Agilent Technologies MagnisDx NGS Prep System
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies MagnisDx NGS Prep System

Agilent Technologies MagnisDx NGS Prep System — Next generation sequencing with Temperature 15 °C.

The MagnisDx NGS Prep System (K1007A) is an automated liquid-handling platform for IVD use, processing human nucleic acid samples from total RNA or fragmented gDNA through library preparation and target enrichment. It integrates a 75 °C heater, shaker, LCD touchscreen, and USB/LAN interfaces, and is validated with Agilent Magnis NGS kits. It differs from the research-use-only Magnis (G9710A) by its in vitro diagnostic labeling.

Temperature
15 °C

Agilent Technologies NGS Automation Platforms Series — Model Comparison & Selection

Model Temperature, °COperating temperature, °CAcoustic pressure, dBElectrostatic discharge immunity, voltsLabware temperature stability, °CLibrary storage duration after run, hours
MagnisDx NGS Prep System Current 157570800012072
Magnis NGS Prep System 157570800012072

Full specifications

Core Performance & Functional Specifications
Temperature 15 °C
Electrostatic Discharge Immunity 8000 volts
Library Storage Duration After Run 72 hours
Lifting capacity 100 kg
Temperature settings 75 °C
Uv Decontamination Extended Cycle Duration 2 hours
Uv Decontamination Quick Cycle Duration 30 minutes
Shaker maximum speed mfr 1800 % rpm
Intended use mfr Automated liquid handling system for NGS library
End result mfr Target-enriched DNA library ready for sequencing
Validated reagents mfr Agilent Magnis NGS kits kits
Materials provided mfr Instrument with preloaded touchscreen software;
Cleaning supplies mfr Alcohol wipes; dry, lint-free, scratch-free wipe
Dimensions, instrument door closed (H × D × W) mfr 71 cm
Dimensions, instrument door open (H × D × W) mfr 107 cm
Clearance, sides mfr 4 inches
Clearance, back mfr 7 inches
Clearance, front mfr 2 inches
Clearance, top mfr 44 inches
Lifting capacity for moving instrument mfr 220 lbs
Maximum shaker speed mfr 1 % tr/min
Maximum speed (rpm) mfr 1800
ISM Classification mfr 1
Diluted bleach (10%) wipes mfr Surface-cleaning of the instrument deck deck
Alcohol (70%) wipes mfr Surface-cleaning of the instrument exterior and
Dry, lint-free, scratch-free, laboratory wipes mfr Surface-cleaning of the barcode scanner window w
White mfr Ready Ready
Blue mfr Ready Ready
Green mfr Running Running
Red mfr Error Error
Set up and run protocols mfr Yes Yes
Update the number of PCR cycles during run setup mfr No Yes Yes
Edit user accounts of other users mfr No Yes Yes
Install firmware updates mfr No Yes Yes
Delete diagnostic reports mfr No Yes Yes
Bypass reagent expiration error mfr No Yes*
Home icon mfr Provides access to the Home screen. Press to exi
User name mfr Displays the user name of the currently logged-i
Door status mfr Indicates the position and lock status of the in
Error icon mfr Indicates the occurrence of a system error or fa
Time and date mfr Displays the time and date according to the syst
User Management mfr Opens the “User Management screen” for viewing u
System Settings mfr Ouvre la Fenêtre « System Settings » permettant
Export mfr Opens the “Export Files screen” for exporting po
Auto Teaching mfr Ouvre la Fenêtre « Auto Teach » permettant d’exé
Hardware Usage Tracking mfr Ouvre la Fenêtre « Hardware Usage Tracking » per
Date & Time Settings mfr Opens the “Date & Time Settings screen” for sett
Chiller Setting mfr Ouvre la Fenêtre « Chiller Setting » permettant
Other Settings mfr Opens the “Other Settings screen” which has a se
Firmware Update mfr Opens the “Firmware Update screen” for installin
Lift capacity required for moving mfr 100 kg
Side clearance mfr 10 cm
Back clearance mfr 18 cm
Chiller hold time before auto shutoff mfr 2 hours
Clearance side mfr 10 cm
Clearance back mfr 18 cm
Clearance front mfr 5 cm
Clearance top mfr 111 cm
ESD threshold mfr 8 volts
Module du thermocycleur mfr Température maximale du bloc thermique thermique
Module chauffage/agitateur/aimant mfr Vitesse maximale maximale
Module refroidisseur mfr Plage de température rature
Prise d’alimentation mfr Puissance maximale maximale
Connecteurs d’entrée/de sortie (E/S) mfr Tension nominale maximale maximale
Système mfr Poids Poids
Conditions environnementales* mfr Altitude Altitude
Configuration et exécution de protocoles mfr Oui Oui
Affichage des données des protocoles exécutés par eux-mêmes mfr Oui Oui
Mise à jour du nombre de cycles de réaction de polymérisation en chaîne lors de la configuration du protocole mfr Non Oui Oui
Affichage des données des protocoles exécutés par d’autres utilisateurs mfr Non Oui Oui
Modification des comptes d’autres utilisateurs mfr Non Oui Oui
Installation des mises à jour du firmware mfr Non Oui Oui
Installation des mises à jour de protocoles mfr Non Oui Oui
Suppression des rapports de diagnostic mfr Non Oui Oui
Modification de la version par défaut des protocoles mfr Non Oui Oui
Contournement de l’erreur de péremption du réactif mfr Non Oui*
Date & time Settings mfr Ouvre la Fenêtre « Date & Time Settings » permet
Lingettes à l'eau de Javel diluée (10 %) mfr Nettoyage en surface du plateau de l'instrument
Lingettes imbibées d'alcool (70 %) mfr Nettoyage en surface de l'extérieur et du platea
Lingettes de laboratoire douces, non pelucheuses et sèches mfr Nettoyage de surface de la fenêtre du lecteur de
Blanc mfr « Ready »
Bleu mfr « Ready »
Vert mfr « Running »
Rouge mfr « Error »
Icône en forme de maison mfr Permet d’accéder à la fenêtre « Home ». Appuyez
Nom d’utilisateur mfr Affiche le nom de l’utilisateur actuellement con
État de la porte mfr Indique la position et l’état de verrouillage de
UV tube lifetime mfr 630 heures
Post-run chiller hold time mfr 2 heures
Post-run sample hold time mfr 72 heures
Lift capacity required mfr 100 kg
Rear clearance mfr 18 cm
Dimensions, Weight & Mechanical Design
Weight 95 kg
Dimensions, instrument door closed (height × depth × width) mfr 71 cm
Dimensions, instrument door open (height × depth × width) mfr 107 cm
Dimensions, instrument door closed (Height × Depth × Width) mfr 71 cm
Dimensions, instrument door open (Height × Depth × Width) mfr 107 cm
Electrical, Power & Drive Specifications
Power consumption 1000 W
Power supply 100...240 VAC
AC voltage mfr 100...240
AC frequency mfr 50 Hz
ISM classification mfr 1 A
Power supply frequency mfr 50 Hz
Maximum voltage rating mfr 3.3 VDC
Interfaces, Connections & Control
USB 2.0 port maximum voltage rating mfr 5 VDC
Control interface mfr LCD touchscreen touchscreen
USB 2.0 port maximum nominal voltage mfr 5 VCC
LAN port maximum nominal voltage mfr 3.3 VCC
LAN port maximum voltage rating mfr 3.3 VDC
USB port voltage mfr 5 VDC
Ethernet port voltage mfr 3.3 VDC
USB 2.0 port maximum rated voltage mfr 5 VCC
LAN port maximum rated voltage mfr 3.3 VCC
View data from their own protocol runs mfr Yes Yes
View data from protocol runs of other users mfr No Yes Yes
Install protocol updates mfr No Yes Yes
Change default protocol version mfr No Yes Yes
Protocol Update mfr Ouvre la Fenêtre « Protocol Update » permettant
Electrostatic discharge threshold affecting USB mfr 8000 volts
Install firmware and protocol updates mfr No Yes Yes
Materials, Media & Process Compatibility
Starting material mfr Total RNA or fragmented genomic DNA purified fro
Operating Conditions, Environment & Protection
Acoustic Pressure 70 dBA
Labware Temperature Stability 120 °C
Library Storage Temperature After Run 12 °C
Maximum Altitude 2000 m
Maximum temperature 75 °C
Minimum Temperature Of Thermal Block 4 °C
Sound pressure 70 dBA
Operating humidity mfr 30...70 condensing
Humidity mfr 30 condensing
Maximum altitude for instrument operation mfr 2000 m

Description

Construction and series positioning

The MagnisDx NGS Prep System is the in vitro diagnostic (IVD) variant of the Magnis NGS Prep series; its sibling, the Magnis NGS Prep System (G9710A), is labeled for research use only. Both are automated liquid-handling systems intended for NGS library preparation and/or target enrichment of human nucleic acid samples. The MagnisDx is distinguished by its IVD labeling and is supplied with preloaded touchscreen software, a power cord, and a functional test certificate. The instrument includes a heater with a maximum temperature of 75 °C, a shaker, an LCD touchscreen for control, and USB 2.0 and LAN ports.

Working range and sample input

The system accepts total RNA or fragmented genomic DNA purified from cell or tissue samples, blood samples, or FFPE samples. The end result is a target-enriched DNA library ready for sequencing. Validated reagents are Agilent Magnis NGS kits. The heater supports processes up to 75 °C, and the shaker provides mixing; the maximum shaker speed is listed as 1 tr/min in the available specifications, which is insufficient to define a usable mixing range and is therefore not further characterized here.

Control, interfaces, and operation

Operation is via an LCD touchscreen with preloaded software. The USB 2.0 port has a maximum nominal voltage of 5 V, and the LAN port has a maximum nominal voltage of 3.3 V. The user guide (GEN MAN K1007 90000, 90010, 90011) covers both the Magnis and MagnisDx systems; detailed installation, connection, and operating procedures are provided in those documents. No certificates or applications are listed in the verified facts.

Integration & Operating Notes

Electrical
Do not exceed 5 V on the USB 2.0 port

The USB 2.0 port is rated for a maximum nominal voltage of 5 V. Connecting a device that sources or applies a higher voltage can damage the port or the instrument. Verify that any USB peripheral or adapter is compatible with this limit before connection.

Factory Specification Sheet
Electrical
LAN port maximum nominal voltage is 3.3 V

The LAN port is rated for a maximum nominal voltage of 3.3 V. Use only standard Ethernet equipment that complies with this limit; non-compliant PoE injectors or miswired cables may exceed the rating and cause damage.

Factory Specification Sheet
Process Limits
Heater temperature must not exceed 75 °C

The heater has a maximum temperature of 75 °C. Process protocols must be validated within this limit; exceeding it may degrade samples or reagents and is outside the instrument's specified operating range.

Factory Specification Sheet
Process Limits
Use only Agilent Magnis NGS kits for validated runs

The system is validated with Agilent Magnis NGS kits. Substituting other reagents may produce invalid results and is not supported by the IVD labeling. Ensure kit compatibility before starting a run.

Factory Specification Sheet
Maintenance
Clean with alcohol wipes and lint-free wipes only

The materials provided include alcohol wipes and dry, lint-free, scratch-free wipes for cleaning. Avoid abrasive or solvent-based cleaners that could damage the touchscreen or instrument surfaces. Follow the user guide for cleaning frequency and procedure.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the MagnisDx NGS Prep System instead of the Magnis NGS Prep System?

Choose the MagnisDx (K1007A) when the workflow requires in vitro diagnostic use, as it is labeled for IVD. The Magnis (G9710A) is for research use only and must not be used for diagnostic procedures. Both automate NGS library prep and target enrichment with Agilent Magnis NGS kits.

What are the interface voltage limits for connecting peripherals to the MagnisDx?

The USB 2.0 port has a maximum nominal voltage of 5 V, and the LAN port has a maximum nominal voltage of 3.3 V. Ensure any connected device or network equipment does not exceed these ratings to avoid damage.

What maintenance supplies are required, and what is included in the base delivery?

The base delivery includes the instrument with preloaded touchscreen software, a power cord, and a functional test certificate. Cleaning requires alcohol wipes and dry, lint-free, scratch-free wipes, which are not listed as included and must be supplied separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Magnis and MagnisDx NGS Prep System User Guide
Manual ·3,1 МБ ·96 pp.
View PDF
PDF
Magnis/MagnisDx NGS Prep System Quick Reference Guide
Manual ·497,6 КБ ·17 pp.
View PDF
PDF
Magnis and MagnisDx NGS Prep System User Guide
Manual ·4,2 МБ ·100 pp.
View PDF