2026
Technical Reference Summary
Covered Models & Document Scope
This operating manual (M (E3.1), issue 06/2026, Art. No. 7001-0441) covers the BINDER M-series drying and heating ovens with forced convection and advanced program functions, equipped with the MB2 microprocessor chamber controller. Covered models and order codes: M 56 (E3.1) / M056-230V / Art. No. 9010-0371, 9110-0371; M 115 (E3.1) / M115-230V / Art. No. 9010-0372, 9110-0372; M 260 (E3.1) / M260-400V / Art. No. 9010-0373, 9110-0373; M 720 (E3.1) / M720-400V / Art. No. 9010-0374, 9110-0374. Options and accessories addressed include APT-COM 4 Multi Management Software, flexible Pt100 object temperature sensor (display or control), analog output for temperature, HEPA fresh air filter, mostly gas-tight version, and inert gas connection with mostly gas-tight version.
Section-by-Section Technical Breakdown
Describes the chamber layout, the rear panel connections and the triangular instrument panel. The rear provides a SUB-D socket 'ALARM / CONTACT' for a zero-voltage relay alarm contact and a SUB-D socket 'ANALOG OUTPUT' (option). Sizes 260 and 720 have a main power switch on the chamber rear.
- SUB-D socket 'ALARM / CONTACT' (6) provides zero-voltage relay alarm contact
- SUB-D socket 'ANALOG OUTPUT' (7) is an option for analog temperature output
- Main power switch (11) is located on the chamber rear for sizes 260 and 720
- Triangular instrument panel and zero-voltage switching contact are described in chap. 2.3 and 2.5
Covers installation location and ambient conditions, electrical connection, optional connection to an exhaust/ventilation system, and insertion of racks. The chamber must not be installed in unventilated recesses and prescribed minimum distances must be observed. Electrical connection data are given per model.
- Observe prescribed minimum distances when installing the chamber (chap. 3.4)
- Ensure sufficient ventilation for dispersal of heat; do not install in unventilated recesses
- Do not install or operate chambers in hazardous locations or damp areas/puddles
- Optional connection to an exhaust/ventilation system is described in chap. 4.2
- Racks are inserted as described in chap. 4.3
Introduces the MB2 microprocessor controller: normal display, program display and chart-recorder display views, controller icons, operating modes, menu structure (main menu, Settings submenu, Service submenu), entry principle, and behavior during power failures and when opening the door.
- Display views: Normal display, program display, chart-recorder display (chap. 5.2)
- Controller menu structure includes Main menu, 'Settings' submenu and 'Service' submenu (chap. 5.5)
- Performance during and after power failures described in chap. 5.7
- Performance when opening the door described in chap. 5.8
Chapter 6 describes the failsafe function of the chamber controller. Details are contained in the referenced chapter.
- Failsafe function is covered in chapter 6
Covers turning on the chamber, controller settings upon start up, and loading the chamber. These steps precede set-point entry and program operation.
- Turning on the chamber (chap. 7.1)
- Controller settings upon start up (chap. 7.2)
- Loading the chamber (chap. 7.3)
Describes set-point entry for temperature and fan speed through the 'Setpoints' menu, adjustment of the air flap position, direct setpoint entry via the Normal display, and special controller functions.
- Set-point entry for temperature and fan speed through the 'Setpoints' menu (chap. 8.1)
- Adjusting the air flap position through the 'Setpoints' menu (chap. 8.2)
- Direct setpoint entry via Normal display (chap. 8.3)
- Special controller functions (chap. 8.4)
Describes starting a timer program, performance during program delay time, cancelling a running timer program, and performance after the end of the program.
- Starting a timer program (chap. 9.1)
- Performance during program delay time (chap. 9.1.1)
- Cancelling a running timer program (chap. 9.2)
- Performance after the end of the program (chap. 9.3)
Covers starting, stopping, pausing and cancelling time programs, performance after program end, creating new time programs, program editor and section editor management, value entry for program sections (duration, set-point ramp and step, special functions, temperature/fan/air flap, tolerance ranges, section repetition), and saving the time program.
- Program editor: program management and deleting a time program (chap. 10.5)
- Section editor: add, copy/insert/replace, delete program sections (chap. 10.6)
- Value entry for a program section: section duration, set-point ramp and set-point step (chap. 10.7)
- Tolerance ranges and repeating one or several sections within a time program (chap. 10.7.5, 10.7.6)
- Saving the time program (chap. 10.7.7)
Covers starting and cancelling week programs, creating new week programs, program editor and section editor management, and value entry for program sections including set-point ramp/step modes, weekday, start time, temperature/fan/air flap settings and special controller functions.
- Program editor: program management and deleting a week program (chap. 11.4)
- Section editor: add, copy/insert/replace, delete program sections (chap. 11.5)
- Value entry: set-point ramp and set-point step modes, weekday, start time (chap. 11.6)
- Setpoint entry for temperature and fan speed, setting the air flap position (chap. 11.6.4)
Provides an overview of notification and alarm messages, describes the state of alarm, resetting an alarm and the list of active alarms, tolerance range settings, and activating/deactivating the audible alarm buzzer.
- Notification and alarm messages overview (chap. 12.1)
- State of alarm (chap. 12.2)
- Resetting an alarm, list of active alarms (chap. 12.3)
- Tolerance range settings (chap. 12.4)
- Activating/deactivating the audible alarm (alarm buzzer) (chap. 12.5)
Describes the overtemperature protective device class 1 and the overtemperature safety controller class 2 / 3.1, including selection between class 2 (temperature limiter) and class 3.1, safety controller mode, setting the safety controller, messages and measures in the state of alarm, function check, and the zero-voltage relay alarm contact.
- Overtemperature protective device class 1: message and measures in the state of alarm (chap. 13.1.1)
- Selecting between safety controller class 2 (temperature limiter) or class 3.1 (chap. 13.2.1)
- Safety controller mode and setting the safety controller (chap. 13.2.2, 13.2.3)
- Message and measures in the state of alarm for class 2/3.1 (chap. 13.2.4)
- Function check (chap. 13.2.5)
- Zero-voltage relay alarm contact (chap. 13.3)
Covers authorization levels and password protection, log in/log out, user change, password assignment and change, deleting a password for an individual authorization level, new password assignment for 'service' or 'admin' authorization level when the password function was deactivated, and activation code.
- Authorization levels and password protection (chap. 14.1)
- Log in, log out, user change (chap. 14.2-14.4)
- Password assignment and password change (chap. 14.5)
- New password assignment for 'service' or 'admin' authorization level when the password function was deactivated (chap. 14.5.3)
- Activation code (chap. 14.6)
Covers selecting the controller menu language, setting date and time, selecting the temperature unit, display configuration (adapting display parameters, touchscreen calibration), network and communication (Ethernet configuration, MAC address display, E-Mail), and USB data transfer.
- Selecting the controller's menu language (chap. 15.1)
- Setting date and time (chap. 15.2)
- Selecting the temperature unit (chap. 15.3)
- Display configuration and touchscreen calibration (chap. 15.4)
- Ethernet configuration and display of the MAC address (chap. 15.5.1)
- E-Mail settings (chap. 15.5.2)
- USB menu: data transfer via USB interface (chap. 15.6)
Provides the service contact page, current operating parameters, event list, and technical chamber information.
- Service contact page (chap. 16.1)
- Current operating parameters (chap. 16.2)
- Event list (chap. 16.3)
- Technical chamber information (chap. 16.4)
Describes the chart recorder views, showing/hiding the legend, switching between legend pages, showing/hiding specific indications, history display, and setting the parameters.
- Views and legend handling (chap. 17.1)
- History display (chap. 17.1.4)
- Setting the parameters (chap. 17.2)
Describes APT-COM 4 Multi Management Software (accessory), object temperature display and object temperature control with flexible Pt100 temperature sensor (options), analog output for temperature (option), HEPA fresh air filter (option), mostly gas-tight version (option), and inert gas connection with mostly gas-tight version (option).
- APT-COM 4 Multi Management Software permits networking up to 100 chambers and connecting them to a PC (chap. 18.1)
- Object temperature display with flexible Pt100 temperature sensor (option) (chap. 18.2)
- Object temperature control with flexible Pt100 temperature sensor (option) (chap. 18.3)
- Analog output for temperature (option) (chap. 18.4)
- HEPA fresh air filter (option) (chap. 18.5)
- Mostly gas-tight version (option) and inert gas connection (chap. 18.6, 18.7)
Covers cleaning and decontamination / chemical disinfection of the chamber.
- Cleaning (chap. 19.1)
- Decontamination / chemical disinfection (chap. 19.2)
Covers personnel qualification for maintenance, maintenance intervals and service, simple troubleshooting by operating personnel, and sending the chamber back to BINDER GmbH. Annual repeat electrical inspection is required per EN 50678/VDE 0701 and EN 50699/VDE 0702. Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER.
- Maintenance work at least once a year; warranty void if performed by non-authorized personnel
- Disconnect chamber before maintenance; do not remove rear panel; chamber must not become wet
- Replace door gasket only when cold
- Annual repeat inspection and test before initial startup and before resuming operation after maintenance/repair; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702
- Simple troubleshooting table covers general, temperature and heating faults (chap. 20.3)
- Detailed troubleshooting and repair by BINDER Service or qualified service partners per Service Manual
Covers factory calibration and adjustment, definition of usable volume, over current protection, M technical data, equipment and options, accessories and spare parts, and dimension drawings for sizes 56, 115, 260 and 720.
- Factory calibration and adjustment (chap. 22.1)
- Definition of usable volume (chap. 22.2)
- Over current protection (chap. 22.3)
- M technical data (chap. 22.4)
- Equipment and options extract (chap. 22.5)
- Accessories and spare parts extract (chap. 22.6)
- Dimensions size 56, 115, 260, 720 (chap. 22.7-22.10)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| M 115 (E3.1) type plate example: nominal temperature 300 °C (572 °F), nominal power 2.40 kW, nominal current 10.5 A, nominal voltage 230 V / 50 Hz ±10% and 230 V / 60 Hz, current type 1 N ~, IP protection 20. p.1 ↗ | Electrical supply and protection must be sized for 230 V single-phase, 10.5 A, 2.40 kW; verify supply voltage within ±10% at indicated power frequency. | M 115 (E3.1), Art. No. 9010-0372 / 9110-0372 |
| Temperature safety device according to DIN 12880, class 2/3.1 (selectable). p.1 ↗ | The chamber provides a selectable overtemperature safety controller class 2 (temperature limiter) or class 3.1; selection and setting are described in chap. 13.2. | M series (all sizes) |
| IP protection 20 according to EN 60529. p.1 ↗ | The chamber is protected against solid objects >12.5 mm but not against water; do not install in damp areas or puddles and set up splash-proof. | M series (all sizes) |
| The chamber must NOT become wet during operation, cleaning, or maintenance; do NOT install in damp areas or in puddles; set up in a splash-proof manner. p.1 ↗ | Prevents deadly electric shock from water entering the chamber. | M series (all sizes) |
| Do NOT install the chamber in unventilated recesses; ensure sufficient ventilation for dispersal of heat and observe prescribed minimum distances (chap. 3.4). p.1 ↗ | Prevents overheating and chamber damage due to lack of ventilation. | M series (all sizes) |
| Do NOT operate the chamber in potentially explosive areas; keep combustible dust or air-solvent mixtures away from the chamber. The chambers do not dispose of any measures of explosion protection. p.1 ↗ | The chamber is not suitable for explosive atmospheres; site selection must exclude combustible dusts and explosive mixtures. | M series (all sizes) |
| Do NOT introduce any substance into the chamber which is combustible or explosive at working temperature; the temperature inside the chamber must lie below the flash point or below the sublimation point of the charging material. p.1 ↗ | Charging material selection must consider flash point/sublimation point relative to the set-point temperature. | M series (all sizes) |
| During and shortly after operation, the inner surfaces almost equal the set-point; inner chamber, exhaust duct, door gaskets and access ports (option) become hot. p.1 ↗ | Do not touch inner surfaces, exhaust duct, access ports, door gaskets or charging material during operation; do not place power cable over the door gap when chamber is hot. | M series (all sizes) |
| Maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER; warranty becomes void if maintenance work is conducted by non-authorized personnel. p.20 ↗ | Plan annual maintenance and use authorized personnel to preserve warranty. | M series (all sizes) |
| Disconnect the chamber before conducting maintenance work: turn off with the on/off switch (4) or main power switch (11) and pull the power plug. Do NOT remove the rear panel of the chamber. p.20 ↗ | Electrical hazard during live maintenance work; rear panel removal is prohibited. | M series (all sizes) |
| Replace the door gasket only when cold; otherwise the door gasket may become damaged. p.20 ↗ | Allow chamber to cool before door gasket replacement. | M series (all sizes) |
| Annual repeat inspection as well as a test prior to initial startup and prior to resuming operation after maintenance or repair are required; recommended testing under EN 50678/VDE 0701 and EN 50699/VDE 0702. p.20 ↗ | Electrical safety testing must be scheduled at commissioning, after maintenance/repair, and annually. | M series (all sizes) |
| After maintenance, the chamber must be tested prior to resuming operation. p.20 ↗ | Do not return chamber to service after maintenance without testing. | M series (all sizes) |
| If the overtemperature protective device class 1 has turned off the chamber: disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, contact BINDER service. p.20 ↗ | Provides a defined reset procedure and escalation path for class 1 protective device trip. | M series (all sizes) |
| APT-COM 4 Multi Management Software permits networking up to 100 chambers and connecting them to a PC. p.2 ↗ | Network planning must account for up to 100 chambers per APT-COM 4 installation. | M series with APT-COM 4 accessory |
| SUB-D socket 'ALARM / CONTACT' (6) provides a zero-voltage relay alarm contact; SUB-D socket 'ANALOG OUTPUT' (7) is an option for analog temperature output. p.2 ↗ | External alarm monitoring and analog temperature recording can be wired via the rear SUB-D sockets. | M series (all sizes) |
| Main power switch (11) is located on the chamber rear for sizes 260 and 720. p.2 ↗ | For M 260 and M 720, the main power switch is at the rear; ensure accessible location during installation. | M 260 (E3.1), M 720 (E3.1) |
| The chamber must only be installed, tested, and started up by personnel qualified for assembly, startup, and operation; operation only by laboratory personnel trained for this purpose. p.1 ↗ | Personnel qualification requirements apply to installation, commissioning and operation. | M series (all sizes) |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber without function. p.110 ↗ | No power supply. | Check connection to power supply. Check whether the chamber is turned on with the On/Off switch (4) or main power switch (11). | chap. 20.3 |
| Chamber without function. p.110 ↗ | Wrong voltage. | Check power supply for correct voltage (chap. 4.1). | chap. 20.3 |
| Chamber without function. p.110 ↗ | Chamber fuse has responded. | Check chamber fuse and replace it if appropriate (chap. 22.3). If it responds again, contact BINDER service. | chap. 20.3 |
| Chamber without function. p.110 ↗ | Defective controller. | Contact BINDER service. | chap. 20.3 |
| Chamber without function. p.110 ↗ | Overtemperature protective device class 1 has turned off the chamber. | Disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, contact BINDER service. | chap. 20.3 |
| Set-point temperature is not reached after specified time. p.110 ↗ | Chamber door not properly closed. | Completely close chamber door. | chap. 20.3 |
| Set-point temperature is not reached after specified time. p.110 ↗ | Defective door gasket. | Replace door gasket. | chap. 20.3 |
| Set-point temperature is not reached after specified time. p.110 ↗ | Controller not adjusted. | Calibrate and adjust controller. | chap. 20.3 |
| Set-point temperature is not reached after specified time. p.110 ↗ | Wrong voltage. | Check the power supply for correct voltage (chap. 4.1). | chap. 20.3 |
| Chamber heating permanently, set-point not held. p.110 ↗ | Defective controller. | Contact BINDER service. | chap. 20.3 |
| Chamber heating permanently, set-point not held. p.110 ↗ | Defective Pt 100 sensor. | Contact BINDER service. | chap. 20.3 |
| Chamber heating permanently, set-point not held. p.110 ↗ | Defective semiconductor relay. | Contact BINDER service. | chap. 20.3 |
| Chamber heating permanently, set-point not held. p.110 ↗ | Controller not adjusted. | Calibrate and adjust controller. | chap. 20.3 |
| Chamber doesn't heat up. Heating icon is displayed. p.110 ↗ | Defective heating element. | Contact BINDER service. | chap. 20.3 |
| Chamber doesn't heat up. Heating icon is displayed. p.110 ↗ | Defective semiconductor relay. | Contact BINDER service. | chap. 20.3 |