Technical Reference Summary

Covered Models & Document Scope

This operating manual (Issue 01/2025, Art. No. 7001-0001) covers BINDER incubators B 28 and drying/heating ovens E 28 with mechanical adjustment. Variants: B 28 (B028-230V, Art. 9010-0002 / 9110-0002), B 28 120 V (B028-120V, Art. 9010-0067 / 9110-0067), and E 28 (E028-230V, Art. 9010-0001 / 9110-0001).

Section-by-Section Technical Breakdown

pp. 16-18
Chamber description and overview

The chambers are electrically heated and naturally ventilated. The inner chamber, pre-heating chamber and interior door sides are stainless steel V2A (1.4301 / AISI 304); the housing is powder-coated RAL 7035 or RAL 9003. B 28 is an incubator with control thermometer, thermostat, main power switch, pilot lamps and thermostat brake; E 28 is a drying/heating oven with a 0–120 minute timer.

  • B 28 front elements: (1) control thermometer, (2) main power switch, (3) green pilot lamp (ready to operate), (4) thermostat, (5) yellow pilot lamp (heating active), (6) thermostat brake.
  • Chamber rear: (8) chamber fuses (230 V chambers: 2 fuses; 120 V chambers: 1 fuse), (9) power cable.
  • Temperature range at ambient +18 °C to +40 °C: B 28 setting range 30 °C to 70 °C, control range 5 °C above room temperature up to 70 °C; E 28 60 °C to 230 °C.
  • E 28 operation above 150 °C may cause harmless yellowish-brown or blue discoloration of metallic surfaces by natural oxidation.
pp. 19-20
Installation location and ambient conditions

The chamber must be set up indoors on an even, non-flammable, vibration-free surface, aligned with a spirit level, in a well-ventilated dry location. Minimum clearances and ambient limits are specified to avoid overheating and to ensure performance.

  • Permissible ambient temperature during operation: +18 °C to +40 °C; technical data relate to +22 °C ± 3 °C.
  • Permissible ambient humidity: max. 70 % r.H., non-condensing.
  • Maximum installation height: 2000 m above sea level.
  • Minimum distances: 250 mm between chambers side by side; rear wall 100 mm; sides 160 mm; at least 200 mm above the chamber.
  • Do not place anything on top of the chamber and do not cover the air exhaust slots.
  • Chambers are NOT intended for stacking.
  • The power plug must remain easily accessible to completely separate the chamber from the power supply.
pp. 18, 21
Electrical connection

The chambers are supplied ready for connection with a fixed power cable of at least 1800 mm length. Connection data depend on model and voltage variant.

  • B 28 and E 28: grounded plug, 230 V at 50 Hz or 230 V at 60 Hz, nominal voltage tolerance ±10 %.
  • B 28 120 V variant: 120 V supply (model B028-120V).
  • 230 V chambers have 2 chamber fuses; 120 V chambers have 1 chamber fuse.
  • E 28 type plate example: 0.80 kW, 3.5 A, 230 V / 50 Hz or 230 V / 60 Hz, 1 N~.
pp. 22-23
Start-up, temperature setting, fresh air and hot-air sterilization

Start-up differs between B 28 (main power switch) and E 28 (0–120 minute timer). Temperature is set via the mechanical thermostat; fresh air supply and hot-air sterilization are described for E 28.

  • B 28 is turned on by the main power switch; E 28 by the 0–120 minutes timer (position ∞ or selected time).
  • Temperature setting is performed at the thermostat; the control thermometer displays the actual temperature.
  • Fresh air supply is adjustable.
  • E 28 can be used for hot-air sterilization.
pp. 23
Overtemperature protective device (Class 1)

The chambers are equipped with a temperature safety device corresponding to Class 1 according to DIN 12880:2007. The device limits chamber temperature and is set using the thermostat brake on B 28.

  • Temperature safety device Class 1 according to DIN 12880:2007.
  • B 28: thermostat brake (6) is used to set the overtemperature protection.
  • If the overtemperature protective device has tripped, the chamber is switched off and heating stops.
pp. 24-26
Cleaning and decontamination

Cleaning and chemical disinfection/decontamination procedures are specified for the chamber. Water must not enter the chamber, and cleaning while the chamber is switched on is prohibited.

  • The chamber must NOT become wet during operation, cleaning or maintenance.
  • Do not use cleaning agents that could damage stainless steel or powder-coated surfaces.
  • Cleaning activity while the chamber is turned on is a foreseeable misuse.
pp. 27-28
Maintenance, service and electrical testing

Regular maintenance is required at least once a year and must be performed by licensed electricians or experts authorized by BINDER. Electrical repeat inspection is required annually and after maintenance or repair.

  • Maintenance at least once a year; warranty void if performed by non-authorized personnel.
  • Before maintenance, turn off the chamber at the main power switch (B 28) or with the timer (E 28) and disconnect the power plug.
  • Do NOT remove the rear panel of the chamber.
  • Replace the door gasket only when cold.
  • Annual repeat electrical inspection and tests before initial start-up and after maintenance/repair; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702.
  • After maintenance, the chamber must be tested before resuming operation.
pp. 28
Simple troubleshooting

Troubleshooting by operating personnel without technical intervention or disassembly. If an error cannot be clearly identified, BINDER Service must be contacted; repairs only by BINDER-authorized qualified service personnel.

  • Check power supply, main power switch/timer position, correct voltage, chamber fuse, and timer setting.
  • If the chamber fuse responds again after replacement, contact BINDER Service.
  • If the overtemperature protective device Class 1 has turned off the chamber, heating is disabled.
  • Defective Pt 100 sensor or semiconductor relay requires turning off the chamber, pulling the power plug and contacting BINDER Service.
  • E 28 heating-up time deviations may be caused by full loading; reduce load or allow longer heating-up times.
pp. 32-36
Technical description and data

Technical data for B 28 and E 28 include temperature ranges, nominal power, current, voltage, IP protection, usable volume definition, overcurrent protection, equipment/options and accessories/spare parts.

  • B 28 temperature range: 30 °C to 70 °C (control range 5 °C above ambient to 70 °C).
  • E 28 temperature range: 60 °C to 230 °C.
  • E 28 example type plate: nominal temperature 230 °C / 446 °F, nominal power 0.80 kW, nominal current 3.5 A, 230 V / 50 Hz or 60 Hz, IP 20, Class 1.
  • Factory calibration and adjustment are described.
  • Usable volume definition and overcurrent protection are specified.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
B 28 temperature setting range is 30 °C to 70 °C; control range is 5 °C above room temperature up to 70 °C at ambient +18 °C to +40 °C. p.16 ↗ Select B 28 for incubation near 37 °C; ensure ambient temperature is at least 5 °C below the desired set-point. B 28
E 28 temperature range is 60 °C to 230 °C at ambient +18 °C to +40 °C. p.16 ↗ E 28 is suitable for drying and heat treatment up to 230 °C; do not use below 60 °C. E 28
Minimum distances: 250 mm between chambers side by side, rear wall 100 mm, sides 160 mm, at least 200 mm above the chamber. p.20 ↗ Plan installation space accordingly to prevent overheating and ensure natural ventilation. B 28 and E 28
Permissible ambient temperature during operation: +18 °C to +40 °C; technical data relate to +22 °C ± 3 °C. p.20 ↗ Performance specifications may deviate outside the reference ambient conditions; avoid elevated ambient temperatures. B 28 and E 28
Permissible ambient humidity: max. 70 % r.H., non-condensing; maximum installation height 2000 m above sea level. p.20 ↗ Do not install in damp areas or at high altitude beyond the limit. B 28 and E 28
Power cable length at least 1800 mm; 230 V at 50 Hz or 60 Hz, ±10 %; grounded plug. p.21 ↗ Provide a nearby grounded outlet; ensure the plug remains accessible for emergency disconnection. B 28 and E 28 230 V variants
230 V chambers have 2 chamber fuses; 120 V chambers have 1 chamber fuse. p.18 ↗ Use the correct number and rating of fuses when replacing; if a fuse responds again, contact BINDER Service. B 28 and E 28
E 28 type plate example: nominal power 0.80 kW, nominal current 3.5 A, nominal voltage 230 V / 50 Hz or 60 Hz, IP protection 20, safety device Class 1 per DIN 12880. p.8 ↗ Verify electrical supply capacity and protection; IP 20 means no water protection. E 28
Temperature safety device Class 1 according to DIN 12880:2007 is installed. p.23 ↗ The chamber provides overtemperature protection; the device must not be bypassed or changed. B 28 and E 28
Maintenance at least once a year by licensed electricians or BINDER-authorized experts; annual electrical repeat inspection recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.27 ↗ Plan annual service and electrical safety testing; warranty void if maintenance by non-authorized personnel. B 28 and E 28
Replace the door gasket only when cold. p.27 ↗ Allow the chamber to cool before gasket replacement to avoid damaging the gasket. B 28 and E 28
Transport and storage ambient temperature range: -10 °C to +60 °C; storage humidity max. 70 % r.H., non-condensing. p.19 ↗ After storage in a cold location, wait at least one hour at ambient temperature before start-up to avoid condensation. B 28 and E 28
Chambers are NOT intended for stacking. p.20 ↗ Do not place chambers on top of each other; provide separate floor or bench space. B 28 and E 28
Inner chamber, pre-heating chamber and interior door sides are stainless steel V2A (1.4301 / AISI 304); housing is powder-coated RAL 7035 or RAL 9003. p.16 ↗ Avoid corrosive charging materials such as acids and halides that may damage stainless steel. B 28 and E 28
E 28 operation above 150 °C may cause harmless yellowish-brown or blue discoloration of metallic surfaces by natural oxidation. p.16 ↗ Discoloration is not a malfunction and does not impair function or quality. E 28
The chamber must NOT become wet during operation, cleaning or maintenance; do not install in damp areas or puddles. p.10 ↗ Set up splash-proof; IP 20 provides no water ingress protection. B 28 and E 28
Do NOT remove the rear panel of the chamber; before maintenance turn off at main power switch (B 28) or timer (E 28) and disconnect power plug. p.27 ↗ Only authorized personnel may perform maintenance; live parts are accessible if the rear panel is removed. B 28 and E 28
B 28 front panel includes a thermostat brake (6) for setting the overtemperature protection. p.17 ↗ Use the thermostat brake to set the Class 1 temperature safety device; do not confuse with the main thermostat. B 28

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Chamber without function. Green pilot lamp (3) is off. p.28 ↗ B 28: Main power switch (2) in position I. E 28: Timer in position ∞ or selected time. No power supply. Check connection to power supply. Check whether the chamber is turned on at the main power switch or the timer. Section 8.3
Chamber without function. Green pilot lamp (3) is off. p.28 ↗ Wrong voltage. Check power supply for correct voltage. Section 8.3
Chamber without function. Green pilot lamp (3) is off. p.28 ↗ Chamber fuse has responded. Check chamber fuse and replace it if appropriate. If it responds again, contact BINDER service. Section 8.3
Chamber without function. Green pilot lamp (3) is off. p.28 ↗ E 28: Timer expired. Check timer setting. Section 8.3
Chamber without function. Green pilot lamp (3) is off. p.28 ↗ Chamber defective. Contact BINDER service. Section 8.3
Chamber without function. Green pilot lamp (3) is off. p.28 ↗ Overtemperature protective device class 1 has turned off the chamber. Heating Not specified in table; contact BINDER Service if fault persists. Section 8.3
Set-point temperature is not reached after specified time. p.28 ↗ Chamber door not properly closed. Completely close chamber door. Section 8.3
Set-point temperature is not reached after specified time. p.28 ↗ Door gasket defective. Replace door gasket. Section 8.3
Set-point temperature is not reached after specified time. p.28 ↗ E 28: Timer defective Contact BINDER service. Section 8.3
Chamber heating permanently, set-point exceeded. p.28 ↗ Pt 100 sensor defective. Turn off the chamber and pull the power plug. Contact BINDER service. Section 8.3
Chamber heating permanently, set-point exceeded. p.28 ↗ Semiconductor relay defective. Turn off the chamber and pull the power plug. Contact BINDER service. Section 8.3
Chamber doesn’t heat up. Green pilot lamp (3) is lit. p.28 ↗ Heating element defective. Contact BINDER service. Section 8.3
Chamber doesn’t heat up. Green pilot lamp (3) is lit. p.28 ↗ Semiconductor relay defective. Contact BINDER service. Section 8.3
E 28: Deviations from the indicated heating-up times. p.28 ↗ Chamber fully loaded. Load the chamber less or consider longer heating-up times. Section 8.3