KBF+KBF PRO (E7) 09 2025
Technical Reference Summary
Covered Models & Document Scope
This operating manual (KBF/KBF-UL/KBF PRO (E7), issue 09/2025, Art. No. 7001-0569) covers the constant climate chambers with program control of the KBF series (KBF 130, 260, 470, 720, 1060, 1600 and their -UL variants) and the KBF PRO series with enlarged temperature and humidity range (KBF PRO 130, 260, 470, 720, 1060, 1600). Art. Nos. are listed on page 1 (e.g. KBF 130 9020-0478/9120-0478, KBF 130-UL 9020-0494/9120-0494, KBF 260 9020-0479/9120-0479, KBF 260-UL 9020-0495/9120-0495, KBF PRO 130 9020-0439/9120-0439, etc.). The document also covers options and accessories such as APT-COM 4, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, flexible Pt 100 object temperature sensor, external freshwater/wastewater cans, and BINDER Pure Aqua Service.
Section-by-Section Technical Breakdown
Describes the chamber layout, including the triangular instrument panel for sizes 130 and 260, the signal triangle for sizes 470, 720, 1060 and 1600, the ergonomically adjustable control terminal for larger sizes, and the rear connection panel with water connections.
- Triangular instrument panel used on chamber sizes 130 and 260.
- Signal triangle used on chamber sizes 470, 720, 1060 and 1600.
- Ergonomically adjustable control terminal on sizes 470, 720, 1060 and 1600.
- Connection panel and water connections located on the chamber rear.
Covers wall spacer from size 260, flexible tilt protection kit for sizes 130 and 470, rack mounting with U-rails or telescopic rails, condensate collection pan for KBF/KBF-UL 130 up to serial number 20250000010225, wastewater and freshwater connections, and electrical connection.
- Spacer for wall distance required from chamber size 260 on.
- Flexible tilt protection kit for chamber sizes 130 and 470.
- Condensate collection pan applies to KBF/KBF-UL 130 up to serial number 20250000010225.
- Freshwater supply can be automatic via water pipe or manual via external freshwater can (option).
- Connection kit for water main and optional safety kit with hose burst protection and reflux protection.
Introduces the MB2 controller operating functions, display views (normal, program, chart-recorder), controller icons, operating modes, menu structure, entry principles, and behavior during power failures and door opening.
- Normal display, program display and chart-recorder display views.
- Operating modes include fixed value, timer program, time programs and week programs.
- Main menu, Settings submenu and Service submenu.
- Defined performance during and after power failures and when opening the door.
Describes turning on the chamber, controller settings upon start up, and turning humidity control on or off.
- Turning on the chamber.
- Controller settings upon start up.
- Turning humidity control on/off.
Explains entering set-points for temperature, humidity and fan speed through the Setpoints menu, direct setpoint entry via the Normal display, and special controller functions.
- Set-point entry for temperature, humidity and fan speed via the Setpoints menu.
- Direct setpoint entry for temperature and humidity via Normal display.
- Special controller functions available in fixed value mode.
Covers starting a timer program, performance during program delay time, cancelling a running timer program, and performance after the end of the program.
- Starting a timer program.
- Performance during program delay time.
- Cancelling a running timer program.
- Performance after the end of the program.
Describes preparations and the procedure for starting and ending hot-air disinfection.
- Preparations before hot-air disinfection.
- Starting and ending the hot-air disinfection.
Covers starting, stopping, pausing and cancelling time programs, creating new time programs, program and section management, value entry for program sections including set-point ramps and steps, tolerance range, repeating sections, and saving the time program.
- Starting an existing time program and performance during program delay time.
- Stopping, pausing and cancelling a running time program.
- Creating a new time program and managing programs and sections.
- Value entry for a program section: section duration, set-point ramp and step, tolerance range.
- Repeating one or several sections within a time program and saving the time program.
Covers starting and cancelling week programs, creating new week programs, program and section management, and value entry for program sections including weekday, start time, set-point ramp and step, and special controller functions.
- Starting an existing week program and cancelling a running week program.
- Creating a new week program and managing programs and sections.
- Value entry for a program section: set-point ramp and step modes, weekday, start time, setpoint entry.
- Special controller functions in week programs.
Provides an overview of notification and alarm messages, including alarm messages and messages concerning the humidity system, state of alarm, resetting alarms and listing active alarms, tolerance range settings, and activating/deactivating the audible alarm.
- Notifications and alarm messages overview.
- Messages concerning the humidity system.
- State of alarm and resetting an alarm, list of active alarms.
- Tolerance range settings.
- Activating/deactivating the audible alarm (alarm buzzer).
Describes the over temperature protective device (class 1), the overtemperature safety controller class 2/3.3 including selection between temperature limiter and temperature protection, safety controller modes, setting limit and offset values, undertemperature protection class 3.2, messages and measures in the state of alarm, and function check.
- Over temperature protective device (class 1).
- Overtemperature safety controller class 2/3.3 with selectable temperature limiter or temperature protection.
- Setting safety controller mode, limit value for overtemperature, and offset value for over- and undertemperature.
- Safety controller class 3.2 (undertemperature protection) and setting value in Limit mode.
- Message and measures in the state of alarm and function check.
Covers authorization levels and password protection, logging in and out, user change, password assignment and change, deleting passwords, new password assignment when password function is deactivated, and activation code.
- Authorization levels and password protection.
- Log in, log out and user change.
- Password assignment and password change.
- Deleting the password for an individual authorization level.
- New password assignment for service or admin authorization level when password function was deactivated.
- Activation code.
Covers selecting the controller menu language, setting date and time, selecting the temperature unit, display configuration including adapting display parameters and touchscreen calibration, network and communication (Ethernet, E-Mail), and USB data transfer.
- Selecting the controller’s menu language.
- Setting date and time and selecting the temperature unit.
- Display configuration: adapting display parameters and touchscreen calibration.
- Network and communication: Ethernet and E-Mail settings.
- USB menu for data transfer via USB interface.
Provides service contact page, current operating parameters, event list, technical chamber information, and self-test function.
- Service contact page.
- Current operating parameters.
- Event list.
- Technical chamber information.
- Self-test function.
Describes chart recorder views, showing and hiding the legend, switching between legend pages, showing and hiding specific indications, history display, and setting the parameters.
- Views of the chart recorder display.
- Show and hide legend and switch between legend pages.
- Show and hide specific indications.
- History display.
- Setting the parameters.
Describes the function of the humidifying and dehumidifying system, including humidifying system up to and from serial number 20250000010225/20250000010226, freshwater requirements, wastewater/condensate handling up to and from serial number 20250000010225/20250000010226, and the dehumidifying system.
- Humidifying system up to serial number 20250000010225 and from serial number 20250000010226.
- Freshwater requirements.
- Wastewater/condensate up to serial number 20250000010225 and from serial number 20250000010226.
- Dehumidifying system.
Covers defrosting during refrigerating operation.
- Defrosting procedure at refrigerating operation.
Covers anti-condensation protection measures.
- Anti-condensation protection.
Describes available options and accessories: APT-COM 4 Multi Management Software (Basic Edition), analog outputs for temperature and humidity, zero-voltage alarm contacts for collective alarm output, internal LED light, door lock, object temperature control with flexible Pt 100 temperature sensor (activation, sensitivity, maximum deviation), external freshwater and wastewater cans, and BINDER Pure Aqua Service.
- APT-COM 4 Multi Management Software (accessory) and Basic Edition.
- Analog outputs for temperature and humidity (option).
- Zero-voltage alarm contacts for collective alarm output (option).
- Internal LED light (option) and door lock (option).
- Object temperature control with flexible Pt 100 temperature sensor (option): activating/deactivating, setting sensitivity and maximum deviation.
- External freshwater and wastewater cans (accessories) with connections and pump box.
- BINDER Pure Aqua Service (accessory).
Covers cleaning, decontamination/chemical disinfection, and hot-air disinfection.
- Cleaning procedures.
- Decontamination / chemical disinfection.
- Hot-air disinfection.
Covers general information and personnel qualification, maintenance intervals and service, simple troubleshooting, and sending the chamber back to BINDER GmbH. Includes the maintenance reminder message after 8760 operating hours or two years.
- Maintenance intervals: regular maintenance at least once a year; steam humidifier at least once a year; cleaning condensers at least twice a year.
- Electrical testing: annual repeat inspection and test prior to initial startup and prior to resuming operation after maintenance or repair, recommended under EN 50678/VDE 0701 and EN 50699/VDE 0702.
- After 8760 operating hours or two years a maintenance message appears; after confirmation it reappears every two weeks until reset by BINDER Service.
- Simple troubleshooting by operating personnel without technical intervention or disassembly.
- Detailed troubleshooting, repair and testing by BINDER Service or qualified service partners.
Provides factory calibration and adjustment, over current protection, definition of usable volume, technical data for KBF/KBF-UL and KBF PRO, equipment and options extract, accessories and spare parts extract, and dimensions.
- Factory calibration and adjustment.
- Over current protection.
- Definition of usable volume.
- KBF / KBF-UL technical data and KBF PRO technical data.
- Equipment and options (extract), accessories and spare parts (extract), dimensions.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal temperature of the chamber is 100 °C / 212 °F. p.12 ↗ | Maximum rated operating temperature for the chamber is 100 °C; do not exceed this value. | KBF 260 example type plate; whole series |
| IP protection is IP 20 according to EN 60529. p.12 ↗ | The chamber is protected against solid objects larger than 12.5 mm but not against water; do not spray with water. | KBF 260 example type plate; whole series |
| Temperature safety device according to DIN 12880:2007, class 2/3.3. p.12 ↗ | The chamber is equipped with a temperature safety device class 2 or 3.3 per DIN 12880; verify the configured class for the application. | KBF 260 example type plate; whole series |
| Electrical ratings example: 220-240 V / 50 Hz, 1.40 kW / 5.8 A; also 220-240 V / 60 Hz and 120 V / 50 Hz variants; 1 N~. p.12 ↗ | Verify site voltage and frequency match the chamber rating before connection; use appropriate overcurrent protection. | KBF 260 and KBF 260-UL example type plates |
| Refrigerant R600A, charge -0.095 kg; contains hydrocarbon gases. p.12 ↗ | Refrigeration system contains flammable hydrocarbon refrigerant; maintenance on the refrigeration system must be performed by qualified personnel trained per EN 13313:2010. | KBF 260 and KBF 260-UL example type plates |
| Maximum operating pressure 11 bar. p.12 ↗ | Water or pressure connections must be rated for at least 11 bar; observe pressure limits when connecting to the water main. | KBF 260 example type plate |
| Regular maintenance work must be performed at least once a year. p.133 ↗ | Plan annual maintenance; warranty becomes void if maintenance is conducted by non-authorized personnel. | Whole series |
| Have regular maintenance work on the steam humidifier conducted at least once a year. p.134 ↗ | Humidifier operating behavior and maintenance intervals depend on available water quality and amount of steam produced. | Whole series |
| We recommend cleaning the condensers at least twice a year; a qualified technician must perform cleaning. p.134 ↗ | Schedule condenser cleaning twice per year; with increased dust in ambient air, clean the condenser several times a year by suction. | Whole series |
| Replace the door gasket only when cold. p.134 ↗ | Replacing the door gasket while warm may damage it. | Whole series |
| After 8760 operating hours or two years the maintenance message appears; after confirmation with the Confirm icon, the message window pops up again every two weeks until reset by BINDER Service. p.134 ↗ | Plan service reset by BINDER Service at or before 8760 operating hours or two years to avoid recurring reminders. | Whole series |
| Annual repeat inspection and a test prior to initial startup and prior to resuming operation after maintenance or repair are required; recommended under EN 50678/VDE 0701 and EN 50699/VDE 0702. p.133 ↗ | Include electrical safety testing in commissioning and annual maintenance schedules. | Whole series |
| Maintenance work at the refrigeration system must only be conducted by qualified personnel who underwent training in accordance with EN 13313:2010 (e.g. refrigeration technician with certified expert knowledge acc. to Regulation (EC) no. 303/2008). p.133 ↗ | Ensure refrigeration service personnel hold the required certification before any work on the refrigeration system. | Whole series |
| Condensate collection pan applies to KBF / KBF-UL 130 up to serial number 20250000010225. p.34 ↗ | For KBF/KBF-UL 130 chambers with serial number up to 20250000010225, install and use the condensate collection pan as described. | KBF 130, KBF 130-UL up to serial number 20250000010225 |
| Humidifying system design differs up to serial number 20250000010225 and from serial number 20250000010226. p.117 ↗ | Identify the chamber serial number to select the correct humidifying system description and spare parts. | KBF / KBF-UL / KBF PRO |
| Wastewater/condensate handling differs up to serial number 20250000010225 and from serial number 20250000010226. p.118 ↗ | Identify the chamber serial number to select the correct wastewater/condensate connection method. | KBF / KBF-UL / KBF PRO |
| Spacer for wall distance required from chamber size 260 on. p.33 ↗ | Maintain the specified wall distance for sizes 260 and larger to ensure proper operation. | KBF / KBF-UL / KBF PRO sizes 260 and larger |
| Flexible tilt protection kit is mounted on chamber sizes 130 and 470. p.33 ↗ | Install the tilt protection kit on sizes 130 and 470 to prevent tipping. | KBF / KBF-UL / KBF PRO sizes 130 and 470 |
| Humidity sensor resistance table gives maximum workplace threshold limit values and tolerated concentrations with permanent load for substances including ammonia (NH3: 20 ppm / 14 mg/m3 TLV; 5500 ppm / 4000 mg/m3 permanent load), acetone (500 ppm / 1200 mg/m3; 3300 ppm / 8000 mg/m3), ethanol (500 ppm / 960 mg/m3; 3500 ppm / 6000 mg/m3), formaldehyde (0.3 ppm / 0.37 mg/m3; 2400 ppm / 3000 mg/m3), ozone (0.1 ppm / 0.2 mg/m3; 0.5 ppm / 1 mg/m3), hydrochloric acid (2 ppm / 3 mg/m3; 300 ppm / 500 mg/m3), hydrogen sulphide (10 ppm / 15 mg/m3; 350 ppm / 500 mg/m3), nitrogen oxides (5 ppm / 9 mg/m3; 5 ppm / 9 mg/m3), sulphur dioxide (5 ppm / 13 mg/m3; 5 ppm / 13 mg/m3). p.20 ↗ | Do not expose the humidity sensor to substances above the listed tolerated concentrations; observe the maximum workplace threshold limit values. | Whole series |
| Tolerated error of the humidity sensor is +/- 2 % r.h. p.20 ↗ | Account for ±2 % relative humidity measurement tolerance when interpreting humidity readings. | Whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| After 8760 operating hours or two years the following message appears p.134 ↗ | Maintenance interval reached (8760 operating hours or two years). | Confirm with the Confirm icon; the message window will pop up again every two weeks until it is reset by BINDER Service. | Section 23.2 |