BINDER B 28
BINDER B 28 — Incubators with Temperature 30…70 °C.
The BINDER B 28 is a 28-litre naturally ventilated incubator with mechanical thermostat and Class 1 overtemperature protection, set for 30–70 °C. It differs from the E 28 oven sibling by its low-temperature incubation range and thermostat brake. It serves microbiology and sample incubation where precise, stable warmth near 37 °C is required.
Direct drop-in replacements
Full specifications
| Temperature | 30...70 °C |
| Interior volume | 28 L |
| Load per rack | 10 kg |
| Nominal power | 0.25 kW |
| Permitted load | 25 kg |
| Temperature fluctuation | 1.0 K |
| Temperature uniformity | 1.5 K |
| Back wall clearance | 160 mm |
| Side wall clearance | 100 mm (B028-230V / B028-120V) |
| Article Number mfr | 9010 |
| Depth | 425 mm (B028-230V / B028-120V) |
| Height | 405 mm (B028-230V / B028-120V) |
| Interior width | 400 mm (B028-230V / B028-120V) |
| Width | 580 mm (B028-230V / B028-120V) |
| Net weight of the unit (empty) mfr | 23 kg |
| Power frequency | 50 Hz 120 V UL variant: 60 Hz |
| Rated voltage | 230 V 120 V UL variant: 120 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
The B 28 shares its stainless-steel V2A (1.4301 / AISI 304) inner chamber, pre-heating chamber and interior door sides with the E 28 drying/heating oven, but is configured as an incubator: mechanical thermostat, control thermometer, main power switch, pilot lamps and a thermostat brake for setting the Class 1 overtemperature safety device. The E 28 instead carries a 0–120 minute timer and a 60–230 °C range. The B 28 is therefore the low-temperature member of the 28-litre mechanical-adjustment pair, not a drying oven.
Working range and behaviour
- Temperature setting range: 30 °C to 70 °C; control range is 5 °C above room temperature up to 70 °C at ambient +18 °C to +40 °C.
- Temperature fluctuation 1.0 K and uniformity 1.5 K, referenced to +22 °C ± 3 °C ambient.
- Natural ventilation and electric heating; no forced-air circulation is specified.
- Class 1 temperature safety device per DIN 12880:2007 limits chamber temperature and is set via the thermostat brake.
Installation and operation
- Indoor, level, non-flammable, vibration-free surface; align with a spirit level.
- Clearances: 160 mm rear, 100 mm sidewise, 250 mm between chambers side by side, at least 200 mm above; not intended for stacking.
- Ambient limits: +18 °C to +40 °C, max. 70 % r.H. non-condensing, max. 2000 m altitude.
- Fixed power cable at least 1800 mm; 230 V 1~ 50/60 Hz (CEE 7/7) or 120 V 1~ 60 Hz (NEMA 5-15); 0.25 kW nominal; 6.3 A unit fuse; 2 chamber fuses on 230 V, 1 on 120 V.
- IP 20: no water protection; chamber must not become wet.
Integration & Operating Notes
Natural ventilation depends on free air around the housing; the manual specifies 160 mm at the rear, 100 mm at the sides, 250 mm between adjacent chambers and at least 200 mm above. The unit must not be stacked. Violating these distances risks overheating and invalidates the temperature uniformity figures.
The control range starts 5 °C above room temperature and ends at 70 °C, valid only for ambient +18 °C to +40 °C. A 37 °C incubation therefore requires ambient no higher than about 32 °C; in a warm room the chamber cannot regulate near the lower set-points.
The 230 V variant (Art. 9010-0002) uses a CEE 7/7 plug and two chamber fuses; the 120 V UL variant (Art. 9010-0067) uses a NEMA 5-15 plug and one chamber fuse. The 6.3 A unit fuse rating and 0.25 kW nominal power apply; a repeated fuse response requires BINDER Service rather than a higher-rated replacement.
The overtemperature protective device per DIN 12880:2007 is adjusted with the thermostat brake on the front panel, separate from the main thermostat. It must not be bypassed or altered; the brake is the only means of setting the safety limit on this mechanical model.
Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, with electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 annually and after any repair. The rear panel must not be removed by operating personnel; isolate power at the main switch and unplug before service.
The inner chamber and interior door sides are V2A (1.4301 / AISI 304). Charging materials containing acids or halides can attack the stainless steel; the chamber must not become wet during operation or cleaning, and cleaning while switched on is prohibited.
Selection & Integration Q&A
Pick the B 28 for incubation between 30 °C and 70 °C, typically near 37 °C, where the thermostat brake sets Class 1 overtemperature protection. Choose the E 28 only for drying or heat treatment from 60 °C to 230 °C with its 0–120 minute timer; the B 28 cannot reach those temperatures.
Provide a grounded 230 V 1~ 50/60 Hz outlet within reach of the fixed 1800 mm cable, using the CEE 7/7 plug. The unit draws 0.25 kW and carries a 6.3 A unit fuse plus two chamber fuses; the plug must remain accessible for emergency disconnection.
No stacking is permitted. Ambient must stay within +18 °C to +40 °C at max. 70 % r.H. non-condensing, and the set-point must be at least 5 K above ambient, so a warm room limits low-temperature incubation. Keep 160 mm rear, 100 mm side and 250 mm between chambers.
Plan annual maintenance by BINDER-authorized personnel plus electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. Door gaskets are replaced only when cold. No cables or accessories are included beyond the fixed 1800 mm power cable; fuses must match the variant (two for 230 V, one for 120 V).