Screens for login, settings, protocols, diagnostics, run data
Procedurepp. 91-94
Troubleshooting suggestions for labware and library issues
pp. 95-96
Document revision history and part number
p. 9table, 29 rows
p. 25table, 1 rows
p. 95table, 2 rows
p. 11table, 3 rows
p. 21table, 4 rows
p. 24table, 1 rows
p. 26table, 11 rows
p. 59table, 8 rows
p. 63table, 7 rows
p. 67table, 5 rows
Table 1 Technical specifications of the instrument
System Component
Specification
Thermal cycler module
Minimum temperature of thermal block
4°C (39.2°F)
Maximum temperature of thermal block
99°C (210.2°F)
Heater/Shaker/Magnet module
Heater
Temperature maximum
75°C (167°F)
Shaker
Maximum speed (rpm)
1800 rpm ±5%
Chiller module
Temperature range
4–12°C (39.2–53.6°F)
Power inlet
AC voltage
100–240 VAC ± 10%
AC frequency
50/60 Hz
Maximum power
1000 W
Input/output (I/O) connectors
USB 2.0 port
Maximum voltage rating
5 VDC
LAN port (for Cat 5 cable‡)
Maximum voltage rating
3.3 VDC
System
ISM Classification
ISM Group 1 Class A According to CISPR 11
Acoustic sound pressure
70 dBA
Dimensions, instrument door closed (Height × Depth × Width)
71 cm × 72 cm × 62 cm (28 inches × 28 inches × 24 inches)
Dimensions, instrument door open (Height × Depth × Width)
107 cm × 72 cm × 62 cm (42 inches × 28 inches × 24 inches)
Weight
95 kg (209 lbs)
Environmental conditions*
Temperature
Operating: 15°C to 25°C (53.6°F to 77°F) Transport and storage: –40°C to 70°C (–40°F to 158°F)
Humidity
Operating: 30% to 70%, non-condensing
Altitude
2000 m (6562 feet)
* The listed conditions apply to instrument operation. Conditions required for assay performance may differ. ‡The maximum length of the LAN cable used for EMC testing is 1.5 meters.
Table 5 Preparatory activities (continued)
Step
Description
На странице показана только шапка таблицы (Step, Description); строки данных на данной странице отсутствуют.
Revision Log
Revision
Change
E.01
• Additional details added to the Warning statement on moving the instrument with a forklift or lift table. • Added 10% allowance to AC voltage specification.
F.00
• New troubleshooting guidance on library collection error. • Updated firmware screen images for version 1.5.1. • Symbol usage updated per EN ISO 15223-1:2021/A1.
Table 3 Recommended supplies for cleaning the system instrument
Description
Purpose
Vendor
Diluted bleach (10%) wipes
Surface-cleaning of the instrument deck
Hype-Wipe Bleach Towelettes, or equivalent
Alcohol (70%) wipes
Surface-cleaning of the instrument exterior and instrument deck
VWR Pre-Moistened Clean Wipes, or equivalent
Dry, lint-free, scratch-free, laboratory wipes
Surface-cleaning of the barcode scanner window
Kimwipes, or equivalent
Table 4 Colors and descriptions of the LED indicator lights in the plate filling area
LED Color
Instrument Status
Description
White
Ready
The lights are illuminated white when the system is idle but the door is open, when the system is performing auto-teaching or teach-point verification, and when a user is setting up a protocol run.
Blue
Ready
The lights are illuminated blue anytime the system is idle and the door is closed, including at completion of a protocol run. The lights are also blue when the system is performing diagnostic tests.
Green
Running
The lights are illuminated green when the system is running a protocol.
Red
Error
The lights are illuminated red when the system has encountered an error. Check the touchscreen for an error message containing further details.
WARNING: During UV decontamination, the indicator lights are off and the instrument deck is illuminated by the UV light. Do not look directly at the UV light.
Table 5 Preparatory activities
Step
Description
Hardware initialization
During hardware initialization, the system returns all motorized parts (i.e., the gantry, HSM module, and thermal cycler) back to their home positions. Additionally, the system checks for the presence of any tips on the micropipettor and, if necessary, disposes of those tips in the tip waste bin.
Table 6
Action
Permitted by Standard users?
Permitted by Advanced users?
Set up and run protocols
Yes
Yes
View data from their own protocol runs
Yes
Yes
Update the number of PCR cycles during run setup
No
Yes
View data from protocol runs of other users
No
Yes
Edit user accounts of other users
No
Yes
Edit chiller temperature setting
No
Yes
Install firmware updates
No
Yes
Install protocol updates
No
Yes
Delete diagnostic reports
No
Yes
Change default protocol version
No
Yes
Bypass reagent expiration error
No
Yes*
* Only permissible on the G9710A Magnis NGS Prep System.
Table 7 Descriptions of the top and bottom banner elements
Element
Description
Top banner
Instrument status
Displays the status of the instrument (Ready, Running, or Error). See “Instrument Status Indicator Lights” on page 21 for more information on the possible instrument statuses.
Bottom banner
Home icon
Provides access to the Home screen. Press to exit the current screen and navigate to the Home screen.
User name
Displays the user name of the currently logged-in user. Press to access a button for logging out.
Door status
Indicates the position and lock status of the instrument door (Door Opened, Door Closed, or Door Unlocked).
Error icon
Indicates the occurrence of a system error or failed diagnostic test. Press to view information on the error(s).
Time and date
Displays the time and date according to the system settings. Press to open the Date & Time Settings screen.
Table 8
Settings screen buttons
Button
Description
User Management
Opens the “User Management screen” for viewing user accounts and accessing tools for adding and editing user accounts.
System Settings
Opens the “System Settings screen” for accessing tools to configure system-wide settings.
Export
Opens the “Export Files screen” for exporting post-run data files and log files.
Protocol Update
Opens the “Protocol Update screen” for installing new protocol files on the system.
Auto Teaching
Opens the “Auto Teach screen” for running auto-teaching.
Self Diagnostic
Opens the “Instrument Diagnostic screen” for running diagnostic tests and viewing reports for diagnostic tests and instrument health checks.
Hardware Usage Tracking
Opens the “Hardware Usage Tracking screen” for viewing a usage counter for the UV-tube.
Table 9
System Settings screen buttons
Button
Description
Instrument Settings
Opens the “Instrument Settings screen” for viewing and/or setting the instrument name, serial number, firmware software version, and device version.
Date & Time Settings
Opens the “Date & Time Settings screen” for setting the date and time of the system.
Chiller Setting
Opens the “Chiller Setting screen” for setting the temperature of the chiller module.
Other Settings
Opens the “Other Settings screen” which has a setting to verify teach points during the IHC.
Firmware Update
Opens the “Firmware Update screen” for installing a new version of the firmware software from a connected USB drive.
Engineering Facts
14 facts · every claim linked to a page
Installation
The instrument must be installed by an Agilent engineer or authorized service provider, not by the end user.
Installation labor and scheduling must be planned with Agilent before delivery.
Source
“The system must be installed by an Agilent engineer or Agilent authorized service provider.”
Installation
Moving the instrument requires a forklift or lift table rated for at least 100 kg plus two people for final positioning.
Site access and rigging equipment must be arranged for the 95 kg unit.
Source
“To move the instrument, use an automated forklift or lift table that can lift at least 100 kg (220 lbs).”
Installation
Clearance of 10 cm on each side and 18 cm at the back must be maintained for ventilation.
Bench layout must reserve these clearances to avoid overheating.
Source
“Maintain 10 cm (4 inches) of cleared space on each side of the instrument and 18 cm (7 inches) of cleared space at the back of the instrument.”
Installation
The instrument needs a grounded, dedicated or independent AC circuit and must not share a circuit with high-current devices.
Circuit planning must account for the 1000 W load and avoid nuisance trips.
Source
“If possible, connect the instrument to an independent or dedicated AC circuit.”
Installation
Only the Agilent-supplied regional power cord may be used; substitutes are prohibited.
The correct regional cord must be ordered and kept with the unit.
Source
“Do not substitute with a power cord from another source.”
Operation
The instrument must be placed in a post-PCR laboratory environment.
Room assignment must follow the pre-/post-PCR workflow zoning.
Source
“Place the instrument in a post-PCR laboratory environment.”
Safety
Electrostatic discharges above 8000 V can disrupt the USB ports, so anti-static precautions are required in high-static areas.
ESD controls (wrist strap, flooring) must be planned for the installation site.
Source
“Electrostatic discharges greater than 8000 volts may interfere with the normal operation of the USB ports on the instrument.”
Operation
USB ports must not be used while a protocol is running and cannot accept encrypted drives.
File transfer procedures must be scheduled outside of runs.
Source
“Do not use USB ports while a Magnis protocol is running.”
Operation
Prepared libraries are held at 12 °C for up to 72 hours and must be collected within that window.
Workflow scheduling must guarantee sample retrieval within 72 hours.
Source
“which is held at 12°C for up to 72 hours. Collect the samples within that 72-hour period.”
Operation
The chiller stays at its set temperature for 2 hours after library-ready notification before shutting off automatically.
Operators must retrieve libraries within the 2-hour chiller countdown.
Source
“The chiller remains at its set temperature (12°C, by default) for 2 hours before automatically shutting off.”
Maintenance
No calibration is required, but annual preventive maintenance by Agilent is mandatory.
Service contracts and annual downtime must be budgeted.
Source
“The Magnis/MagnisDx NGS Prep System does not require calibration, however, it does require preventive maintenance annually.”
Maintenance
The barcode scanner window should be cleaned only with a soft cloth and water or mild detergent, never touched otherwise.
Cleaning SOPs must protect the scanner window from damage.
Source
“Do not touch the window with anything other than the soft cloth or wipe that you are using for cleaning.”
Performance
Operating humidity outside 30–70% non-condensing can degrade instrument performance.
HVAC requirements must be verified for the lab room.
Source
“Operating the system in humidity levels outside of this range can impact performance.”
Performance
Vibration-generating or vibration-sensitive equipment nearby can impact performance.
Bench placement must avoid adjacent centrifuges and similar equipment.
Source
“Proximity to lab equipment that generates vibrations can impact performance.”