Cary 3500 UV-Vis Spectrophotometer

Contents

  1. Overview pp. 1-2
    Product introduction and modular design benefits
  2. Overview pp. 3-8
    Regulatory compliance, data integrity, and software features
  3. Spec pp. 9-10
    System overview and high-concentration measurement performance
  4. Config pp. 11-12
    Available accessories and compact model options
  5. Spec pp. 13-15
    Temperature ramping and multi-cell measurement capabilities
  6. Services p. 16
    Agilent CrossLab support and service offerings
  7. p. 13 table, 6 rows

The measured Tm values for the herring sperm DNA sample at each temperature ramp rate

Ramp rate (°C/min)Replicate 1 Tm (°C)Replicate 2 Tm (°C)Replicate 3 Tm (°C)Average Tm (°C)Standard Deviation
187.187.187.187.10.0
587.086.686.586.70.2
1086.787.187.086.90.2
2087.187.187.187.10.0
3087.187.086.686.90.2
4086.687.087.086.90.2

Engineering Facts

14 facts · every claim linked to a page
Selection

The series shares a common UV-Vis engine with interchangeable modules, so the module choice defines whether the lab gets single-sample, multicell, or flexible solid/liquid capability.

Determines which module configuration must be specified at purchase to match cuvette, solid-sample, or high-throughput needs.

Source
“The Agilent Cary 3500 UV-Vis spectrophotometer series share a common UV-Vis engine, which produces monochromatic light that is measured by the various UV-Vis sample measurement modules.”
Selection

The Multicell module supports up to seven samples plus a reference in eight cell positions, which sets the maximum batch throughput per measurement.

Defines the sample capacity needed when specifying the instrument for multi-standard calibration or batch workflows.

Source
“The Cary 3500 Multicell UV-Vis is designed for measuring up to seven samples and a reference (or other combinations in the eight cell positions).”
Operation

The xenon flash lamp only fires during acquisition, which protects photosensitive samples and removes warm-up time before use.

Affects sample handling for photolabile analytes and allows immediate measurement after power-on without a warm-up wait.

Source
“The lamp only illuminates the sample when data is acquired, protecting sensitive samples from photodegradation, and reducing power consumption.”
Maintenance

The xenon flash lamp carries a 10-year replacement warranty, which substantially lowers long-term lamp replacement cost and downtime.

Reduces total cost of ownership and planned maintenance compared with conventional lamps.

Source
“The xenon flash lamp comes with 10-year replacement warranty (for Cary 3500 instruments purchased from Agilent or participating partners)”
Performance

The engine acquires 250 data points per second with wavelength drive up to 2500 nm/s, so fast kinetics are not lost to acquisition rate.

Confirms the instrument can capture fast reactions without undersampling when specifying for kinetics work.

Source
“With the xenon lamp generating 250 data points per second and the wavelength drive moving at up to 2500 nm per second, you’ll never miss vital data”
Compliance

OpenLab provides access control, secure storage, e-signatures and audit trail to support FDA 21 CFR Part 11 and related regulated environments.

Determines whether the data system configuration meets the lab's regulatory data-integrity requirements.

Source
“These controls include access control and secure storage in a local or central database, electronic signature workflows, and an advanced audit trail review.”
Compliance

Automated operational qualification tests align with USP <857>, Ph. Eur. 2.2.25 and JP 2.24, so passing them demonstrates pharmacopeial performance.

Supports qualification planning and demonstrates compliance for pharmaceutical laboratories.

Source
“These tests align with the global pharmacopeia requirements of U.S. Pharmacopeia (USP) general chapter <857>, European Pharmacopeia (Ph. Eur.) chapter 2.2.25, and the Japanese Pharmacopeia (JP) chapter 2.24.”
Installation

The Sipper accessory uses a three-channel pump to fill and rinse up to three flow cells simultaneously, reducing manual cuvette handling errors.

Affects accessory selection and plumbing layout when automating sample introduction.

Source
“It includes a novel three-channel pump that fills and rinses up to three flow cells at the same time, reducing sample measurement time and removing operator-induced errors”
Performance

Spectral bandwidth is variable from 0.1 to 5 nm, allowing resolution to be matched to the application.

Determines whether the instrument meets resolution requirements for narrow-peak or high-absorbance measurements.

Source
“The spectral bandwidth of the Cary 3500 UV-Vis engine is variable and can be adjusted from (0.1 to 5 nm).”
Performance

The Flexible module measures absorbances exceeding 6 AU, enabling high-concentration analysis with fewer dilutions.

Reduces dilution steps and errors when specifying for concentrated samples.

Source
“The Cary 3500 Flexible UV-Vis spectrophotometer provides high linear dynamic range across the wavelength range, with absorbances exceeding 6 absorbance units.”
Installation

Variable-path-length rectangular and cylindrical cell holders change path length toollessly among 2, 4, 5 and 10 cm, eliminating alignment procedures.

Simplifies accessory setup and reduces operator alignment errors during installation and use.

Source
“This toolless design eliminates the time-consuming alignment procedure and can be used for many applications”
Installation

Waterless Peltier temperature control up to 110 °C removes the need for a water circulator, avoiding plumbing, flood risk and maintenance.

Affects bench space planning and eliminates water supply/drain requirements at installation.

Source
“The integrated, air-cooled, Peltier-driven temperature control does not require a space-consuming water circulator.”
Performance

Temperature can be controlled from –5 to 110 °C with ramp rates up to 40 °C per minute without loss of accuracy.

Confirms the temperature range and ramp capability needed for thermal denaturation and ramping experiments.

Source
“It provides fast and accurate temperature measurements between –5 to 110 °C.”
Performance

The Multicell module scans the full wavelength range on all eight channels in less than one second.

Sets throughput expectations for simultaneous standards, samples and controls.

Source
“The Cary 3500 Multicell UV-Vis spectrophotometer allows customers to simultaneously scan a full wavelength range on all eight channels in less than one second.”