KT (E6.1) 12 2025

Technical Reference Summary

Covered Models & Document Scope

Covers BINDER KT series cooling incubators with Peltier refrigerating technology, HIT insulation and T4.12 program controller: KT 53 (KT053-230V, Art. 9020-0311/9120-0311), KT 53-UL (KT053UL-120V, 9020-0312/9120-0312), KT 115 (KT115-230V, 9020-0313/9120-0313), KT 115-UL (KT115UL-120V, 9020-0314/9120-0314), KT 170 (KT170-230V, 9020-0289/9120-0289) and KT 170-UL (KT170UL-120V, 9020-0310/9120-0310). Options covered include door heating, zero-voltage relay alarm/control outputs, object temperature display with flexible Pt 100 sensor, and APT-COM 4 software.

Section-by-Section Technical Breakdown

pp. 18-20
Chamber description and instrument panel

Describes the chamber layout, the APT.line preheating chamber system, the instrument panel and the chamber rear. The controller is the T4.12 program controller; optional APT-COM 4 software is referenced.

  • APT.line preheating chamber system for spatial and temporal temperature accuracy
  • Instrument panel with T4.12 controller
  • Chamber rear connections described in chap. 2.3
pp. 24-25
Installation, wall spacing and electrical connection

Covers installation of the chamber, required spacer for wall distance, and electrical connection. Sufficient ventilation and prescribed minimum distances must be observed to avoid overheating.

  • Spacer for wall distance (chap. 4.1)
  • Electrical connection (chap. 4.2)
  • Do not install in unventilated recesses; observe minimum distances per chap. 3.4
pp. 26
Start up and behavior when opening the door

Describes start-up of the chamber and the behavior of the controller/chamber when the door is opened. Door opening affects temperature control and alarm delay settings.

  • Start-up procedure (chap. 5)
  • Behavior when opening the door (chap. 5.1)
pp. 26-32
Functional overview of the T4.12 chamber controller

Introduces the T4.12 controller menu structure, operating modes, behavior during and after power failure, and information displays. Includes general menu, quick menu and User menu.

  • Menu structure: general menu, quick menu, User menu
  • Operating modes including control off and fixed value
  • Performance during and after power failure
  • Information displays
pp. 33-35
Configuration of optional equipment

Covers setting the optional door heating, switching optional zero-voltage relay control outputs on/off, functional test of the optional alarm output, and switching the optional object temperature display on/off.

  • Optional door heating setting
  • Zero-voltage relay control outputs on/off
  • Functional test of optional alarm output
  • Optional object temperature display on/off
pp. 36-38
Set-point entry in fixed value operating mode

Describes setting ranges and entry of set-points via the quick menu or the general menu in fixed value operating mode.

  • Setting ranges (chap. 8.1)
  • Set-point entry via quick menu
  • Set-point entry via general menu
pp. 39-57
Time programs

Covers starting, running, cancelling, creating and deleting time programs, including section handling, temperature setpoints, section duration, repetition, tolerance range, set-point ramp/step modes and optional zero-voltage relay outputs.

  • Start/run existing time program
  • Cancel running time program
  • Create new time program with sections, setpoints, durations and repetitions
  • Tolerance range and set-point ramp/step modes
  • Switch optional zero-voltage relay outputs within programs
pp. 58-71
Week programs

Covers starting, running, cancelling, creating and deleting week programs, including section handling, temperature setpoints, day of week, time of day, shift-point activity and optional zero-voltage relay outputs.

  • Start/run existing week program
  • Cancel running week program
  • Create new week program with day/time shift points
  • Switch optional zero-voltage relay outputs within programs
pp. 72-75
Key lock

Describes direct activation of the key lock, automatic key lock, and changing the password for unlocking the key lock.

  • Direct key lock activation
  • Automatic key lock
  • Change unlock password
pp. 76-85
General controller settings

Covers setup wizard, date/time, menu language, display brightness, temperature unit, data recording rate, factory reset, network configuration and chamber configuration display.

  • Setup wizard
  • Date and time settings
  • Menu language selection
  • Display brightness and temperature unit
  • Data recording rate
  • Factory reset
  • Network configuration and display
pp. 86-87
Data transfer via USB interface

Describes exporting data to a USB drive and importing data from a USB drive.

  • Export data to USB drive
  • Import data from USB drive
pp. 88-93
Notifications and alarms

Covers notification and alarm overviews, alarm status (set, acknowledged, cleared), confirming a set alarm, alarm configuration, active alarm list, history, buzzer activation/test, global alarm on/off and door alarm delay.

  • Alarm states: set, acknowledged, cleared
  • Confirm set alarm via Info > Reset
  • Alarm buzzer test and on/off
  • Activate/deactivate all alarm functions
  • Set delay time after opening the door
  • Optional zero-voltage relay alarm contact activated on door open, temperature deviation, power failure and main switch off
pp. 94
Event list

The event list records the entire sequence of alarm events (set – acknowledged – cleared) and other messages for the current day, with the most recent message at the end.

  • Access via Menu > Event list or User > View event list
  • Shows set, acknowledged and cleared alarm events with timestamps
pp. 95-97
Graphical display of measured values

Covers setting the sampling rate, defining the display range and selecting parameters for the graphical display.

  • Sampling rate setting
  • Display range definition
  • Parameter selection
pp. 97-101
Temperature safety devices

Describes the overtemperature protective device (class 1) and the overtemperature safety controller (temperature safety device class 3.1), including safety controller modes and setting.

  • Overtemperature protective device class 1
  • Overtemperature safety controller class 3.1
  • Safety controller modes and setting
pp. 102
Notes on refrigerating operation

Provides operating notes specific to the Peltier refrigerating system.

  • Peltier refrigerating operation notes
pp. 102-104
Options

Covers optional APT-COM 4 Multi Management Software, object temperature display with flexible Pt 100 sensor, zero-voltage relay alarm output and zero-voltage relay control outputs.

  • APT-COM 4 software option
  • Object temperature display with flexible Pt 100 sensor
  • Zero-voltage relay alarm output
  • Zero-voltage relay control outputs
pp. 105-107
Cleaning and decontamination

Covers cleaning and decontamination/chemical disinfection of the chamber.

  • Cleaning procedure
  • Decontamination / chemical disinfection
pp. 108-111
Maintenance, service, troubleshooting and testing

Covers personnel qualification, maintenance intervals, simple troubleshooting, electrical testing and returning the chamber to BINDER. Annual maintenance and annual electrical repeat inspection are required.

  • Maintenance at least once a year by licensed electricians or BINDER-authorized experts
  • Annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 recommended
  • Simple troubleshooting by operating personnel without disassembly
  • Replace door gasket only when cold
  • Clean Peltier fan grid several times a year in dusty ambient air
pp. 116-122
Technical description and data

Covers factory calibration and adjustment, overcurrent protection, definition of usable volume, KT (E6.1) technical data, equipment/options extract, spare parts/accessories extract and dimension drawings for KT 53, KT 115 and KT 170.

  • Factory calibration and adjustment
  • Overcurrent protection
  • Definition of usable volume
  • Technical data table
  • Equipment, options, spare parts and accessories extracts
  • Dimension drawings for KT/KT-UL 53, 115 and 170

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal temperature 100 °C / 212 °F; nominal power 0.80 kW; nominal current 3.0 A; nominal voltage 200-230 V / 50 Hz or 200-230 V / 60 Hz, 1 N~ (example type plate KT 170). p.1 ↗ Verify supply voltage and frequency within the nominal voltage range ±10% at the indicated power frequency before connection. KT 170 (example type plate); voltage class applies to 230 V models
IP protection 20 according to EN 60529; temperature safety device according to DIN 12880:2007, class 3.1. p.1 ↗ Chamber is protected against solid objects >12.5 mm but not water; install in splash-proof manner and do not wet the chamber. All KT models
Intended temperature application range for cultivation of microorganisms with narrow temperature optimum: 4 °C / 39.2 °F to 37 °C / 98.6 °F; typical uses include long-term storage at 4 °C, refrigerated incubation 20–25 °C, incubation at 37 °C and alternating 37 °C / 4 °C. p.1 ↗ Select the KT model and setpoints for operation within this range; do not exceed the intended use. KT series
Charging material must not contain explosive or flammable solvents; operating temperature must lie below the flash point or sublimation point of the charging material; no explosive mixture with air may form. p.1 ↗ Assess physical/chemical properties of the load before operation; do not introduce combustible dust or air-solvent mixtures. All KT models
Do not install the chamber in unventilated recesses; ensure sufficient ventilation for heat dispersal and observe prescribed minimum distances (chap. 3.4). p.1 ↗ Maintain wall clearances to prevent overheating and chamber damage. All KT models
Optional zero-voltage relay alarm contact is activated on alarm messages 'door open' and temperature deviations, as well as on power failure and when turning off the main power switch. p.14 ↗ Use this contact for remote alarm monitoring; note it also signals power loss and main switch off. Models with optional zero-voltage relay alarm output
Alarm states are 'set' (active, buzzer sounds if activated), 'acknowledged' (active, buzzer off, alarm icon remains) and 'cleared' (cause disappeared, icon gone, message remains in event list). p.14 ↗ Acknowledging an alarm mutes the buzzer but does not clear the alarm until the cause is remedied. All KT models with T4.12 controller
Alarm buzzer test: 'Buzzer test: On' emits an intermittent signal; set to 'Buzzer test: Off' to stop. p.14 ↗ Use Menu > Alarms > Buzzer test to verify audible alarm function. All KT models with T4.12 controller
Door alarm delay time can be set (chap. 14.5.5). p.14 ↗ Adjust delay to avoid nuisance door-open alarms during routine access. All KT models with T4.12 controller
Maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER; warranty becomes void if maintenance is conducted by non-authorized personnel. p.21 ↗ Schedule annual maintenance and use authorized personnel only. All KT models
Annual repeat electrical inspection and a test prior to initial startup and prior to resuming operation after maintenance or repair are required; EN 50678/VDE 0701 and EN 50699/VDE 0702 testing recommended. p.21 ↗ Include electrical safety testing in commissioning and annual maintenance plans. All KT models
Replace the door gasket only when cold; otherwise the door gasket may become damaged. p.21 ↗ Allow chamber to cool before gasket replacement. All KT models
With increased dust in ambient air, clean the Peltier fan grid (7) by suction or blowing several times a year. p.21 ↗ Dusty environments require more frequent fan-grid cleaning to maintain cooling performance. All KT models
Before maintenance work, turn off the chamber at the main power switch and disconnect the power plug; do not remove the rear panel. p.21 ↗ Isolate power before service; rear panel removal is prohibited. All KT models
Simple troubleshooting by operating personnel does not require technical intervention into the chamber nor disassembly of chamber parts; detailed troubleshooting and repair must be performed by BINDER Service or BINDER-qualified partners per the Service Manual. p.21 ↗ Limit on-site troubleshooting to the simple-troubleshooting table; escalate other faults. All KT models
After maintenance, the chamber must be tested prior to resuming operation. p.21 ↗ Include functional/electrical test after any maintenance or repair. All KT models
Type plate is located on the left chamber side, bottom right; UKCA label sticks next to the type plate on the left side, bottom right-hand. p.1 ↗ Use type plate data for model, serial number, voltage and safety-device identification. All KT models
UL chambers are certified by Underwriters Laboratories Inc. according to UL 61010-1 3rd Ed., CAN/CSA-C22.2 No. 61010-1 3rd Ed., and IEC/EN 61010-1:2014 3rd Ed. p.1 ↗ UL variants are intended for the North American market and carry UL certification. KT 53-UL, KT 115-UL, KT 170-UL

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Chamber without function. p.109 ↗ No power supply. Check connection to power supply. Check whether the chamber is turned on at the main power switch. chap. 21.3
Chamber without function. p.109 ↗ Wrong voltage. Check power supply for voltage of 100-120V or 200-240V. chap. 21.3
Chamber without function. p.109 ↗ Chamber fuse has responded. Check chamber fuse and replace it if appropriate. If it responds again, contact BINDER service. chap. 21.3
Chamber without function. p.109 ↗ Controller defective. Contact BINDER service. chap. 21.3
Nominal temperature exceeded by 10° due to chamber failure. p.109 ↗ Over temperature protective device (class 1) responds. Heating chap. 21.3
Overtemperature. Chamber heating permanently, exceeding the set-point p.109 ↗ Controller defective. Confirm the alarm (chap. 14.4). Contact BINDER service. chap. 21.3
Notification "Heating active". Alarm message "Temp. range" p.109 ↗ Semiconductor relay defective. Controller not adjusted. Confirm the alarm (chap. 14.4). Calibrate and adjust controller. chap. 21.3