BINDER KT 53
BINDER ·Made in Germany · In Active Production

BINDER KT 53

BINDER KT 53 — Incubators with Temperature 4…100 °C.

The BINDER KT 53 is a 53 L Peltier-cooled incubator (KT053-230V, Art. 9020-0311) operating from 4 to 100 °C with 0.1 K fluctuation and 0.3 K uniformity. It is the smallest chamber in the KT series, distinguished by a 0.4 kW nominal rating and 40 kg permitted load. It serves refrigerated incubation, long-term 4 °C storage, and alternating 37/4 °C protocols in microbiology.

Temperature
4…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KT Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KT 53 Current 4–1006363.55340015
KT 115 4–10065.571.510260030
KT 170 4–10065.5102.516360030

Full specifications

Core Performance & Functional Specifications
Temperature 4...100 °C
Interior volume 53 L
Load per rack 15 kg
Max. heat compensation 100 W
Nominal power 0.4 kW
Permitted load 40 kg (KT053-230V / KT053UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Back wall clearance 100 mm (KT053-230V / KT053UL-120V)
Side wall clearance 240 mm (KT053-230V / KT053UL-120V)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 2 min
Dimensions, Weight & Mechanical Design
Depth 630 mm (KT053-230V / KT053UL-120V)
Height 635 mm (KT053-230V / KT053UL-120V)
Interior width 400 mm (KT053-230V / KT053UL-120V)
Width 660 mm (KT053-230V / KT053UL-120V)
Net weight of the unit (empty) mfr 63 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 200...240 V
Unit fuse 10.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in operation

Installation, control and operation

Integration & Operating Notes

Mounting
Maintain 100 mm rear and 240 mm side wall clearance

The Peltier system rejects heat at the rear; insufficient clearance or installation in an unventilated recess causes overheating and chamber damage. Plan the bench or floor position with these minimum distances to walls and adjacent equipment before commissioning.

Data Sheet KT 53; Instruction Manual KT (E6.1), chap. 1.8 and 3.4
Electrical
Supply must stay within 200–240 V and match 50 Hz rating

The KT053-230V is rated 200...240 V, 1~, 50 Hz with a 10.0 A unit fuse and 0.4 kW nominal power. Verify the supply voltage and frequency before connection; the UL variant (KT053UL-120V) is a separate 100...120 V model with NEMA 5-20 plug and must not be connected to a 230 V circuit.

Data Sheet Model KT 53 en.pdf; Factory Specification Sheet
Process Limits
Load must be below flash point and free of flammable solvents

Charging material must not contain explosive or flammable solvents, and the operating temperature must lie below the flash or sublimation point of the load. Assess the physical and chemical properties of every charge before operation to avoid forming an explosive mixture with air inside the chamber.

Instruction Manual KT (E6.1), chap. 1.8/1.9
Control & I/O
Acknowledging an alarm only mutes the buzzer

Alarm states are set (active, buzzer on), acknowledged (active, buzzer off, icon remains) and cleared (cause gone). Acknowledging does not clear the alarm until the cause is remedied; operators must not assume the process is back in specification after pressing acknowledge.

Instruction Manual KT (E6.1), chap. 14.3
Maintenance
Annual maintenance by BINDER-authorized personnel is mandatory

Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER; warranty is void if non-authorized personnel conduct it. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required, plus a test before initial startup and after any repair.

Instruction Manual KT (E6.1), chap. 21.1/21.2
Maintenance
Clean Peltier fan grid several times a year in dusty air

With increased dust in ambient air, the Peltier fan grid must be cleaned by suction or blowing several times a year to maintain cooling performance. Before any maintenance, switch off at the main power switch, disconnect the plug, and do not remove the rear panel.

Instruction Manual KT (E6.1), chap. 21.2
Maintenance
Replace door gasket only when the chamber is cold

Replacing the door gasket while the chamber is warm can damage the gasket. Allow the chamber to cool to ambient before gasket replacement, and test the chamber before resuming operation after any maintenance or repair.

Instruction Manual KT (E6.1), chap. 21.2
Environment
IP20 enclosure requires splash-proof installation

The chamber is protected against solid objects larger than 12.5 mm but not against water (IP20 per EN 60529). Install it in a dry, splash-proof location and do not wet the chamber during cleaning or decontamination.

Instruction Manual KT (E6.1), chap. 1.6

Selection & Integration Q&A

When should I choose the KT 53 instead of the KT 115 or KT 170?

Pick the KT 53 when 53 L interior (400×630×635 mm) and 40 kg load capacity cover your workload and bench space is limited. Move to KT 115 or KT 170 only for larger batches; all three share the 4...100 °C range and Peltier technology.

What supply and plug does the KT 53 require, and can I use the UL variant on a 230 V line?

The KT053-230V needs 200...240 V, 1~, 50 Hz with CEE 7/7 plug and 10.0 A fuse. The KT053UL-120V is a separate 100...120 V model with NEMA 5-20 plug and UL certification; it must not be connected to a 230 V supply.

What mounting clearances and ambient conditions must I provide for the KT 53?

Keep 100 mm clearance at the rear and 240 mm at the sides for Peltier heat rejection, and do not install in an unventilated recess. The IP20 enclosure must be in a dry, splash-proof location; dusty air requires more frequent fan-grid cleaning.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance by BINDER-authorized personnel and annual electrical repeat inspection per EN 50678/EN 50699 are mandatory. Optional items not in base delivery include APT-COM 4 software, Pt 100 object temperature sensor, door heating and zero-voltage relay outputs.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KT 53 en
Datasheet ·379,4 КБ ·7 pp.
View PDF